Medical physics最新文献

筛选
英文 中文
Beam collimation and filtration optimization for a novel orthovoltage radiotherapy system 一种新型正电压放射治疗系统的光束准直和滤波优化。
IF 3.2 2区 医学
Medical physics Pub Date : 2025-02-06 DOI: 10.1002/mp.17662
Nathan Clements, Olivia Masella, Deae-Eddine Krim, Lane Braun, Magdalena Bazalova-Carter
{"title":"Beam collimation and filtration optimization for a novel orthovoltage radiotherapy system","authors":"Nathan Clements, Olivia Masella, Deae-Eddine Krim, Lane Braun, Magdalena Bazalova-Carter","doi":"10.1002/mp.17662","DOIUrl":"10.1002/mp.17662","url":null,"abstract":"<div>\u0000 \u0000 \u0000 <section>\u0000 \u0000 <h3> Background</h3>\u0000 \u0000 <p>The inaccessibility of clinical linear accelerators in low- and middle-income countries creates a need for low-cost alternatives. Kilovoltage (kV) x-ray tubes have shown promise as a source that could meet this need. However, performing radiotherapy with a kV x-ray tube has numerous difficulties, including high skin dose, rapid dose fall-off, and low dose rates. These limitations create a need for highly effective beam collimation and filtration.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Purpose</h3>\u0000 \u0000 <p>To improve the treatment potential of a novel kV x-ray system by optimizing an iris collimator and beam filtration using Bayesian techniques and Monte Carlo (MC) simulations.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Methods</h3>\u0000 \u0000 <p>The Kilovoltage Optimized AcceLerated Adaptive therapy system's current beam configuration consists of a 225 kVp x-ray tube, a 12-leaflet tungsten iris collimator, and a 0.1 mm copper filter. A Bayesian optimization was performed for the large and small focal spot sizes of the kV x-ray tube source at 220 kVp using TopasOpt, an open-source library for optimization in TOPAS. Collimator thickness, copper filter thickness, source-to-collimator distance (SCD), and source-to-surface distance (SSD) were the variables considered in the optimization. The objective function was designed to maximize the dose rate and the dose at a depth of 5 cm while minimizing the beam penumbra width and the out-of-field dose (OFD), all evaluated in a water phantom. Post-optimization, the optimal beam configuration was simulated and compared to the existing configuration.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Results</h3>\u0000 \u0000 <p>The optimal collimation setup consisted of 2.5 mm thick tungsten leaflets for the iris collimator and a 350 mm SSD for both focal spot sizes. The optimal copper filtration was 0.22 mm for the large focal spot and 0.15 mm for the small focal spot, with a SCD of 148.5 mm for the large focal spot and 125.8 mm for the small focal spot. For the large focal spot, the surface dose rate decreased by 9.4%, while the PDD at 5cm depth (<span></span><math>\u0000 <semantics>\u0000 <msub>\u0000 <mtext>PDD</mtext>\u0000 <mrow>\u0000 <mn>5</mn>\u0000 <mi>c</mi>\u0000 <mi>m</mi>\u0000 </mrow>\u0000 </msub>\u0000 <annotation>$text{PDD}_{5textnormal {cm}}$</annotation>\u0000 </semantics></math>) increased by 7.7% compared ","PeriodicalId":18384,"journal":{"name":"Medical physics","volume":"52 5","pages":"3204-3215"},"PeriodicalIF":3.2,"publicationDate":"2025-02-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/mp.17662","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"143257701","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Enhancing gamma-ray detection: Processing grooved microstructures on LYSO crystal with femtosecond laser 增强伽玛射线探测:用飞秒激光处理LYSO晶体的沟槽微结构。
IF 3.2 2区 医学
Medical physics Pub Date : 2025-02-06 DOI: 10.1002/mp.17665
Xi Zhang, Xin Yu, Hua Cheng, Yuli Wang, Hamid Sabet, Siwei Xie, Jianfeng Xu, Qiyu Peng
{"title":"Enhancing gamma-ray detection: Processing grooved microstructures on LYSO crystal with femtosecond laser","authors":"Xi Zhang, Xin Yu, Hua Cheng, Yuli Wang, Hamid Sabet, Siwei Xie, Jianfeng Xu, Qiyu Peng","doi":"10.1002/mp.17665","DOIUrl":"10.1002/mp.17665","url":null,"abstract":"<div>\u0000 \u0000 \u0000 <section>\u0000 \u0000 <h3> Background</h3>\u0000 \u0000 <p>Gamma-ray detection plays a crucial role in the fields of biomedicine, space exploration, national defense, and security. High-precision gamma photon detection relies on scintillation crystals, which attenuate gamma rays through mechanisms such as photoelectric effect and Compton scattering. These interactions generate light signals within the scintillation crystal, which are subsequently converted into electronic signals using photodetectors, enabling accurate readout, and analysis.