Henner Huflage, Robin Hendel, Andreas Steven Kunz, Süleyman Ergün, Saif Afat, Nils Petri, Viktor Hartung, Philipp Gruschwitz, Thorsten Alexander Bley, Jan-Peter Grunz
{"title":"研究超高分辨率肺部光子计数 CT 中的小像素效应。","authors":"Henner Huflage, Robin Hendel, Andreas Steven Kunz, Süleyman Ergün, Saif Afat, Nils Petri, Viktor Hartung, Philipp Gruschwitz, Thorsten Alexander Bley, Jan-Peter Grunz","doi":"10.1097/RLI.0000000000001013","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>The aim of this study was to investigate potential benefits of ultra-high resolution (UHR) over standard resolution scan mode in ultra-low dose photon-counting detector CT (PCD-CT) of the lung.</p><p><strong>Materials and methods: </strong>Six cadaveric specimens were examined with 5 dose settings using tin prefiltration, each in UHR (120 × 0.2 mm) and standard mode (144 × 0.4 mm), on a first-generation PCD-CT scanner. Image quality was evaluated quantitatively by noise comparisons in the trachea and both main bronchi. In addition, 16 readers (14 radiologists and 2 internal medicine physicians) independently completed a browser-based pairwise forced-choice comparison task for assessment of subjective image quality. The Kendall rank coefficient ( W ) was calculated to assess interrater agreement, and Pearson's correlation coefficient ( r ) was used to analyze the relationship between noise measurements and image quality rankings.</p><p><strong>Results: </strong>Across all dose levels, image noise in UHR mode was lower than in standard mode for scan protocols matched by CTDI vol ( P < 0.001). UHR examinations exhibited noise levels comparable to the next higher dose setting in standard mode ( P ≥ 0.275). Subjective ranking of protocols based on 5760 pairwise tests showed high interrater agreement ( W = 0.99; P ