{"title":"A passive thin-film mechanical device for visual respiratory monitoring during oxygen mask use: development and bench validation.","authors":"Dylan Rowe, Mariam Rowe, Phillip Melksham","doi":"10.1088/1873-4030/aeb4bd","DOIUrl":null,"url":null,"abstract":"<p><p>Continuous respiratory monitoring during transfer from the operating theatre to the post-anaesthesia care unit is frequently unavailable despite this being a period of recognised patient vulnerability. Existing electronic respiratory monitors introduce cost, complexity and set-up that may limit use in fast-paced clinical environments. We developed Flow-Bag, a fully passive mechanical breath-indicating device consisting of a thin heat-sealed high-density polyethylene film bag that attaches externally to an oxygen mask. Preliminary film-thickness screening was undertaken using a programmable breathing simulator (PEMDx Pty Ltd, Australia). A 6<i>µ</i>m (per ply) prototype was assessed across respiratory rates (RRs) of 5-60 breaths min<sup>-1</sup>and tidal volumes of 50-600 ml. Three blinded observers positioned 5 m from the device counted visible inflation-deflation cycles over 60 s intervals, with counts compared against simulator-delivered breaths. Performance was additionally assessed across four commercially available oxygen mask models. In screening, the lowest detected tidal volume fell from 400 ml with 20<i>µ</i>m film to 150 ml with 6<i>µ</i>m film, the thinnest film that could be reproducibly manufactured using the processes available to us. The final prototype produced observable breath indications in 42 of 56 tested conditions (all tidal volumes ⩾150 ml) but not at 50 or 100 ml. Within these detectable conditions, observer counts agreed 100% with simulator-delivered breaths. Performance was maintained across four oxygen mask models. However, sensitivity was reduced by incomplete sealing at either the mask-to-face or device-to-mask interface. Flow-Bag is a fully passive mechanical breath indicator. Under controlled bench conditions using a mechanical breathing simulator, it provided reliable visual indication of simulated breathing across RRs of 5-60 breaths min<sup>-1</sup>and tidal volumes of 150-600 ml. Subject to clinical validation and safety testing, this device may offer a potentially low-cost adjunct alongside standard monitoring.</p>","PeriodicalId":49836,"journal":{"name":"Medical Engineering & Physics","volume":" ","pages":""},"PeriodicalIF":2.4000,"publicationDate":"2026-10-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical Engineering & Physics","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1088/1873-4030/aeb4bd","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0
Abstract
Continuous respiratory monitoring during transfer from the operating theatre to the post-anaesthesia care unit is frequently unavailable despite this being a period of recognised patient vulnerability. Existing electronic respiratory monitors introduce cost, complexity and set-up that may limit use in fast-paced clinical environments. We developed Flow-Bag, a fully passive mechanical breath-indicating device consisting of a thin heat-sealed high-density polyethylene film bag that attaches externally to an oxygen mask. Preliminary film-thickness screening was undertaken using a programmable breathing simulator (PEMDx Pty Ltd, Australia). A 6µm (per ply) prototype was assessed across respiratory rates (RRs) of 5-60 breaths min-1and tidal volumes of 50-600 ml. Three blinded observers positioned 5 m from the device counted visible inflation-deflation cycles over 60 s intervals, with counts compared against simulator-delivered breaths. Performance was additionally assessed across four commercially available oxygen mask models. In screening, the lowest detected tidal volume fell from 400 ml with 20µm film to 150 ml with 6µm film, the thinnest film that could be reproducibly manufactured using the processes available to us. The final prototype produced observable breath indications in 42 of 56 tested conditions (all tidal volumes ⩾150 ml) but not at 50 or 100 ml. Within these detectable conditions, observer counts agreed 100% with simulator-delivered breaths. Performance was maintained across four oxygen mask models. However, sensitivity was reduced by incomplete sealing at either the mask-to-face or device-to-mask interface. Flow-Bag is a fully passive mechanical breath indicator. Under controlled bench conditions using a mechanical breathing simulator, it provided reliable visual indication of simulated breathing across RRs of 5-60 breaths min-1and tidal volumes of 150-600 ml. Subject to clinical validation and safety testing, this device may offer a potentially low-cost adjunct alongside standard monitoring.
期刊介绍:
Medical Engineering & Physics provides a forum for the publication of the latest developments in biomedical engineering, and reflects the essential multidisciplinary nature of the subject. The journal publishes in-depth critical reviews, scientific papers and technical notes. Our focus encompasses the application of the basic principles of physics and engineering to the development of medical devices and technology, with the ultimate aim of producing improvements in the quality of health care.Topics covered include biomechanics, biomaterials, mechanobiology, rehabilitation engineering, biomedical signal processing and medical device development. Medical Engineering & Physics aims to keep both engineers and clinicians abreast of the latest applications of technology to health care.