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Purpose</h3>\u0000 \u0000 <p>Improving the readout efficiency of visible photons in crystal detectors can significantly improve the efficiency of gamma-ray detection. Scintillator crystals are usually hard and brittle materials, which are difficult to process. In this paper, we innovatively propose the method of using a femtosecond laser to process grooved microstructures on the light output surface of scintillator crystals to improve the detection efficiency, and thus enhance the comprehensive performance of gamma-ray detection.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Methods</h3>\u0000 \u0000 <p>Optical simulation software is first used to explore the enhancement of the light output performance by the grooved microstructures. Subsequently, a 5-dimension system for femtosecond laser processing of scintillator crystals was constructed, which can achieve accurate processing of grooved structures. Finally, the feasibility of the study was verified by applying grooved microstructure on crystal bars and crystal arrays.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Results</h3>\u0000 \u0000 <p>TracePro simulation results showed an average efficiency improvement in light output of 33.56% within the groove parameters: spacing from 20 to 140 µm, depth from 8 to 28 µm, and width from 10 to 30 µm. A custom-designed readout electronic system for gamma detection and a laser processing platform was then constructed to evaluate the feasibility of applying grooved structures to the lutetium-yttrium oxyorthosilicate (LYSO) crystal surface. According to the simulation results, 12 groups of crystal bars were fabricated with spacings from 60 to 140 µm, depths from 7 to 16 µm, and widths from 11 to 14 µm. Experimental results showed an average improvement of 20.4% in light output for the crystal bars, and that of the crystal arrays can be improved by 6.85% on average.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Conclusions</h3>\u0000 \u0000 <p>This study introduces a method of using femtosecond lasers to fabricate grooved microstructures on LYSO crystal surfaces, which has demonstrated a significant improvement i","PeriodicalId":18384,"journal":{"name":"Medical physics","volume":"52 4","pages":"2644-2660"},"PeriodicalIF":3.2,"publicationDate":"2025-02-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"143257636","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Spatial resolution improvements for photon counting detectors using coincidence counting and frequency weighted reconstruction 利用符合计数和频率加权重建提高光子计数探测器的空间分辨率。
IF 3.2 2区 医学
Medical physics Pub Date : 2025-02-06 DOI: 10.1002/mp.17664
Scott S. Hsieh, Katsuyuki Taguchi, Shuai Leng, Cynthia H. McCollough
{"title":"Spatial resolution improvements for photon counting detectors using coincidence counting and frequency weighted reconstruction","authors":"Scott S. Hsieh,&nbsp;Katsuyuki Taguchi,&nbsp;Shuai Leng,&nbsp;Cynthia H. McCollough","doi":"10.1002/mp.17664","DOIUrl":"10.1002/mp.17664","url":null,"abstract":"<div>\u0000 \u0000 \u0000 <section>\u0000 \u0000 <h3> Background</h3>\u0000 \u0000 <p>Photon counting X-ray detectors (PCDs) provide better spatial resolution than energy integrating X-ray detectors, but even higher resolution is desired in some applications. Certain charge sharing compensation techniques such as coincidence counting preferentially detect photons that arrive at the boundary between pixels, and this could be used for subpixel localization of incident photons and enhanced spatial resolution.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Purpose</h3>\u0000 \u0000 <p>To estimate improvements to spatial resolution and detective quantum efficiency that are possible when using coincidence counting for high-resolution, non-spectral imaging.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Methods</h3>\u0000 \u0000 <p>The modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) were estimated using numerical simulations of a two-dimensional parallel-beam CT system. Coincidence counters were modeled either geometrically or with Monte Carlo simulations. The geometric model consisted of narrow coincidence counters that interlaced with wider ordinary pixels, and showed that using standard filtered backprojection decreases low-frequency DQE, but that a frequency weighting technique could be used to restore DQE. The Monte Carlo simulations were used to estimate the possible improvements that could be expected from real systems. The pixel pitch was 0.25 mm and the source apertures considered were 0, 0.125, and 0.5 mm. A numerical stent phantom was also used to illustrate possible improvements.