Preliminary validation of a portable, radiation-free, surface-based system (OmÖS) for sagittal spinal curvature measurement.

IF 2.4 4区 医学 Q3 ENGINEERING, BIOMEDICAL
Osman Uzun, Ramazan Kayacan, Halil Burç
{"title":"Preliminary validation of a portable, radiation-free, surface-based system (OmÖS) for sagittal spinal curvature measurement.","authors":"Osman Uzun, Ramazan Kayacan, Halil Burç","doi":"10.1088/1873-4030/ae8a54","DOIUrl":null,"url":null,"abstract":"<p><p>Radiographic assessment remains the reference standard for measuring sagittal spinal curvatures; however, repeated exposure to ionizing radiation limits its suitability for screening and longitudinal follow-up. This has motivated the development of reliable, radiation-free measurement technologies capable of providing clinically meaningful curvature estimates. This study presents the preliminary validation of a portable, radiation-free, surface-based measurement system (OmÖS) designed for sagittal plane spinal curvature measurement. Thoracic and lumbar curvature angles were measured in 16 volunteers using both lateral spinal radiographs (radiographic Cobb method, RaG) and OmÖS under standardized standing conditions. Curvature angles were calculated using surface-based geometrical parameters derived from palpated spinous process locations. System performance was evaluated through internal calibration by comparing stepper motor-based and linear variable differential transformer (LVDT)-based measurements; validity and agreement analyses against radiographic measurements using Pearson correlation, Bland-Altman analysis, and simple linear regression, as well as test-retest reliability assessment over a two-week interval using intraclass correlation coefficients (ICCs). OmÖS demonstrated a very strong correlation with radiographic measurements for thoracic curvature angles (<i>r</i>= 0.974,<i>p</i>< 0.001) and a strong correlation for lumbar curvature angles (<i>r</i>= 0.852,<i>p</i>< 0.001). Bland-Altman analysis indicated minimal systematic bias for thoracic curvature measurements (0.13°), with limits of agreement within predefined clinically acceptable ranges. For lumbar curvature angles, agreement was characterized by slightly wider limits of agreement, but without evidence of systematic bias. Internal calibration demonstrated excellent agreement between the mechanical and electronic measurement components (<i>r</i>> 0.99). Test-retest reliability analysis demonstrated excellent reproducibility for both regions (ICC = 0.990), with low standard error of measurement (SEM = 0.61°) and minimum detectable change at the 95% confidence level (MDC95 = 1.70°). These findings indicate that OmÖS can provide reliable and reproducible radiation-free measurements of sagittal spinal curvature angles under standardized conditions. Whether the proposed system represents a radiation-free alternative of sagittal spinal curvature assessment for screening needs to be confirmed in future clinical studies involving larger and more diverse populations.</p>","PeriodicalId":49836,"journal":{"name":"Medical Engineering & Physics","volume":" ","pages":""},"PeriodicalIF":2.4000,"publicationDate":"2026-07-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical Engineering & Physics","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1088/1873-4030/ae8a54","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

