{"title":"使用PXI的定时和触发构建测试应用程序","authors":"L. Duraiappah","doi":"10.1109/AUTEST.2002.1047894","DOIUrl":null,"url":null,"abstract":"The PXI platform provides sophisticated synchronization capabilities for automated test applications. This paper provides a technical overview of triggering and synchronization capabilities of PXI, including the PXI Trigger Bus, the Star Trigger Bus, and the back-plane clock. The test applications of triggering and synchronization are discussed and demonstrated, including: synchronization of multiple digitizers/scopes to simultaneously sample multiple channels; synchronization of a scope (or multiple scopes) to a source to accomplish time domain measurements of the DUT response to stimulus; position based measurements (digitizers synchronized to counter/timers); synchronization of a DMM to a switch to access multiple test nodes when the DUT has settled to a steady state; and synchronization of motion control with image acquisition.","PeriodicalId":372875,"journal":{"name":"Proceedings, IEEE AUTOTESTCON","volume":"10 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2002-12-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"3","resultStr":"{\"title\":\"Building test applications using timing and triggering with PXI\",\"authors\":\"L. Duraiappah\",\"doi\":\"10.1109/AUTEST.2002.1047894\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The PXI platform provides sophisticated synchronization capabilities for automated test applications. This paper provides a technical overview of triggering and synchronization capabilities of PXI, including the PXI Trigger Bus, the Star Trigger Bus, and the back-plane clock. The test applications of triggering and synchronization are discussed and demonstrated, including: synchronization of multiple digitizers/scopes to simultaneously sample multiple channels; synchronization of a scope (or multiple scopes) to a source to accomplish time domain measurements of the DUT response to stimulus; position based measurements (digitizers synchronized to counter/timers); synchronization of a DMM to a switch to access multiple test nodes when the DUT has settled to a steady state; and synchronization of motion control with image acquisition.\",\"PeriodicalId\":372875,\"journal\":{\"name\":\"Proceedings, IEEE AUTOTESTCON\",\"volume\":\"10 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2002-12-10\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"3\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Proceedings, IEEE AUTOTESTCON\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/AUTEST.2002.1047894\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Proceedings, IEEE AUTOTESTCON","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/AUTEST.2002.1047894","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Building test applications using timing and triggering with PXI
The PXI platform provides sophisticated synchronization capabilities for automated test applications. This paper provides a technical overview of triggering and synchronization capabilities of PXI, including the PXI Trigger Bus, the Star Trigger Bus, and the back-plane clock. The test applications of triggering and synchronization are discussed and demonstrated, including: synchronization of multiple digitizers/scopes to simultaneously sample multiple channels; synchronization of a scope (or multiple scopes) to a source to accomplish time domain measurements of the DUT response to stimulus; position based measurements (digitizers synchronized to counter/timers); synchronization of a DMM to a switch to access multiple test nodes when the DUT has settled to a steady state; and synchronization of motion control with image acquisition.