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Results</h3>\u0000 \u0000 <p>Assuming a 0.125 mm source aperture and the Monte Carlo model, the limiting MTF (at 10%) increased from 20 to 40 lp/cm using coincidence counters. This can be explained by the increased sampling (and higher Nyquist limit) possible from coincidence counters. In the geometric model, coincidence counters are compared to conventional double sampling techniques such as in-plane flying focal spot, and the MTF at 38 lp/cm increased from 5% to 53%. Without the frequency weighting technique, low-frequency DQE was reduced by about 20%, but these losses are recovered with frequency weighting. Improvements are much more modest with the 0.25 mm source aperture because the system becomes source-limited.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Conclusions</h3>\u0000 \u0000 <p>Coincidence counting could be used to increase spatial resolution in PCDs. The increases in system resolution could be large if a high-resolution X-ray source were available.</p>\u0000 </section>\u0000 </div>","PeriodicalId":18384,"journal":{"name":"Medical physics","volume":"52 4","pages":"2052-2063"},"PeriodicalIF":3.2,"publicationDate":"2025-02-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"143256489","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Absorbed dose differences between twin fetuses for pregnancy patients in CT examinations 双胎妊娠患者CT检查吸收剂量差异分析。
IF 3.2 2区 医学
Medical physics Pub Date : 2025-02-05 DOI: 10.1002/mp.17659
Shuiyin Qu, Haoran Jia, Habib Zaidi, Tianwu Xie
{"title":"Absorbed dose differences between twin fetuses for pregnancy patients in CT examinations","authors":"Shuiyin Qu,&nbsp;Haoran Jia,&nbsp;Habib Zaidi,&nbsp;Tianwu Xie","doi":"10.1002/mp.17659","DOIUrl":"10.1002/mp.17659","url":null,"abstract":"<div>\u0000 \u0000 \u0000 <section>\u0000 \u0000 <h3> Background</h3>\u0000 \u0000 <p>Estimation of the radiation dose to the fetus is essential for the assessment of radiation risks and benefits to pregnant patients undergoing radiological examinations. During the past decade, the global twinning rate has soared resulting from embryo assistance and increased delivery age. However, to the best of our knowledge, radiation dosimetry in multiple pregnancies from radiological imaging has never been reported before.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Purpose</h3>\u0000 \u0000 <p>The purpose of this study is to develop personalized computational models for twin fetuses based on clinical CT images of real pregnant patients and to estimate personalized radiation doses for twin fetuses from abdominal/pelvic CT examinations.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Methods</h3>\u0000 \u0000 <p>Personalized computational phantoms representing pregnant females with twins at the second and third trimesters were constructed based on CT images of two pregnant patients. Monte Carlo calculations were performed using the MCNP transport code and three validated CT scanners to estimate the radiation dose of twin fetuses during abdominal and pelvic CT examinations.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Results</h3>\u0000 \u0000 <p>The absorbed fetal organ dose was calculated and compared between twins. For the same patient, the absolute difference in fetal organ dose between twins varies between 0.63% and 39.64% with an average value of 12.85%. The estimated total-body dose differences for twin fetuses were 11.55% and 7.51%, respectively, for pregnant patients at 22 and 30 weeks gestational age.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Conclusion</h3>\u0000 \u0000 <p>The variations of body weight and organ mass affect the absorbed dose of twin fetuses. Personalized computational models provide more accurate fetal radiation dosimetry estimates for pregnant patients with twins. This work also contributes to a better understanding of model-induced uncertainties in external radiation dosimetry for the developing fetus.</p>\u0000 </section>\u0000 </div>","PeriodicalId":18384,"journal":{"name":"Medical physics","volume":"52 4","pages":"2632-2643"},"PeriodicalIF":3.2,"publicationDate":"2025-02-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"143257696","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Impact of porous lung substitute on linear energy transfer (LET) assessed via Monte Carlo simulation and CR-39 measurement with a carbon-ion beam 通过蒙特卡罗模拟和碳离子束CR-39测量评估多孔肺替代品对线性能量传递(LET)的影响。
IF 3.2 2区 医学
Medical physics Pub Date : 2025-02-05 DOI: 10.1002/mp.17671
Jingfang Zhao, Bo Chen, Michael Moyers, Zhiyuan Yang, Shiyan Yang, Weiwei Wang
