在颈动脉支架植入术中防止远端滤器型栓塞保护装置被卡住的重要注意事项。

IF 2.4 4区 医学 Q2 CLINICAL NEUROLOGY
Neurologia medico-chirurgica Pub Date : 2024-10-15 Epub Date: 2024-08-28 DOI:10.2176/jns-nmc.2024-0013
Juntaro Fujita, Nobuyuki Hirotsune, Yuki Matsuda, Osamu Hamasaki, Shoichi Fukuda, Yoshihiro Okada, Aiko Kobashi, Shohei Nishigaki, Keigo Makino, Yusuke Tomita, Naoya Kidani, Kenichiro Muraoka, Shigeki Nishino
{"title":"在颈动脉支架植入术中防止远端滤器型栓塞保护装置被卡住的重要注意事项。","authors":"Juntaro Fujita, Nobuyuki Hirotsune, Yuki Matsuda, Osamu Hamasaki, Shoichi Fukuda, Yoshihiro Okada, Aiko Kobashi, Shohei Nishigaki, Keigo Makino, Yusuke Tomita, Naoya Kidani, Kenichiro Muraoka, Shigeki Nishino","doi":"10.2176/jns-nmc.2024-0013","DOIUrl":null,"url":null,"abstract":"<p><p>Failure to retrieve a distal filter-based embolic protection device (EPD) is a potential complication of carotid artery stenting. This may be caused by trapping of the proximal marker of the EPD within the stent tip marker. Maintaining an adequate distance between the two can prevent this. We examined the behavior of several stent-filter-based EPD combinations, focusing on their propensity to become trapped or disengage in vitro. Four physicians subjectively rated the force required to result in trapping using a 5-point scale. Moreover, the force required to disengage trapped devices was evaluated. The Casper stent-Spider FX EPD combination was difficult to disengage when entrapment occurred, which suggested that this phenomenon tended to occur with this combination. The stent tip marker of the closed-cell stents advanced as they shortened, which may be a unique feature of closed-cell stents. Although trapping is uncommon, it can cause serious complications. To prevent these complications, device characteristics should be well understood before they are used in patients.</p>","PeriodicalId":19225,"journal":{"name":"Neurologia medico-chirurgica","volume":" ","pages":"382-386"},"PeriodicalIF":2.4000,"publicationDate":"2024-10-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11538804/pdf/","citationCount":"0","resultStr":"{\"title\":\"Important Notes for Preventing Entrapment of Distal Filter-based Embolic Protection Device in Carotid Artery Stenting.\",\"authors\":\"Juntaro Fujita, Nobuyuki Hirotsune, Yuki Matsuda, Osamu Hamasaki, Shoichi Fukuda, Yoshihiro Okada, Aiko Kobashi, Shohei Nishigaki, Keigo Makino, Yusuke Tomita, Naoya Kidani, Kenichiro Muraoka, Shigeki Nishino\",\"doi\":\"10.2176/jns-nmc.2024-0013\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Failure to retrieve a distal filter-based embolic protection device (EPD) is a potential complication of carotid artery stenting. This may be caused by trapping of the proximal marker of the EPD within the stent tip marker. Maintaining an adequate distance between the two can prevent this. We examined the behavior of several stent-filter-based EPD combinations, focusing on their propensity to become trapped or disengage in vitro. Four physicians subjectively rated the force required to result in trapping using a 5-point scale. Moreover, the force required to disengage trapped devices was evaluated. The Casper stent-Spider FX EPD combination was difficult to disengage when entrapment occurred, which suggested that this phenomenon tended to occur with this combination. The stent tip marker of the closed-cell stents advanced as they shortened, which may be a unique feature of closed-cell stents. Although trapping is uncommon, it can cause serious complications. To prevent these complications, device characteristics should be well understood before they are used in patients.</p>\",\"PeriodicalId\":19225,\"journal\":{\"name\":\"Neurologia medico-chirurgica\",\"volume\":\" \",\"pages\":\"382-386\"},\"PeriodicalIF\":2.4000,\"publicationDate\":\"2024-10-15\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11538804/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Neurologia medico-chirurgica\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.2176/jns-nmc.2024-0013\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/8/28 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q2\",\"JCRName\":\"CLINICAL NEUROLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Neurologia medico-chirurgica","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.2176/jns-nmc.2024-0013","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/8/28 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"CLINICAL NEUROLOGY","Score":null,"Total":0}
引用次数: 0

摘要

无法取回远端过滤式栓塞保护装置(EPD)是颈动脉支架术的一个潜在并发症。其原因可能是 EPD 的近端标记被支架顶端标记卡住。在两者之间保持足够的距离可以避免这种情况的发生。我们研究了几种基于支架-过滤器的 EPD 组合的行为,重点是它们在体外被卡住或脱离的倾向。四位医生使用 5 点量表主观评定了导致卡住所需的力。此外,还对脱开被困装置所需的力进行了评估。卡斯珀支架-蜘蛛 FX EPD 组合在发生夹持时很难脱离,这表明这种现象往往发生在这种组合上。闭孔支架的支架尖端标记随着缩短而前进,这可能是闭孔支架的独特之处。虽然诱捕现象并不常见,但它可能导致严重的并发症。为防止这些并发症的发生,在将设备用于患者之前,应充分了解设备的特性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Important Notes for Preventing Entrapment of Distal Filter-based Embolic Protection Device in Carotid Artery Stenting.

Failure to retrieve a distal filter-based embolic protection device (EPD) is a potential complication of carotid artery stenting. This may be caused by trapping of the proximal marker of the EPD within the stent tip marker. Maintaining an adequate distance between the two can prevent this. We examined the behavior of several stent-filter-based EPD combinations, focusing on their propensity to become trapped or disengage in vitro. Four physicians subjectively rated the force required to result in trapping using a 5-point scale. Moreover, the force required to disengage trapped devices was evaluated. The Casper stent-Spider FX EPD combination was difficult to disengage when entrapment occurred, which suggested that this phenomenon tended to occur with this combination. The stent tip marker of the closed-cell stents advanced as they shortened, which may be a unique feature of closed-cell stents. Although trapping is uncommon, it can cause serious complications. To prevent these complications, device characteristics should be well understood before they are used in patients.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Neurologia medico-chirurgica
Neurologia medico-chirurgica 医学-临床神经学
CiteScore
3.70
自引率
10.50%
发文量
63
审稿时长
3-8 weeks
期刊介绍: Information not localized
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信