Sara Marquina-Martín, Francisco J Segura-Calvo, Martín Puzo-Bayod, Marta Jiménez-García, Francisco J Castro-Alonso
{"title":"Impact of pupil dilation on IOL power calculation based on biometric measurements with Heidelberg Anterion.","authors":"Sara Marquina-Martín, Francisco J Segura-Calvo, Martín Puzo-Bayod, Marta Jiménez-García, Francisco J Castro-Alonso","doi":"10.1007/s10633-026-10117-7","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate the impact of pharmacologic pupil dilation (PD) on biometric measurements and intraocular lens (IOL) power calculation using different generation formulas in patients undergoing cataract surgery.</p><p><strong>Methods: </strong>This prospective study included 126 eyes from 75 patients. Biometric data were obtained before and after pharmacologic dilation using the Anterion Swept-Source OCT biometer (Heidelberg Engineering). Parameters recorded included axial length (AL), mean keratometry (Km), anterior chamber depth (ACD), lens thickness (LT), and white-to-white (WTW) distance. IOL power was calculated using six formulas: SRK/T, Haigis, Barrett Universal II (BUII), Kane, Hoffer QST, and Okulix. Statistical comparisons of biometric values and predicted IOL power before and after dilation were conducted using mixed-effects models.</p><p><strong>Results: </strong>No significant changes were observed in AL, CCT, LT or Km after dilation. However, ACD, WTW distance and PD showed statistically significant increases (both p < 0.01). IOL power estimations based on SRK/T, Haigis, Hoffer QST, BUII and Kane were not significantly affected after dilation. In contrast, Okulix formulas demonstrated statistically significant variations in calculated IOL power post-dilation (p < 0.0001).</p><p><strong>Conclusions: </strong>Pharmacologic pupil dilation induces changes in ACD and WTW measurements obtained using the Anterion Swept-Source biometer, thereby influencing IOL power estimation in formulas incorporating these parameters. Ray-tracing Okulix formula is more sensitive to dilation, whereas third-generation formulas remain more stable. Consistency in pupil status during biometry acquisition is essential to ensure refractive predictability.</p>","PeriodicalId":11207,"journal":{"name":"Documenta Ophthalmologica","volume":" ","pages":""},"PeriodicalIF":2.0000,"publicationDate":"2026-06-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Documenta Ophthalmologica","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s10633-026-10117-7","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: To evaluate the impact of pharmacologic pupil dilation (PD) on biometric measurements and intraocular lens (IOL) power calculation using different generation formulas in patients undergoing cataract surgery.
Methods: This prospective study included 126 eyes from 75 patients. Biometric data were obtained before and after pharmacologic dilation using the Anterion Swept-Source OCT biometer (Heidelberg Engineering). Parameters recorded included axial length (AL), mean keratometry (Km), anterior chamber depth (ACD), lens thickness (LT), and white-to-white (WTW) distance. IOL power was calculated using six formulas: SRK/T, Haigis, Barrett Universal II (BUII), Kane, Hoffer QST, and Okulix. Statistical comparisons of biometric values and predicted IOL power before and after dilation were conducted using mixed-effects models.
Results: No significant changes were observed in AL, CCT, LT or Km after dilation. However, ACD, WTW distance and PD showed statistically significant increases (both p < 0.01). IOL power estimations based on SRK/T, Haigis, Hoffer QST, BUII and Kane were not significantly affected after dilation. In contrast, Okulix formulas demonstrated statistically significant variations in calculated IOL power post-dilation (p < 0.0001).
Conclusions: Pharmacologic pupil dilation induces changes in ACD and WTW measurements obtained using the Anterion Swept-Source biometer, thereby influencing IOL power estimation in formulas incorporating these parameters. Ray-tracing Okulix formula is more sensitive to dilation, whereas third-generation formulas remain more stable. Consistency in pupil status during biometry acquisition is essential to ensure refractive predictability.
期刊介绍:
Documenta Ophthalmologica is an official publication of the International Society for Clinical Electrophysiology of Vision. The purpose of the journal is to promote the understanding and application of clinical electrophysiology of vision. Documenta Ophthalmologica will publish reviews, research articles, technical notes, brief reports and case studies which inform the readers about basic and clinical sciences related to visual electrodiagnosis and means to improve diagnosis and clinical management of patients using visual electrophysiology. Studies may involve animals or humans. In either case appropriate care must be taken to follow the Declaration of Helsinki for human subject or appropriate humane standards of animal care (e.g., the ARVO standards on Animal Care and Use).