{"title":"Impact of porous lung substitute on linear energy transfer (LET) assessed via Monte Carlo simulation and CR-39 measurement with a carbon-ion beam","authors":"Jingfang Zhao,&nbsp;Bo Chen,&nbsp;Michael Moyers,&nbsp;Zhiyuan Yang,&nbsp;Shiyan Yang,&nbsp;Weiwei Wang","doi":"10.1002/mp.17671","DOIUrl":"10.1002/mp.17671","url":null,"abstract":"&lt;div&gt;\u0000 \u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Background&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;Carbon-ion beam radiotherapy offers substantial physical and biological advantages due to its distinct Bragg peak (BP) depth dose distribution and higher linear energy transfer (LET) in the peak region that enhances its efficacy in tumor eradication compared to x-ray beams. Porous structures, such as those found in lung and lung-equivalent tissues, unfortunately, introduce significant uncertainties in both dose and LET distributions, which current treatment planning systems (TPS) inadequately address.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Purpose&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;This study aims to investigate the effects of porous lung-equivalent structures on LET distribution using Monte Carlo (MC) simulations and CR-39 measurements. It seeks to understand how porous structures influence LET spectra and dose-averaged LET (LET&lt;sub&gt;d&lt;/sub&gt;) in carbon-ion beams.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Methods&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;A Gammex LN300 phantom and a binary voxel virtual phantom composed of water and air were used to represent lung-equivalent tissues for measurements and MC simulations. LET spectra measured with CR-39 at different depths within the LN300 slabs were compared with MC-calculated LET&lt;sub&gt;d&lt;/sub&gt; distributions. The impact of porous structures on dose and LET&lt;sub&gt;d&lt;/sub&gt; distributions was evaluated using various beam configurations, including single-beam and multi-beam setups. Additionally, a convolution method with modulation power (&lt;i&gt;P&lt;/i&gt;&lt;sub&gt;mod&lt;/sub&gt;) was proposed to improve LET&lt;sub&gt;d&lt;/sub&gt; prediction in porous media.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Results&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;The study demonstrated that porous structures broaden both the dose and LET&lt;sub&gt;d&lt;/sub&gt; distributions, especially around the BP region. Multiple beam angles helped mitigate dose degradation but did not resolve discrepancies in the LET&lt;sub&gt;d&lt;/sub&gt; distributions. Compared with calculation results based on CT images, intensity-modulated particle therapy (IMPT) using a distal LET&lt;sub&gt;d&lt;/sub&gt; patching method in porous structure increased the median LET&lt;sub&gt;d&lt;/sub&gt; in the target from 67.2 to 69.6 keV/µm, and the minimum LET&lt;sub&gt;d&lt;/sub&gt; from 51.5 to 58.0 keV/µm, respectively. Moreover, to improve the prediction of LET&lt;sub&gt;d&lt;/sub&gt; in porous structures, analytical convolution-based predictions showed good agreement with the MC simulations, with mean LET&lt;sub&gt;d&lt;/sub&gt; deviations of −1.9% ± 1.6% in the plateau, −3.1% ± 4.9% in the BP, and −1.1% ± 7.7% in the tail region.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Conclusions&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;Porous lung-equivalent structures significantly affect LET&lt;sub&gt;d&lt;/sub&gt; distribu","PeriodicalId":18384,"journal":{"name":"Medical physics","volume":"52 4","pages":"2593-2605"},"PeriodicalIF":3.2,"publicationDate":"2025-02-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"143256092","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Fast and accurate peak skin dose estimation method for interventional fluoroscopy patients 介入透视患者皮肤峰值剂量快速准确估算方法。
IF 3.2 2区 医学
Medical physics Pub Date : 2025-02-05 DOI: 10.1002/mp.17667
Haegin Han, Seth W. Streitmatter, Cari M. Kitahara, Choonsik Lee
