{"title":"Real-Time Water Pressure Monitoring in Unilateral Biportal Endoscopic Spine Surgery.","authors":"Dongwoo Yu, Ikchan Jeon, Sang Woo Kim","doi":"10.14245/ns.2550648.324","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Unilateral biportal endoscopic (UBE) spine surgery is a minimally invasive technique that uses continuous irrigation to improve visualization and control bleeding. Effective water pressure management is crucial for patient safety, particularly at the cervical and thoracic levels where spinal cord injury risk is higher. However, real-time pressure monitoring remains underexplored. This study evaluates the impact of real-time water pressure monitoring on safety during UBE surgery.</p><p><strong>Methods: </strong>A prospective study was conducted involving 20 patients undergoing UBE lumbar spine surgery. Patients were divided into 2 groups based on the irrigation system: gravity-based or infusion pump. Real-time water pressure was monitored using a digital sensor throughout surgery. Each procedure was categorized into 3 phases: phase I, working space preparation; phase II, laminectomy; phase III, flavectomy, dura exposure, and discectomy. Data was analyzed according to the type of irrigation system and surgical phase.</p><p><strong>Results: </strong>The mean water pressure in the surgical field during UBE spine surgery was 17.98± 8.07 mmHg, with no significant differences between surgical phases. However, the infusion pump system maintained significantly lower mean pressure (12.10±3.51 mmHg) compared to the gravity-based system (23.86±6.97 mmHg, p=0.001). The infusion pump system consistently maintained a significantly lower mean water pressure compared to the gravity-based system.</p><p><strong>Conclusion: </strong>Real-time water pressure monitoring during UBE surgery enhances safety by enabling improved control of pressure within the surgical field. Both the gravity-based and infusion pump systems safely maintained working space pressure, with the pump system showing significantly lower pressure levels.</p>","PeriodicalId":19269,"journal":{"name":"Neurospine","volume":"22 3","pages":"812-818"},"PeriodicalIF":3.6000,"publicationDate":"2025-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12518894/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Neurospine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.14245/ns.2550648.324","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/9/30 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"CLINICAL NEUROLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: Unilateral biportal endoscopic (UBE) spine surgery is a minimally invasive technique that uses continuous irrigation to improve visualization and control bleeding. Effective water pressure management is crucial for patient safety, particularly at the cervical and thoracic levels where spinal cord injury risk is higher. However, real-time pressure monitoring remains underexplored. This study evaluates the impact of real-time water pressure monitoring on safety during UBE surgery.
Methods: A prospective study was conducted involving 20 patients undergoing UBE lumbar spine surgery. Patients were divided into 2 groups based on the irrigation system: gravity-based or infusion pump. Real-time water pressure was monitored using a digital sensor throughout surgery. Each procedure was categorized into 3 phases: phase I, working space preparation; phase II, laminectomy; phase III, flavectomy, dura exposure, and discectomy. Data was analyzed according to the type of irrigation system and surgical phase.
Results: The mean water pressure in the surgical field during UBE spine surgery was 17.98± 8.07 mmHg, with no significant differences between surgical phases. However, the infusion pump system maintained significantly lower mean pressure (12.10±3.51 mmHg) compared to the gravity-based system (23.86±6.97 mmHg, p=0.001). The infusion pump system consistently maintained a significantly lower mean water pressure compared to the gravity-based system.
Conclusion: Real-time water pressure monitoring during UBE surgery enhances safety by enabling improved control of pressure within the surgical field. Both the gravity-based and infusion pump systems safely maintained working space pressure, with the pump system showing significantly lower pressure levels.