{"title":"利用有限元分析评估钛和碳纤维增强聚醚醚酮迷你钢板在 Le Fort I 型推进截骨固定中的生物力学效应和稳定性。","authors":"Kübra Yakut, Sezai Çiftçi","doi":"10.1016/j.jormas.2024.102153","DOIUrl":null,"url":null,"abstract":"<div><h3>Aim</h3><div>This study aimed to investigate the biomechanical properties of 60 % carbon fiber-reinforced polyetheretherketone (Cfr-PEEK), which exhibits high mechanical strength and can address the limitations of titanium mini plates used in Le Fort I osteotomy.</div></div><div><h3>Material and Method</h3><div>Models were created using the FEA method based on tomography images of adult individuals. A 5 mm maxillary advancement was applied to the models following Le Fort I osteotomy. Mini plates made of titanium and 60 % Cfr-PEEK were used. Support was provided by the nasomaxillary and zygomaticomaxillary buttresses to fix a total of four l-shaped mini plates. Oblique loads of 125 N, directed from palatal to buccal, and a total of 250 N compression loads were applied to the central fossa of the premolar and molar teeth in the maxillary model at a 30° angle relative to the long axis of the teeth. Displacement values at the osteotomy line, Von Mises stresses in the mini plate-screws, and principal stresses in the bone were compared.</div></div><div><h3>Results</h3><div>Examination of stress values in the fixation systems of the models revealed higher stress values in the Cfr-PEEK model compared to the titanium model. However, these stresses did not reach levels that would deform the Cfr-PEEK fixation systems. Stress and displacement values in the bone were lower in the Cfr-PEEK model compared to the titanium model.</div></div><div><h3>Conclusion</h3><div>According to the findings of our study, Cfr-PEEK represents a viable alternative to titanium for mini plate material in Le Fort I osteotomy, offering biomechanical advantages.</div></div>","PeriodicalId":55993,"journal":{"name":"Journal of Stomatology Oral and Maxillofacial Surgery","volume":"126 6","pages":"Article 102153"},"PeriodicalIF":1.8000,"publicationDate":"2024-11-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Evaluation of the biomechanical effects and stability of titanium and carbon fiber-reinforced polyetheretherketone mini plates in Le Fort I advancement osteotomy fixation using finite element analysis\",\"authors\":\"Kübra Yakut, Sezai Çiftçi\",\"doi\":\"10.1016/j.jormas.2024.102153\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Aim</h3><div>This study aimed to investigate the biomechanical properties of 60 % carbon fiber-reinforced polyetheretherketone (Cfr-PEEK), which exhibits high mechanical strength and can address the limitations of titanium mini plates used in Le Fort I osteotomy.</div></div><div><h3>Material and Method</h3><div>Models were created using the FEA method based on tomography images of adult individuals. A 5 mm maxillary advancement was applied to the models following Le Fort I osteotomy. Mini plates made of titanium and 60 % Cfr-PEEK were used. Support was provided by the nasomaxillary and zygomaticomaxillary buttresses to fix a total of four l-shaped mini plates. Oblique loads of 125 N, directed from palatal to buccal, and a total of 250 N compression loads were applied to the central fossa of the premolar and molar teeth in the maxillary model at a 30° angle relative to the long axis of the teeth. Displacement values at the osteotomy line, Von Mises stresses in the mini plate-screws, and principal stresses in the bone were compared.</div></div><div><h3>Results</h3><div>Examination of stress values in the fixation systems of the models revealed higher stress values in the Cfr-PEEK model compared to the titanium model. However, these stresses did not reach levels that would deform the Cfr-PEEK fixation systems. Stress and displacement values in the bone were lower in the Cfr-PEEK model compared to the titanium model.</div></div><div><h3>Conclusion</h3><div>According to the findings of our study, Cfr-PEEK represents a viable alternative to titanium for mini plate material in Le Fort I osteotomy, offering biomechanical advantages.</div></div>\",\"PeriodicalId\":55993,\"journal\":{\"name\":\"Journal of Stomatology Oral and Maxillofacial Surgery\",\"volume\":\"126 6\",\"pages\":\"Article 102153\"},\"PeriodicalIF\":1.8000,\"publicationDate\":\"2024-11-15\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Stomatology Oral and Maxillofacial Surgery\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2468785524004427\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"DENTISTRY, ORAL SURGERY & MEDICINE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Stomatology Oral and Maxillofacial Surgery","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2468785524004427","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0
摘要
目的:本研究旨在探讨60%碳纤维增强聚醚醚酮(Cfr-PEEK)的生物力学特性,该材料具有较高的机械强度,可解决勒堡I型截骨术中使用的钛迷你板的局限性:根据成人的断层扫描图像,使用有限元分析法创建模型。在 Le Fort I 型截骨术后,对模型进行 5 毫米的上颌骨前移。使用了由钛和 60% Cfr-PEEK 制成的迷你板。鼻颌托和颧颌托提供了支撑,共固定了四块 L 形迷你板。从腭侧到颊侧施加 125 牛顿的斜向载荷,并在上颌模型的前磨牙和磨牙中心窝施加总计 250 牛顿的压缩载荷,与牙齿长轴成 30° 角。比较了截骨线处的位移值、微型板状螺钉中的 Von Mises 应力和骨中的主应力:结果:对模型固定系统应力值的检查显示,Cfr-PEEK 模型的应力值高于钛模型。然而,这些应力并未达到导致 Cfr-PEEK 固定系统变形的水平。与钛模型相比,Cfr-PEEK 模型中骨骼的应力和位移值较低:根据我们的研究结果,在 Le Fort I 截骨术中,Cfr-PEEK 是钛迷你钢板材料的可行替代品,具有生物力学优势。
Evaluation of the biomechanical effects and stability of titanium and carbon fiber-reinforced polyetheretherketone mini plates in Le Fort I advancement osteotomy fixation using finite element analysis
Aim
This study aimed to investigate the biomechanical properties of 60 % carbon fiber-reinforced polyetheretherketone (Cfr-PEEK), which exhibits high mechanical strength and can address the limitations of titanium mini plates used in Le Fort I osteotomy.
Material and Method
Models were created using the FEA method based on tomography images of adult individuals. A 5 mm maxillary advancement was applied to the models following Le Fort I osteotomy. Mini plates made of titanium and 60 % Cfr-PEEK were used. Support was provided by the nasomaxillary and zygomaticomaxillary buttresses to fix a total of four l-shaped mini plates. Oblique loads of 125 N, directed from palatal to buccal, and a total of 250 N compression loads were applied to the central fossa of the premolar and molar teeth in the maxillary model at a 30° angle relative to the long axis of the teeth. Displacement values at the osteotomy line, Von Mises stresses in the mini plate-screws, and principal stresses in the bone were compared.
Results
Examination of stress values in the fixation systems of the models revealed higher stress values in the Cfr-PEEK model compared to the titanium model. However, these stresses did not reach levels that would deform the Cfr-PEEK fixation systems. Stress and displacement values in the bone were lower in the Cfr-PEEK model compared to the titanium model.
Conclusion
According to the findings of our study, Cfr-PEEK represents a viable alternative to titanium for mini plate material in Le Fort I osteotomy, offering biomechanical advantages.