{"title":"Fast and accurate peak skin dose estimation method for interventional fluoroscopy patients","authors":"Haegin Han,&nbsp;Seth W. Streitmatter,&nbsp;Cari M. Kitahara,&nbsp;Choonsik Lee","doi":"10.1002/mp.17667","DOIUrl":"10.1002/mp.17667","url":null,"abstract":"&lt;div&gt;\u0000 \u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Background&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;In fluoroscopy, particularly fluoroscopically guided interventions (FGIs), accurate estimation of peak skin dose (PSD) is crucial for identifying potential radiation-induced skin injuries. Most current methodologies for PSD calculation methods rely on analytical methods, which may introduce uncertainty due to their limited consideration of the complexities of x-ray beam conditions, patient geometry, and positioning. Methods based on full Monte Carlo (MC) simulations can enhance accuracy, but their practical application is limited due to the intensive requirement of computational resources and time.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Purpose&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;We aimed to develop a novel method that combines MC simulation with a noise reduction technique to calculate PSD, as well as skin dose distributions, more efficiently and accurately. The goal was to overcome the limitations of current methods, providing a more practical solution for clinical and academic use.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Methods&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;Our method to calculate the PSD and skin dose distributions consists of two steps of rough MC simulation and iterative noise reduction. The performance of the methodology was demonstrated for six fluoroscopy scenarios, with results compared against those from full MC simulation with high particle history, which is considered a gold standard for radiation dosimetry relative to conventional analytical methods.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Results&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;Our method was demonstrated for various fluoroscopy scenarios, and the result showed that the iterative noise reduction procedure successfully estimates PSD and skin dose distribution for rough MC simulations with a maximum dose statistical error of up to 20%. For successful dose estimations, PSD discrepancies from the values obtained by full MC simulation were within 3%, and voxel-wise dose differences in skin dose distributions were less than 10% of the average skin dose. The computation time of our method was on the order of a few seconds on a personal computer, which is estimated to be at least 10&lt;sup&gt;4&lt;/sup&gt; times faster than full MC simulation when using the same computing resources.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Conclusion&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;Our method rapidly and accurately calculates PSD and skin dose distribution, making it a useful tool for research and clinical applications. The planned integration of our method into the National Cancer Institute Dosimetry System for Ra","PeriodicalId":18384,"journal":{"name":"Medical physics","volume":"52 4","pages":"2551-2559"},"PeriodicalIF":3.2,"publicationDate":"2025-02-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/mp.17667","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"143257658","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Chan–Vese aided fuzzy C-means approach for whole breast and fibroglandular tissue segmentation: Preliminary application to real-world breast MRI Chan-Vese辅助模糊c均值法在全乳及纤维腺组织分割中的初步应用。
IF 3.2 2区 医学
Medical physics Pub Date : 2025-02-05 DOI: 10.1002/mp.17660
Syed Furqan Qadri, Chao Rong, Mubashir Ahmad, Jing Li, Salman Qadri, Syeda Shamaila Zareen, Zeyu Zhuang, Salabat Khan, Hongxiang Lin
{"title":"Chan–Vese aided fuzzy C-means approach for whole breast and fibroglandular tissue segmentation: Preliminary application to real-world breast MRI","authors":"Syed Furqan Qadri,&nbsp;Chao Rong,&nbsp;Mubashir Ahmad,&nbsp;Jing Li,&nbsp;Salman Qadri,&nbsp;Syeda Shamaila Zareen,&nbsp;Zeyu Zhuang,&nbsp;Salabat Khan,&nbsp;Hongxiang Lin","doi":"10.1002/mp.17660","DOIUrl":"10.1002/mp.17660","url":null,"abstract":"<div>\u0000 \u0000 \u0000 <section>\u0000 \u0000 <h3> Background</h3>\u0000 \u0000 <p>Magnetic resonance imaging (MRI) is a highly sensitive modality for diagnosing breast cancer, providing an expanding range of clinical usages that are crucial for the care of women at elevated risk of breast cancer development. Segmentation of the whole breast and fibroglandular tissue (FGT), used to evaluate breast cancer risk, is often manually delineated by radiologists in clinical practice. In this paper, we aim to substitute handcrafted breast density segmentation and categorization. The traditional fuzzy C-means (FCM) enable automatic segmentation but may be susceptible to heterogeneity or sparse FGT distribution in MRI.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Purpose</h3>\u0000 \u0000 <p>We develop a new automated technique for the segmentation of whole breast and FGT for the coronal-view MRI.