A Pneumatically Actuated Robotic Assistant for MRI-Guided Stereotactic Neurosurgery

IF 4.3 Q2 ENGINEERING, BIOMEDICAL
Gang Li;Atharva Paralikar;Pavel Yarmolenko;Ayush Nankani;Wolfgang Loew;Charles Dumoulin;Kevin Cleary;Chima Oluigbo;Reza Monfaredi
{"title":"A Pneumatically Actuated Robotic Assistant for MRI-Guided Stereotactic Neurosurgery","authors":"Gang Li;Atharva Paralikar;Pavel Yarmolenko;Ayush Nankani;Wolfgang Loew;Charles Dumoulin;Kevin Cleary;Chima Oluigbo;Reza Monfaredi","doi":"10.1109/TMRB.2026.3680983","DOIUrl":null,"url":null,"abstract":"This paper presents the development and cadaveric evaluation of a pneumatically actuated robotic assistant for MRI-guided stereotactic neurosurgery. The robot features two coaxial arcs that provide four degrees of freedom to position and orient a needle within an MRI scanner. It is driven by custom-designed MR-Safe pneumatic stepper motors and powered by in-room compressed medical air, enabling precise interventions under intraoperative MRI guidance. To improve anatomical visualization, two custom MR imaging coils were developed and integrated with the system. A dedicated clinical workflow was also proposed for robot-assisted stereotactic procedures under MRI-guidance. Targeting accuracy was first assessed in free space using an optical tracking system, yielding mean absolute errors of <inline-formula> <tex-math>$0.97 \\pm 0.42$ </tex-math></inline-formula> mm for tip position and <inline-formula> <tex-math>$0.46 \\pm 0.23^{\\circ } $ </tex-math></inline-formula> for needle angle. System-level accuracy was then evaluated in MRI-guided phantom studies, showing errors of <inline-formula> <tex-math>$1.38 \\pm 0.80$ </tex-math></inline-formula> mm in tip position and <inline-formula> <tex-math>$1.08 \\pm 0.67^{\\circ } $ </tex-math></inline-formula> in angle. Finally, a cadaver study validated the clinical workflow, with observed errors of 2.11 mm in position and 3.32° in angle. Quantitative imaging analysis confirmed minimal image degradation, with signal-to-noise ratio (SNR) variation below 12.26% and 2D geometric distortion under 0.20%.","PeriodicalId":73318,"journal":{"name":"IEEE transactions on medical robotics and bionics","volume":"8 2","pages":"906-917"},"PeriodicalIF":4.3000,"publicationDate":"2026-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE transactions on medical robotics and bionics","FirstCategoryId":"1085","ListUrlMain":"https://ieeexplore.ieee.org/document/11475210/","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/4/6 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

Abstract

This paper presents the development and cadaveric evaluation of a pneumatically actuated robotic assistant for MRI-guided stereotactic neurosurgery. The robot features two coaxial arcs that provide four degrees of freedom to position and orient a needle within an MRI scanner. It is driven by custom-designed MR-Safe pneumatic stepper motors and powered by in-room compressed medical air, enabling precise interventions under intraoperative MRI guidance. To improve anatomical visualization, two custom MR imaging coils were developed and integrated with the system. A dedicated clinical workflow was also proposed for robot-assisted stereotactic procedures under MRI-guidance. Targeting accuracy was first assessed in free space using an optical tracking system, yielding mean absolute errors of $0.97 \pm 0.42$ mm for tip position and $0.46 \pm 0.23^{\circ } $ for needle angle. System-level accuracy was then evaluated in MRI-guided phantom studies, showing errors of $1.38 \pm 0.80$ mm in tip position and $1.08 \pm 0.67^{\circ } $ in angle. Finally, a cadaver study validated the clinical workflow, with observed errors of 2.11 mm in position and 3.32° in angle. Quantitative imaging analysis confirmed minimal image degradation, with signal-to-noise ratio (SNR) variation below 12.26% and 2D geometric distortion under 0.20%.
mri引导立体定向神经外科的气动驱动机器人助手
本文介绍了一种用于mri引导立体定向神经外科的气动驱动机器人助手的发展和尸体评估。该机器人具有两个同轴弧,提供四个自由度来定位和定位MRI扫描仪内的针头。它由定制设计的MR-Safe气动步进电机驱动,由室内压缩医用空气驱动,在术中MRI指导下实现精确干预。为了提高解剖可视化,开发了两个定制的磁共振成像线圈并与系统集成。一个专门的临床工作流程也提出了在mri指导下机器人辅助立体定向手术。首先使用光学跟踪系统在自由空间中评估瞄准精度,尖端位置的平均绝对误差为$0.97 \pm 0.42$ mm,针角的平均绝对误差为$0.46 \pm 0.23^{\circ} $。然后在mri引导的幻影研究中评估系统级精度,显示尖端位置误差为1.38 \pm 0.80$ mm,角度误差为1.08 \pm 0.67 $ {\circ} $。最后,一项尸体研究验证了临床工作流程,观察到的位置误差为2.11 mm,角度误差为3.32°。定量成像分析证实图像退化最小,信噪比(SNR)变化小于12.26%,二维几何畸变小于0.20%。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
6.80
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书