{"title":"长髓内钉与短髓内钉治疗反向转子前骨折:生物力学研究","authors":"Gilles Udin , Lore Hoffmann , Fabio Becce , Olivier Borens , Alexandre Terrier","doi":"10.1016/j.medengphy.2024.104230","DOIUrl":null,"url":null,"abstract":"<div><p>There is currently no definitive evidence for the implant of choice for the treatment of reverse pertrochanteric fractures. Here, we aimed to compare the stability provided by two implant options: long and short intramedullary nails.</p><p>We performed finite element simulations of different patterns of reverse pertrochanteric fractures with varying bone quality, and compared the short vs long nail stabilization under physiological loads. For each variable combination, the micromotions at the fracture site, bone strain, and implant stress were computed.</p><p>Mean micromotions at the fracture surface and absolute and relative fracture surface with micromotions >150 µm were slightly lower with the short nail (8%, 3%, and 3%, respectively). The distal fracture extension negatively affected the stability, with increasing micromotions on the medial side. Bone strain above 1 % was not affected by the nail length. Fatigue stresses were similar for both implants, and no volume was found above the yield and ultimate stress in the tested conditions.</p><p>This simulation study shows no benefit of long nails for the investigated patterns of reverse pertrochanteric fractures, with similar micromotions at the fracture site, bone strain, and implant stress.</p></div>","PeriodicalId":49836,"journal":{"name":"Medical Engineering & Physics","volume":"131 ","pages":"Article 104230"},"PeriodicalIF":1.7000,"publicationDate":"2024-08-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S1350453324001310/pdfft?md5=1fe9fee30f4491794c3f79c37c757e0b&pid=1-s2.0-S1350453324001310-main.pdf","citationCount":"0","resultStr":"{\"title\":\"Long vs short intramedullary nails for reverse pertrochanteric fractures: A biomechanical study\",\"authors\":\"Gilles Udin , Lore Hoffmann , Fabio Becce , Olivier Borens , Alexandre Terrier\",\"doi\":\"10.1016/j.medengphy.2024.104230\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><p>There is currently no definitive evidence for the implant of choice for the treatment of reverse pertrochanteric fractures. Here, we aimed to compare the stability provided by two implant options: long and short intramedullary nails.</p><p>We performed finite element simulations of different patterns of reverse pertrochanteric fractures with varying bone quality, and compared the short vs long nail stabilization under physiological loads. For each variable combination, the micromotions at the fracture site, bone strain, and implant stress were computed.</p><p>Mean micromotions at the fracture surface and absolute and relative fracture surface with micromotions >150 µm were slightly lower with the short nail (8%, 3%, and 3%, respectively). The distal fracture extension negatively affected the stability, with increasing micromotions on the medial side. Bone strain above 1 % was not affected by the nail length. Fatigue stresses were similar for both implants, and no volume was found above the yield and ultimate stress in the tested conditions.</p><p>This simulation study shows no benefit of long nails for the investigated patterns of reverse pertrochanteric fractures, with similar micromotions at the fracture site, bone strain, and implant stress.</p></div>\",\"PeriodicalId\":49836,\"journal\":{\"name\":\"Medical Engineering & Physics\",\"volume\":\"131 \",\"pages\":\"Article 104230\"},\"PeriodicalIF\":1.7000,\"publicationDate\":\"2024-08-13\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S1350453324001310/pdfft?md5=1fe9fee30f4491794c3f79c37c757e0b&pid=1-s2.0-S1350453324001310-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Medical Engineering & Physics\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S1350453324001310\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"ENGINEERING, BIOMEDICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical Engineering & Physics","FirstCategoryId":"5","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1350453324001310","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
Long vs short intramedullary nails for reverse pertrochanteric fractures: A biomechanical study
There is currently no definitive evidence for the implant of choice for the treatment of reverse pertrochanteric fractures. Here, we aimed to compare the stability provided by two implant options: long and short intramedullary nails.
We performed finite element simulations of different patterns of reverse pertrochanteric fractures with varying bone quality, and compared the short vs long nail stabilization under physiological loads. For each variable combination, the micromotions at the fracture site, bone strain, and implant stress were computed.
Mean micromotions at the fracture surface and absolute and relative fracture surface with micromotions >150 µm were slightly lower with the short nail (8%, 3%, and 3%, respectively). The distal fracture extension negatively affected the stability, with increasing micromotions on the medial side. Bone strain above 1 % was not affected by the nail length. Fatigue stresses were similar for both implants, and no volume was found above the yield and ultimate stress in the tested conditions.
This simulation study shows no benefit of long nails for the investigated patterns of reverse pertrochanteric fractures, with similar micromotions at the fracture site, bone strain, and implant stress.
期刊介绍:
Medical Engineering & Physics provides a forum for the publication of the latest developments in biomedical engineering, and reflects the essential multidisciplinary nature of the subject. The journal publishes in-depth critical reviews, scientific papers and technical notes. Our focus encompasses the application of the basic principles of physics and engineering to the development of medical devices and technology, with the ultimate aim of producing improvements in the quality of health care.Topics covered include biomechanics, biomaterials, mechanobiology, rehabilitation engineering, biomedical signal processing and medical device development. Medical Engineering & Physics aims to keep both engineers and clinicians abreast of the latest applications of technology to health care.