{"title":"The usefulness of X-ray output management in general radiography systems using exposure index.","authors":"Kazuhiro Ogasawara, Shinya Ohwada, Rie Tachibana, Katsuhiko Ogasawara","doi":"10.1177/08953996241299994","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>The periodic quality control of X-ray devices is important for obtaining optical medical images and determining the appropriate X-ray exposure dose. Additionally, the measurement of the X-ray output is constrained by time, technical aspects, and expenses. Therefore, we investigated the usefulness of a simple method for managing X-ray output using an Exposure Index (EI).</p><p><strong>Methods: </strong>The entire surface of the flat panel detector was X-ray-irradiated every Friday at the time of end-of-work inspection under the condition that the recorded EI was approximately 1000. The EI and exposure dose were measured, and the linearity and accuracy were evaluated.</p><p><strong>Results: </strong>The output gradually decreased from the start of the measurements in Room 1 and stabilized after the output was adjusted. The relationship between exposure dose and EI showed high linearity, with R<sup>2</sup> > 0.99, and the CV of EI was less than 2.41%, indicating high reproducibility.</p><p><strong>Conclusions: </strong>We demonstrated that the results of constancy tests can be easily quantified using EI. The EI method can manage the X-ray output with good reproducibility.</p>","PeriodicalId":49948,"journal":{"name":"Journal of X-Ray Science and Technology","volume":" ","pages":"204-210"},"PeriodicalIF":1.7000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of X-Ray Science and Technology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/08953996241299994","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/13 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"INSTRUMENTS & INSTRUMENTATION","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: The periodic quality control of X-ray devices is important for obtaining optical medical images and determining the appropriate X-ray exposure dose. Additionally, the measurement of the X-ray output is constrained by time, technical aspects, and expenses. Therefore, we investigated the usefulness of a simple method for managing X-ray output using an Exposure Index (EI).
Methods: The entire surface of the flat panel detector was X-ray-irradiated every Friday at the time of end-of-work inspection under the condition that the recorded EI was approximately 1000. The EI and exposure dose were measured, and the linearity and accuracy were evaluated.
Results: The output gradually decreased from the start of the measurements in Room 1 and stabilized after the output was adjusted. The relationship between exposure dose and EI showed high linearity, with R2 > 0.99, and the CV of EI was less than 2.41%, indicating high reproducibility.
Conclusions: We demonstrated that the results of constancy tests can be easily quantified using EI. The EI method can manage the X-ray output with good reproducibility.
期刊介绍:
Research areas within the scope of the journal include:
Interaction of x-rays with matter: x-ray phenomena, biological effects of radiation, radiation safety and optical constants
X-ray sources: x-rays from synchrotrons, x-ray lasers, plasmas, and other sources, conventional or unconventional
Optical elements: grazing incidence optics, multilayer mirrors, zone plates, gratings, other diffraction optics
Optical instruments: interferometers, spectrometers, microscopes, telescopes, microprobes