Abstract

Radiographic assessment remains the reference standard for measuring sagittal spinal curvatures; however, repeated exposure to ionizing radiation limits its suitability for screening and longitudinal follow-up. This has motivated the development of reliable, radiation-free measurement technologies capable of providing clinically meaningful curvature estimates. This study presents the preliminary validation of a portable, radiation-free, surface-based measurement system (OmÖS) designed for sagittal plane spinal curvature measurement. Thoracic and lumbar curvature angles were measured in 16 volunteers using both lateral spinal radiographs (radiographic Cobb method, RaG) and OmÖS under standardized standing conditions. Curvature angles were calculated using surface-based geometrical parameters derived from palpated spinous process locations. System performance was evaluated through internal calibration by comparing stepper motor-based and linear variable differential transformer (LVDT)-based measurements; validity and agreement analyses against radiographic measurements using Pearson correlation, Bland-Altman analysis, and simple linear regression, as well as test-retest reliability assessment over a two-week interval using intraclass correlation coefficients (ICCs). OmÖS demonstrated a very strong correlation with radiographic measurements for thoracic curvature angles (r= 0.974,p< 0.001) and a strong correlation for lumbar curvature angles (r= 0.852,p< 0.001). Bland-Altman analysis indicated minimal systematic bias for thoracic curvature measurements (0.13°), with limits of agreement within predefined clinically acceptable ranges. For lumbar curvature angles, agreement was characterized by slightly wider limits of agreement, but without evidence of systematic bias. Internal calibration demonstrated excellent agreement between the mechanical and electronic measurement components (r> 0.99). Test-retest reliability analysis demonstrated excellent reproducibility for both regions (ICC = 0.990), with low standard error of measurement (SEM = 0.61°) and minimum detectable change at the 95% confidence level (MDC95 = 1.70°). These findings indicate that OmÖS can provide reliable and reproducible radiation-free measurements of sagittal spinal curvature angles under standardized conditions. Whether the proposed system represents a radiation-free alternative of sagittal spinal curvature assessment for screening needs to be confirmed in future clinical studies involving larger and more diverse populations.

初步验证便携式,无辐射,表面为基础的系统(OmÖS)矢状脊柱曲度测量。
放射学评估仍然是测量矢状面脊柱弯曲的参考标准;然而,电离辐射的反复暴露限制了其筛查和纵向随访的适用性。这推动了可靠的、无辐射的测量技术的发展,能够提供有临床意义的曲率估计。本研究提出了一种便携式、无辐射、基于表面的测量系统(OmÖS)的初步验证,该系统设计用于矢状面脊柱曲率测量。在标准化的站立条件下,使用侧位脊柱x线片(放射科布法,RaG)和OmÖS测量16名志愿者的胸椎和腰椎弯曲角。曲率角的计算使用基于触诊棘突位置的表面几何参数。系统性能通过内部校准进行评估,通过比较基于步进电机和线性可变差动变压器(LVDT)的测量有效性和一致性分析,使用Pearson相关、Bland-Altman分析和简单线性回归,以及使用类内相关系数(ICC)在两周间隔内进行重测可靠性评估。OmÖS与胸椎曲度的x线测量结果有很强的相关性(r = 0.974, p < 0.001),与腰椎曲度的x线测量结果有很强的相关性(r = 0.852, p < 0.001)。Bland-Altman分析显示胸曲度测量的系统偏差最小(0.13°),在预先确定的临床可接受范围内具有一定的一致性。对于腰椎弯曲角度,一致性的特点是一致性的范围稍宽,但没有系统性偏差的证据。内部校准表明机械和电子测量元件之间具有良好的一致性(r > 0.99)。测试-重测信度分析显示,两个区域的重现性良好(ICC = 0.990),测量标准误差低(SEM = 0.61°),95%置信水平下的最小可检测变化(MDC95 = 1.70°)。这些发现表明OmÖS可以在标准化条件下提供可靠的、可重复的无辐射脊柱矢状面曲率角测量。所提出的系统是否代表了一种无辐射的矢状面脊柱曲度评估筛查的替代方案,需要在未来涉及更大、更多样化人群的临床研究中得到证实。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Medical Engineering & Physics
Medical Engineering & Physics 工程技术-工程:生物医学
CiteScore
4.30
自引率
4.50%
发文量
172
审稿时长
3.0 months
期刊介绍: Medical Engineering & Physics provides a forum for the publication of the latest developments in biomedical engineering, and reflects the essential multidisciplinary nature of the subject. The journal publishes in-depth critical reviews, scientific papers and technical notes. Our focus encompasses the application of the basic principles of physics and engineering to the development of medical devices and technology, with the ultimate aim of producing improvements in the quality of health care.Topics covered include biomechanics, biomaterials, mechanobiology, rehabilitation engineering, biomedical signal processing and medical device development. Medical Engineering & Physics aims to keep both engineers and clinicians abreast of the latest applications of technology to health care.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书