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Methods</h3>\u0000 \u0000 <p>We propose a Chan–Vese (CV) aided FCM segmentation approach for estimating the FGT in the whole breast using fat-suppressed (FS) precontrast T1-weighted breast MRI. We present a methodology pipeline comprising region-of-interest (ROI) extraction, nonparametric non-uniform intensity normalization N4 algorithm-based intensity inhomogeneity correction, skin-layer extraction, and then whole breast and FGT segmentation. Our approach involves the FCM algorithm to assign membership degrees to pixels, distinguishing FGT regions from surrounding adipose tissues by assessing their probability of belonging to specific FGT regions, and subsequently, the region-based active contour CV model leverages these membership degrees to direct contour evolution and enhance segmentation boundaries. The proposed method adeptly tackles common challenges in MRI, including blurred edges, low contrast, and intensity inhomogeneity, with efficiency.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Results</h3>\u0000 \u0000 <p>We evaluated our approach on the Duke Breast Cancer MRI data (DBCM-data) and achieved good segmentation accuracy in terms of Dice similarity coefficient (DSC), Intersection-over-Union (IoU), and Sensitivity (SEN). Our method demonstrates significant accuracy, achieving a DSC (%) of 93.2 ± 3.3 and 84.1 ± 4.9, IoU (%) of 86.4 ± 3.5 and 73.2 ± 5.1, and SEN 87.3 ± 4.1 and 76.7 ± 4.1 for the segmentations of whole breast and FGT, respectively.</p>\u0000 </section>\u0000 \u0000 <section>\u0000 \u0000 <h3> Conclusion</h3>\u0000 \u0000 <p>Our results demonstrated that the CV-aided FCM approach significantly outperformed the existing methods and resulted in significantly more accurate whole breast and FGT segmentation in MRI data.</p>\u0000 </section>\u0000 </div>","PeriodicalId":18384,"journal":{"name":"Medical physics","volume":"52 5","pages":"2950-2960"},"PeriodicalIF":3.2,"publicationDate":"2025-02-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"143257631","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Patient- and fraction-specific magnetic resonance volume reconstruction from orthogonal images with generative adversarial networks 利用生成式对抗网络从正交图像重建患者和分部特定的磁共振容积。
IF 3.2 2区 医学
Medical physics Pub Date : 2025-02-04 DOI: 10.1002/mp.17668
Hideaki Hirashima, Dejun Zhou, Nobutaka Mukumoto, Haruo Inokuchi, Nobunari Hamaura, Mutsumi Yamagishi, Mai Sakagami, Naoki Mukumoto, Mitsuhiro Nakamura, Keiko Shibuya
{"title":"Patient- and fraction-specific magnetic resonance volume reconstruction from orthogonal images with generative adversarial networks","authors":"Hideaki Hirashima,&nbsp;Dejun Zhou,&nbsp;Nobutaka Mukumoto,&nbsp;Haruo Inokuchi,&nbsp;Nobunari Hamaura,&nbsp;Mutsumi Yamagishi,&nbsp;Mai Sakagami,&nbsp;Naoki Mukumoto,&nbsp;Mitsuhiro Nakamura,&nbsp;Keiko Shibuya","doi":"10.1002/mp.17668","DOIUrl":"10.1002/mp.17668","url":null,"abstract":"&lt;div&gt;\u0000 \u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Background&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;Although deep learning (DL) methods for reconstructing 3D magnetic resonance (MR) volumes from 2D MR images yield promising results, they require large amounts of training data to perform effectively. To overcome this challenge, fine-tuning—a transfer learning technique particularly effective for small datasets—presents a robust solution for developing personalized DL models.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Purpose&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;A 2D to 3D conditional generative adversarial network (GAN) model with a patient- and fraction-specific fine-tuning workflow was developed to reconstruct synthetic 3D MR volumes using orthogonal 2D MR images for online dose adaptation.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Methods&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;A total of 2473 3D MR volumes were collected from 43 patients. The training and test datasets were separated into 34 and 9 patients, respectively. All patients underwent MR-guided adaptive radiotherapy using the same imaging protocol. The population data contained 2047 3D MR volumes from the training dataset. Population data were used to train the population-based GAN model. For each fraction of the remaining patients, the population model was fine-tuned with the 3D MR volumes acquired before beam irradiation of the fraction, named the fine-tuned model. The performance of the fine-tuned model was tested using the 3D MR volume acquired immediately after the beam delivery of the fraction. The model's input was a pair of axial and sagittal MR images at the isocenter level, and the output was a 3D MR volume. Model performance was evaluated using the structural similarity index measure (SSIM), peak signal-to-noise ratio (PSNR), root mean square error (RMSE), and mean absolute error (MAE). Moreover, the prostate, bladder, and rectum in the predicted MR images were manually segmented. To assess geometric accuracy, the 2D Dice Similarity Coefficient (DSC) and 2D Hausdorff Distance (HD) were calculated.