新生儿视网膜闪烁电图的研究。

IF 2.6 4区 医学 Q2 OPHTHALMOLOGY
James V M Hanson, Caroline Weber, Oliver A Pfäffli, Dirk Bassler, Daphne L McCulloch, Christina Gerth-Kahlert
{"title":"新生儿视网膜闪烁电图的研究。","authors":"James V M Hanson,&nbsp;Caroline Weber,&nbsp;Oliver A Pfäffli,&nbsp;Dirk Bassler,&nbsp;Daphne L McCulloch,&nbsp;Christina Gerth-Kahlert","doi":"10.1007/s10633-022-09889-5","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To develop and validate a flicker electroretinogram (ERG) protocol in term-born neonates as a potential tool for assessing preterm infants at risk of developing retinopathy of prematurity.</p><p><strong>Methods: </strong>A custom flicker ERG protocol was developed for use with the hand-held RETeval® electrophysiology device. Feasibility of measuring flicker ERG through closed eyelids and without mydriasis was established in a pilot study enabling optimisation of the test protocol. Following this, healthy term-born neonates (gestational age 37-42 weeks) were recruited at the Neonatology clinic of the University Hospital Zurich. Flicker ERG recordings were performed using proprietary disposable skin electrodes during the first four days of life when the infants were sleeping. Flicker stimuli were presented at 28.3 Hz for a stimulus series at 3, 6, 12, 30, and 50 cd·s/m<sup>2</sup>, with two measurements at each stimulus level. Results were analysed offline. Flicker ERG peak times and amplitudes were derived from the averaged measurements per stimulus level for each subject.</p><p><strong>Results: </strong>28 term-born neonates were included in the analysis. All infants tolerated the testing procedure well. Flicker ERG recording was achieved in all subjects with reproducible flicker ERG waveforms for 30 and 50 cd·s/m<sup>2</sup> stimuli. Reproducible ERGs were recorded in the majority of infants for the weaker stimuli (with detectable ERGs in 20/28, 25/28, and 27/28 at 3, 6, and 12 cd·s/m<sup>2</sup>, respectively). Flicker ERG amplitudes increased with increasing stimulus strength, with peak times concurrently decreasing slightly.</p><p><strong>Conclusion: </strong>Flicker ERG recording is feasible and reliably recorded in sleeping neonates through closed eyelids using skin electrodes and without mydriasis. Flicker ERG amplitude decreases for lower luminance flicker but remains detectable for 3 cd·s/m<sup>2</sup> flicker in the majority of healthy term-born neonates. These data provide a basis to study retinal function in premature infants using this protocol.</p>","PeriodicalId":11207,"journal":{"name":"Documenta Ophthalmologica","volume":"145 3","pages":"175-184"},"PeriodicalIF":2.6000,"publicationDate":"2022-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9653345/pdf/","citationCount":"1","resultStr":"{\"title\":\"Flicker electroretinogram in newborn infants.\",\"authors\":\"James V M Hanson,&nbsp;Caroline Weber,&nbsp;Oliver A Pfäffli,&nbsp;Dirk Bassler,&nbsp;Daphne L McCulloch,&nbsp;Christina Gerth-Kahlert\",\"doi\":\"10.1007/s10633-022-09889-5\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>To develop and validate a flicker electroretinogram (ERG) protocol in term-born neonates as a potential tool for assessing preterm infants at risk of developing retinopathy of prematurity.</p><p><strong>Methods: </strong>A custom flicker ERG protocol was developed for use with the hand-held RETeval® electrophysiology device. Feasibility of measuring flicker ERG through closed eyelids and without mydriasis was established in a pilot study enabling optimisation of the test protocol. Following this, healthy term-born neonates (gestational age 37-42 weeks) were recruited at the Neonatology clinic of the University Hospital Zurich. Flicker ERG recordings were performed using proprietary disposable skin electrodes during the first four days of life when the infants were sleeping. Flicker stimuli were presented at 28.3 Hz for a stimulus series at 3, 6, 12, 30, and 50 cd·s/m<sup>2</sup>, with two measurements at each stimulus level. Results were analysed offline. Flicker ERG peak times and amplitudes were derived from the averaged measurements per stimulus level for each subject.</p><p><strong>Results: </strong>28 term-born neonates were included in the analysis. All infants tolerated the testing procedure well. Flicker ERG recording was achieved in all subjects with reproducible flicker ERG waveforms for 30 and 50 cd·s/m<sup>2</sup> stimuli. Reproducible ERGs were recorded in the majority of infants for the weaker stimuli (with detectable ERGs in 20/28, 25/28, and 27/28 at 3, 6, and 12 cd·s/m<sup>2</sup>, respectively). Flicker ERG amplitudes increased with increasing stimulus strength, with peak times concurrently decreasing slightly.</p><p><strong>Conclusion: </strong>Flicker ERG recording is feasible and reliably recorded in sleeping neonates through closed eyelids using skin electrodes and without mydriasis. Flicker ERG amplitude decreases for lower luminance flicker but remains detectable for 3 cd·s/m<sup>2</sup> flicker in the majority of healthy term-born neonates. These data provide a basis to study retinal function in premature infants using this protocol.</p>\",\"PeriodicalId\":11207,\"journal\":{\"name\":\"Documenta Ophthalmologica\",\"volume\":\"145 3\",\"pages\":\"175-184\"},\"PeriodicalIF\":2.6000,\"publicationDate\":\"2022-12-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9653345/pdf/\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Documenta Ophthalmologica\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1007/s10633-022-09889-5\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"OPHTHALMOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Documenta Ophthalmologica","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s10633-022-09889-5","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 1

