{"title":"Numerical investigation of the flow induced by a transcatheter intra-aortic entrainment pump.","authors":"Yeojin Park, Osman Aycan, Lyes Kadem","doi":"10.1080/10255842.2025.2471517","DOIUrl":null,"url":null,"abstract":"<p><p>This study evaluates the fluid dynamics inside and outside transcatheter blood pump positioned in the aorta. We focus on the pump's impact on blood component damage and arterial wall stress. CFD simulations were performed for rotational speeds ranging from 6000 to 15000 rpm, with a blood flow rate of 1.6 L/min. Results show that significant blood damage may occur at speeds as low as 12000 rpm, and the pump's outflow jet induces elevated wall shear stress, potentially leading to arterial aneurysms. These findings suggest the need for further design improvements to reduce risks when used in prolonged or transplant-related applications.</p>","PeriodicalId":50640,"journal":{"name":"Computer Methods in Biomechanics and Biomedical Engineering","volume":" ","pages":"1-14"},"PeriodicalIF":1.7000,"publicationDate":"2025-02-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Computer Methods in Biomechanics and Biomedical Engineering","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1080/10255842.2025.2471517","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"COMPUTER SCIENCE, INTERDISCIPLINARY APPLICATIONS","Score":null,"Total":0}
引用次数: 0
Abstract
This study evaluates the fluid dynamics inside and outside transcatheter blood pump positioned in the aorta. We focus on the pump's impact on blood component damage and arterial wall stress. CFD simulations were performed for rotational speeds ranging from 6000 to 15000 rpm, with a blood flow rate of 1.6 L/min. Results show that significant blood damage may occur at speeds as low as 12000 rpm, and the pump's outflow jet induces elevated wall shear stress, potentially leading to arterial aneurysms. These findings suggest the need for further design improvements to reduce risks when used in prolonged or transplant-related applications.
期刊介绍:
The primary aims of Computer Methods in Biomechanics and Biomedical Engineering are to provide a means of communicating the advances being made in the areas of biomechanics and biomedical engineering and to stimulate interest in the continually emerging computer based technologies which are being applied in these multidisciplinary subjects. Computer Methods in Biomechanics and Biomedical Engineering will also provide a focus for the importance of integrating the disciplines of engineering with medical technology and clinical expertise. Such integration will have a major impact on health care in the future.