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Results&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;A total of 84 3D MR volumes were included in the performance testing. The mean ± standard deviation (SD) of SSIM, PSNR, RMSE, and MAE were 0.64 ± 0.10, 93.9 ± 1.5 dB, 0.050 ± 0.009, and 0.036 ± 0.007 for the population model and 0.72 ± 0.09, 96.2 ± 1.8 dB, 0.041 ± 0.007, and 0.028 ± 0.006 for the fine-tuned model, respectively. The image quality of the fine-tuned model was significantly better than that of the population model (&lt;i&gt;p&lt;/i&gt; &lt; 0.05). The mean ± SD of DSC and HD of the population model were 0.79 ± 0.08 and 1.70 ± 2.35 mm for prostate, 0.81 ± 0.10 and 2.75 ± 1.53 mm for bladder, and 0.72 ± 0.08 and 1.93 ± 0.59 mm for rectum. Contrarily, the mean ± SD","PeriodicalId":18384,"journal":{"name":"Medical physics","volume":"52 5","pages":"2785-2795"},"PeriodicalIF":3.2,"publicationDate":"2025-02-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"143191659","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Electron beam monitoring of a modified conventional medical accelerator with a portable current transformer system traceable to primary electrical standards 可追溯至初级电气标准的便携式电流互感器系统改进的传统医用加速器电子束监测。
IF 3.2 2区 医学
Medical physics Pub Date : 2025-02-04 DOI: 10.1002/mp.17653
James Renaud, Bryan Richard Muir, Andrew Williams, Malcolm McEwen
{"title":"Electron beam monitoring of a modified conventional medical accelerator with a portable current transformer system traceable to primary electrical standards","authors":"James Renaud,&nbsp;Bryan Richard Muir,&nbsp;Andrew Williams,&nbsp;Malcolm McEwen","doi":"10.1002/mp.17653","DOIUrl":"10.1002/mp.17653","url":null,"abstract":"&lt;div&gt;\u0000 \u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Background&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;Ultra-high dose rate radiotherapy (UHDR) delivers therapeutic doses at rates &gt;40 Gy/s in a fraction of a second, aiming to enhance the therapeutic ratio through the FLASH effect. The substantial increase in UHDR beam current poses serious challenges for conventional active dosimeters. Integrating current transformers (ICT) offer a nondestructive solution for accurate monitoring, enabling the type of fast transient readout that will be crucial for UHDR treatment verification.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Purpose&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;The aim of this study is to build and characterize a clinically deployable ICT system and to develop an accurate calibration methodology for absolute charge determination that is traceable to primary electrical standards.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Methods&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;The ICT was constructed from a Super MuMetal&lt;sup&gt;®&lt;/sup&gt; toroid, and its secondary winding was made from 50 Ω coaxial cable. A 3D-printed case with an internal conductive coating shields the toroid assembly from interference. The ICT readout involves a custom differential amplifier and a commercial flash analog-to-digital converter. The system was calibrated using a bespoke sub-µs current pulser in the range of (2 to 16) mA, which itself is traceable to electrical standards via an in-house built electrometer with a calibrated feedback capacitor. The performance of the ICT was evaluated as a milliampere-scale beam monitor against concurrent absorbed dose graphite calorimetry irradiation measurements acquired on a specially tuned medical accelerator for UHDR delivery.