摘要

目的:研究并验证一种针对足月新生儿的闪烁视网膜电图(ERG)方案,作为评估早产儿发生早产儿视网膜病变风险的潜在工具。方法:为手持式RETeval®电生理设备开发了一种定制的闪烁ERG协议。在一项试点研究中,通过闭眼和无瞳孔测量闪烁ERG的可行性得到了证实,从而优化了测试方案。在此之后,在苏黎世大学医院新生儿科诊所招募了健康的足月新生儿(胎龄37-42周)。在婴儿出生后的前四天,当他们睡觉时,使用专有的一次性皮肤电极进行闪烁ERG记录。在3、6、12、30和50 cd·s/m2的刺激序列中,以28.3 Hz的频率呈现闪烁刺激,每个刺激水平有两次测量。结果离线分析。闪烁ERG峰值时间和振幅由每个受试者每个刺激水平的平均测量值得出。结果:28例足月新生儿纳入分析。所有婴儿都能很好地耐受测试过程。在30和50 cd·s/m2的刺激下,所有受试者的闪烁ERG波形均可重现。在较弱的刺激下,大多数婴儿记录到可重复的ERGs(分别在3、6和12 cd·s/m2时,20/28、25/28和27/28可检测到ERGs)。闪烁ERG振幅随刺激强度的增加而增加,峰值时间同时略有减少。结论:在无瞳孔的情况下,闭眼皮肤电极记录睡眠新生儿的闪烁ERG是可行和可靠的。闪烁ERG振幅在较低亮度闪烁时减小,但在大多数健康足月新生儿中仍可检测到3 cd·s/m2闪烁。这些数据为使用该方案研究早产儿视网膜功能提供了基础。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

Flicker electroretinogram in newborn infants.

Flicker electroretinogram in newborn infants.

Flicker electroretinogram in newborn infants.

Flicker electroretinogram in newborn infants.

Purpose: To develop and validate a flicker electroretinogram (ERG) protocol in term-born neonates as a potential tool for assessing preterm infants at risk of developing retinopathy of prematurity.

Methods: A custom flicker ERG protocol was developed for use with the hand-held RETeval® electrophysiology device. Feasibility of measuring flicker ERG through closed eyelids and without mydriasis was established in a pilot study enabling optimisation of the test protocol. Following this, healthy term-born neonates (gestational age 37-42 weeks) were recruited at the Neonatology clinic of the University Hospital Zurich. Flicker ERG recordings were performed using proprietary disposable skin electrodes during the first four days of life when the infants were sleeping. Flicker stimuli were presented at 28.3 Hz for a stimulus series at 3, 6, 12, 30, and 50 cd·s/m2, with two measurements at each stimulus level. Results were analysed offline. Flicker ERG peak times and amplitudes were derived from the averaged measurements per stimulus level for each subject.

Results: 28 term-born neonates were included in the analysis. All infants tolerated the testing procedure well. Flicker ERG recording was achieved in all subjects with reproducible flicker ERG waveforms for 30 and 50 cd·s/m2 stimuli. Reproducible ERGs were recorded in the majority of infants for the weaker stimuli (with detectable ERGs in 20/28, 25/28, and 27/28 at 3, 6, and 12 cd·s/m2, respectively). Flicker ERG amplitudes increased with increasing stimulus strength, with peak times concurrently decreasing slightly.

Conclusion: Flicker ERG recording is feasible and reliably recorded in sleeping neonates through closed eyelids using skin electrodes and without mydriasis. Flicker ERG amplitude decreases for lower luminance flicker but remains detectable for 3 cd·s/m2 flicker in the majority of healthy term-born neonates. These data provide a basis to study retinal function in premature infants using this protocol.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Documenta Ophthalmologica
Documenta Ophthalmologica 医学-眼科学
CiteScore
3.50
自引率
21.40%
发文量
46
审稿时长
>12 weeks
期刊介绍: Documenta Ophthalmologica is an official publication of the International Society for Clinical Electrophysiology of Vision. The purpose of the journal is to promote the understanding and application of clinical electrophysiology of vision. Documenta Ophthalmologica will publish reviews, research articles, technical notes, brief reports and case studies which inform the readers about basic and clinical sciences related to visual electrodiagnosis and means to improve diagnosis and clinical management of patients using visual electrophysiology. Studies may involve animals or humans. In either case appropriate care must be taken to follow the Declaration of Helsinki for human subject or appropriate humane standards of animal care (e.g., the ARVO standards on Animal Care and Use).
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信