An ultra-high vacuum scanning tunneling microscope with pulse tube and Joule-Thomson cooling operating at sub-pm z-noise.

IF 1.3 4区 工程技术 Q3 INSTRUMENTS & INSTRUMENTATION
Marcus Eßer, Marco Pratzer, Marc Frömming, Jonas Duffhauß, Priyamvada Bhaskar, Michael A Krzyzowski, Markus Morgenstern
{"title":"An ultra-high vacuum scanning tunneling microscope with pulse tube and Joule-Thomson cooling operating at sub-pm z-noise.","authors":"Marcus Eßer, Marco Pratzer, Marc Frömming, Jonas Duffhauß, Priyamvada Bhaskar, Michael A Krzyzowski, Markus Morgenstern","doi":"10.1063/5.0230892","DOIUrl":null,"url":null,"abstract":"<p><p>Low-temperature scanning tunneling spectroscopy is a key method to probe electronic and magnetic properties down to the atomic scale, but suffers from extreme vibrational sensitivity. This makes it challenging to employ closed-cycle cooling with its required pulse-type vibrational excitations, albeit this is mandatory to avoid helium losses for counteracting the continuously raising helium prices. Here, we describe a compact ultra-high vacuum scanning tunneling microscope (STM) system with an integrated primary pulse tube cooler (PTC) for closed-cycle operation. It achieves temperatures down to 1.5 K via a secondary Joule-Thomson stage and a z-noise down to 300 fmRMS in the STM junction for the frequency range of 0.1 Hz-5 kHz (feedback loop off). This is better than many STMs cooled by an external supply of liquid helium. The challenge to combine an effective vibrational decoupling from the PTC with sufficient thermal conduction is tackled by using a multipartite approach including the concept of bellows with minimal stiffness to decouple the PTC vibrationally from the STM and an optimized STM design with minimal vibrational transfer to the STM junction. As important benchmarks, we could reduce the voltage noise in the tunnel junction down to 120 μV and supply radio frequency excitations up to 40 GHz with amplitudes up to 10 mV in the junction via a close-by antenna. The development principally enables other secondary cooling stages such that it opens the perspective for a helium conserving operation of STMs across the whole interesting temperature range.</p>","PeriodicalId":21111,"journal":{"name":"Review of Scientific Instruments","volume":"95 12","pages":""},"PeriodicalIF":1.3000,"publicationDate":"2024-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Review of Scientific Instruments","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1063/5.0230892","RegionNum":4,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"INSTRUMENTS & INSTRUMENTATION","Score":null,"Total":0}
引用次数: 0

Abstract

Low-temperature scanning tunneling spectroscopy is a key method to probe electronic and magnetic properties down to the atomic scale, but suffers from extreme vibrational sensitivity. This makes it challenging to employ closed-cycle cooling with its required pulse-type vibrational excitations, albeit this is mandatory to avoid helium losses for counteracting the continuously raising helium prices. Here, we describe a compact ultra-high vacuum scanning tunneling microscope (STM) system with an integrated primary pulse tube cooler (PTC) for closed-cycle operation. It achieves temperatures down to 1.5 K via a secondary Joule-Thomson stage and a z-noise down to 300 fmRMS in the STM junction for the frequency range of 0.1 Hz-5 kHz (feedback loop off). This is better than many STMs cooled by an external supply of liquid helium. The challenge to combine an effective vibrational decoupling from the PTC with sufficient thermal conduction is tackled by using a multipartite approach including the concept of bellows with minimal stiffness to decouple the PTC vibrationally from the STM and an optimized STM design with minimal vibrational transfer to the STM junction. As important benchmarks, we could reduce the voltage noise in the tunnel junction down to 120 μV and supply radio frequency excitations up to 40 GHz with amplitudes up to 10 mV in the junction via a close-by antenna. The development principally enables other secondary cooling stages such that it opens the perspective for a helium conserving operation of STMs across the whole interesting temperature range.

求助全文
约1分钟内获得全文 求助全文
来源期刊
Review of Scientific Instruments
Review of Scientific Instruments 工程技术-物理:应用
CiteScore
3.00
自引率
12.50%
发文量
758
审稿时长
2.6 months
期刊介绍: Review of Scientific Instruments, is committed to the publication of advances in scientific instruments, apparatuses, and techniques. RSI seeks to meet the needs of engineers and scientists in physics, chemistry, and the life sciences.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信