Ashutosh Sharma;Vineet Gupta;Joon-Gon Son;Abhishek Gupta;János Bohus;József A. Fülöp;Thomas Gebert
{"title":"Linearity of Fast and Highly Sensitive LiTaO$_{3}$ Pyroelectric Detectors in the Terahertz Range","authors":"Ashutosh Sharma;Vineet Gupta;Joon-Gon Son;Abhishek Gupta;János Bohus;József A. Fülöp;Thomas Gebert","doi":"10.1109/TTHZ.2024.3445603","DOIUrl":null,"url":null,"abstract":"In response to single-cycle THz pulses, we measured the linearity parameter of lithium tantalate pyroelectric detectors, over nearly three orders of magnitude in pulse energy, up to 2 \n<inline-formula><tex-math>$\\mu$</tex-math></inline-formula>\nJ. The response was linear up to 1 \n<inline-formula><tex-math>$\\mu$</tex-math></inline-formula>\nJ pulse energy and 25 \n<inline-formula><tex-math>$\\mu$</tex-math></inline-formula>\nJ/cm\n<inline-formula><tex-math>$^{2}$</tex-math></inline-formula>\n energy flux, and sublinear at higher irradiations. This finding shows the importance of characterizing the detector response linearity for accurate THz metrology. The detectors had high sensitivity in single-pulse detection mode up to 10 kHz repetition rate.","PeriodicalId":13258,"journal":{"name":"IEEE Transactions on Terahertz Science and Technology","volume":"14 6","pages":"823-829"},"PeriodicalIF":3.9000,"publicationDate":"2024-08-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ieeexplore.ieee.org/stamp/stamp.jsp?tp=&arnumber=10638827","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE Transactions on Terahertz Science and Technology","FirstCategoryId":"5","ListUrlMain":"https://ieeexplore.ieee.org/document/10638827/","RegionNum":2,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ENGINEERING, ELECTRICAL & ELECTRONIC","Score":null,"Total":0}
引用次数: 0
Abstract
In response to single-cycle THz pulses, we measured the linearity parameter of lithium tantalate pyroelectric detectors, over nearly three orders of magnitude in pulse energy, up to 2
$\mu$
J. The response was linear up to 1
$\mu$
J pulse energy and 25
$\mu$
J/cm
$^{2}$
energy flux, and sublinear at higher irradiations. This finding shows the importance of characterizing the detector response linearity for accurate THz metrology. The detectors had high sensitivity in single-pulse detection mode up to 10 kHz repetition rate.
期刊介绍:
IEEE Transactions on Terahertz Science and Technology focuses on original research on Terahertz theory, techniques, and applications as they relate to components, devices, circuits, and systems involving the generation, transmission, and detection of Terahertz waves.