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Results&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;The ICT responses for nominal test pulses, generated by a function generator and a current pulser, exhibited accurate reproduction of rise and fall times within the 1 ns sampling frequency. A systematic droop effect of 0.6(1)%/µs was observed but is accounted for through the calibration chain. The calibration of the current pulser exhibited a repeatability typically better than 0.05%, with a slowly-varying leakage that can be subtracted using a linear regression of the leakage current. The ICT charge calibration demonstrated a repeatability in the range of (0.5 to &lt; 0.05)% for charge per pulse values in the range of (0.5 to 50) nC, respectively. The ICT response showed a strong linear relationship (adj-R&lt;sup&gt;2&lt;/sup&gt; = 0.99997) to charge per pulse. The in-beam comparison with a graphite calorimeter demonstrated the effectiveness of the ICT as an online beam monitor, independent of pulse repetition frequency in the rang","PeriodicalId":18384,"journal":{"name":"Medical physics","volume":"52 4","pages":"2581-2592"},"PeriodicalIF":3.2,"publicationDate":"2025-02-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/mp.17653","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"143191696","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
A comparison of carbon ions versus protons for integrated mode ion imaging 碳离子与质子集成模式离子成像的比较。
IF 3.2 2区 医学
Medical physics Pub Date : 2025-02-04 DOI: 10.1002/mp.17645
Mikaël Simard, Ryan Fullarton, Lennart Volz, Christoph Schuy, Daniel G. Robertson, Allison Toltz, Colin Baker, Sam Beddar, Christian Graeff, Charles-Antoine Collins Fekete
{"title":"A comparison of carbon ions versus protons for integrated mode ion imaging","authors":"Mikaël Simard,&nbsp;Ryan Fullarton,&nbsp;Lennart Volz,&nbsp;Christoph Schuy,&nbsp;Daniel G. Robertson,&nbsp;Allison Toltz,&nbsp;Colin Baker,&nbsp;Sam Beddar,&nbsp;Christian Graeff,&nbsp;Charles-Antoine Collins Fekete","doi":"10.1002/mp.17645","DOIUrl":"10.1002/mp.17645","url":null,"abstract":"&lt;div&gt;\u0000 \u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Background&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;Incorporating image guidance into ion beam therapy is critical for minimizing beam range uncertainties and realizing the modality's potential. One promising avenue for image guidance is to capture transmission ion radiographs (iRads) before and/or during treatment. iRad image quality is typically maximized using a single-event imaging system, which involves tracking individual ions, albeit the approach is generally not suited to clinical beam settings. An alternative faster and clinically compatible method is integrated mode imaging, where individual pencil beam data is acquired, rather than single ion data. To evaluate the usefulness of transmission ion imaging for image guidance, it is crucial to evaluate the image quality of integrated mode iRad systems.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Purpose&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;We report extensive image quality metrics of integrated mode carbon ion radiographs (cRads) and compare them with proton radiographs (pRads).&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Methods&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;iRads were obtained at the Marburg Ion Beam Therapy Center using a plastic volumetric scintillator equipped with CCD cameras. The detector captures orthogonal views of the 3D energy deposition in the scintillator from individual pencil beams. Four phantoms were scanned using a &lt;span&gt;&lt;/span&gt;&lt;math&gt;\u0000 &lt;semantics&gt;\u0000 &lt;mrow&gt;\u0000 &lt;mn&gt;15&lt;/mn&gt;\u0000 &lt;mo&gt;×&lt;/mo&gt;\u0000 &lt;mn&gt;15&lt;/mn&gt;\u0000 &lt;mspace&gt;&lt;/mspace&gt;\u0000 &lt;msup&gt;\u0000 &lt;mi&gt;cm&lt;/mi&gt;\u0000 &lt;mn&gt;2&lt;/mn&gt;\u0000 &lt;/msup&gt;\u0000 &lt;/mrow&gt;\u0000 &lt;annotation&gt;$15times 15 {rm cm}^2$&lt;/annotation&gt;\u0000 &lt;/semantics&gt;&lt;/math&gt; field of view and a beam spacing of 1 mm. First, 9 tissue-substitute inserts were used to evaluate water equivalent thickness (WET) accuracy. Radiographs of those inserts were reconstructed for beam spacings ranging from 1 to 7 mm to evaluate the impact of spacing on quantitative accuracy. For spatial resolution, custom 3D printed line pair (lp) modules ranging from 0.5 to 10 lp/cm were scanned. To evaluate low contrast detectability, a custom 3D printed low contrast module consisting of 20 holes with depths ranging from 1 to 8 mm and diameters from 1 to 10 mm was scanned. iRads of an anthropomorphic head phantom were also obtained.&lt;/p&gt;\u0000 &lt;/section&gt;\u0000 \u0000 &lt;section&gt;\u0000 \u0000 &lt;h3&gt; Results&lt;/h3&gt;\u0000 \u0000 &lt;p&gt;Spatial resolution and low contrast detectio","PeriodicalId":18384,"journal":{"name":"Medical physics","volume":"52 5","pages":"3097-3106"},"PeriodicalIF":3.2,"publicationDate":"2025-02-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/mp.17645","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"143191693","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
相关产品
×
本文献相关产品
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信