用于肩部肌肉骨骼研究应用的低剂量 CT 扫描的几何精度

IF 1.7 4区 医学 Q3 ENGINEERING, BIOMEDICAL
Rebekah L. Lawrence , Lydia Nicholson , Erin C.S. Lee , Kelby Napier , Benjamin Zmistowski , Michael J. Rainbow
{"title":"用于肩部肌肉骨骼研究应用的低剂量 CT 扫描的几何精度","authors":"Rebekah L. Lawrence ,&nbsp;Lydia Nicholson ,&nbsp;Erin C.S. Lee ,&nbsp;Kelby Napier ,&nbsp;Benjamin Zmistowski ,&nbsp;Michael J. Rainbow","doi":"10.1016/j.medengphy.2024.104214","DOIUrl":null,"url":null,"abstract":"<div><p>Computed tomography (CT) imaging is frequently employed in a variety of musculoskeletal research applications. Although research studies often use imaging protocols developed for clinical applications, lower dose protocols are likely possible when the goal is to reconstruct 3D bone models. Our purpose was to describe the dose-accuracy trade-off between incrementally lower-dose CT scans and the geometric reconstruction accuracy of the humerus, scapula, and clavicle. Six shoulder specimens were acquired and scanned using 5 helical CT protocols: 1) 120 kVp, 450 mA (full-dose); 2) 120 kVp, 120 mA; 3) 120 kVp, 100 mA; 4) 100 kVp, 100 mA; 5) 80 kVp, 80 mA. Scans were segmented and reconstructed into 3D surface meshes. Geometric error was assessed by comparing the surfaces of the low-dose meshes to the full-dose (gold standard) mesh and was described using mean absolute error, bias, precision, and worst-case error. All low-dose protocols resulted in a &gt;70 % reduction in the effective dose. Lower dose scans resulted in higher geometric errors; however, error magnitudes were generally &lt;0.5 mm. These data suggest that the effective dose associated with CT imaging can be substantially reduced without a significant loss of geometric reconstruction accuracy.</p></div>","PeriodicalId":49836,"journal":{"name":"Medical Engineering & Physics","volume":null,"pages":null},"PeriodicalIF":1.7000,"publicationDate":"2024-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S1350453324001152/pdfft?md5=7cf2c5434df8a9bcf734627a4144f509&pid=1-s2.0-S1350453324001152-main.pdf","citationCount":"0","resultStr":"{\"title\":\"Geometric accuracy of low-dose CT scans for use in shoulder musculoskeletal research applications\",\"authors\":\"Rebekah L. Lawrence ,&nbsp;Lydia Nicholson ,&nbsp;Erin C.S. Lee ,&nbsp;Kelby Napier ,&nbsp;Benjamin Zmistowski ,&nbsp;Michael J. Rainbow\",\"doi\":\"10.1016/j.medengphy.2024.104214\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><p>Computed tomography (CT) imaging is frequently employed in a variety of musculoskeletal research applications. Although research studies often use imaging protocols developed for clinical applications, lower dose protocols are likely possible when the goal is to reconstruct 3D bone models. Our purpose was to describe the dose-accuracy trade-off between incrementally lower-dose CT scans and the geometric reconstruction accuracy of the humerus, scapula, and clavicle. Six shoulder specimens were acquired and scanned using 5 helical CT protocols: 1) 120 kVp, 450 mA (full-dose); 2) 120 kVp, 120 mA; 3) 120 kVp, 100 mA; 4) 100 kVp, 100 mA; 5) 80 kVp, 80 mA. Scans were segmented and reconstructed into 3D surface meshes. Geometric error was assessed by comparing the surfaces of the low-dose meshes to the full-dose (gold standard) mesh and was described using mean absolute error, bias, precision, and worst-case error. All low-dose protocols resulted in a &gt;70 % reduction in the effective dose. Lower dose scans resulted in higher geometric errors; however, error magnitudes were generally &lt;0.5 mm. These data suggest that the effective dose associated with CT imaging can be substantially reduced without a significant loss of geometric reconstruction accuracy.</p></div>\",\"PeriodicalId\":49836,\"journal\":{\"name\":\"Medical Engineering & Physics\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.7000,\"publicationDate\":\"2024-08-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S1350453324001152/pdfft?md5=7cf2c5434df8a9bcf734627a4144f509&pid=1-s2.0-S1350453324001152-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Medical Engineering & Physics\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S1350453324001152\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"ENGINEERING, BIOMEDICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical Engineering & Physics","FirstCategoryId":"5","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1350453324001152","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

摘要

计算机断层扫描(CT)成像经常被用于各种肌肉骨骼研究应用中。虽然研究通常使用为临床应用而开发的成像方案,但当目标是重建三维骨骼模型时,低剂量方案也是可能的。我们的目的是描述剂量逐渐降低的 CT 扫描与肱骨、肩胛骨和锁骨几何重建精度之间的剂量-精度权衡。采用 5 种螺旋 CT 方案采集并扫描了 6 个肩部标本:1) 120 kVp,450 mA(全剂量);2) 120 kVp,120 mA;3) 120 kVp,100 mA;4) 100 kVp,100 mA;5) 80 kVp,80 mA。扫描被分割并重建为三维表面网格。通过比较低剂量网格与全剂量(金标准)网格的表面来评估几何误差,并使用平均绝对误差、偏差、精确度和最坏情况误差进行描述。所有低剂量方案都使有效剂量减少了 70%。低剂量扫描导致较高的几何误差,但误差幅度一般为 0.5 毫米。这些数据表明,CT 成像的有效剂量可以大大降低,而几何重建的准确性不会有明显的损失。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Geometric accuracy of low-dose CT scans for use in shoulder musculoskeletal research applications

Computed tomography (CT) imaging is frequently employed in a variety of musculoskeletal research applications. Although research studies often use imaging protocols developed for clinical applications, lower dose protocols are likely possible when the goal is to reconstruct 3D bone models. Our purpose was to describe the dose-accuracy trade-off between incrementally lower-dose CT scans and the geometric reconstruction accuracy of the humerus, scapula, and clavicle. Six shoulder specimens were acquired and scanned using 5 helical CT protocols: 1) 120 kVp, 450 mA (full-dose); 2) 120 kVp, 120 mA; 3) 120 kVp, 100 mA; 4) 100 kVp, 100 mA; 5) 80 kVp, 80 mA. Scans were segmented and reconstructed into 3D surface meshes. Geometric error was assessed by comparing the surfaces of the low-dose meshes to the full-dose (gold standard) mesh and was described using mean absolute error, bias, precision, and worst-case error. All low-dose protocols resulted in a >70 % reduction in the effective dose. Lower dose scans resulted in higher geometric errors; however, error magnitudes were generally <0.5 mm. These data suggest that the effective dose associated with CT imaging can be substantially reduced without a significant loss of geometric reconstruction accuracy.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Medical Engineering & Physics
Medical Engineering & Physics 工程技术-工程:生物医学
CiteScore
4.30
自引率
4.50%
发文量
172
审稿时长
3.0 months
期刊介绍: Medical Engineering & Physics provides a forum for the publication of the latest developments in biomedical engineering, and reflects the essential multidisciplinary nature of the subject. The journal publishes in-depth critical reviews, scientific papers and technical notes. Our focus encompasses the application of the basic principles of physics and engineering to the development of medical devices and technology, with the ultimate aim of producing improvements in the quality of health care.Topics covered include biomechanics, biomaterials, mechanobiology, rehabilitation engineering, biomedical signal processing and medical device development. Medical Engineering & Physics aims to keep both engineers and clinicians abreast of the latest applications of technology to health care.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信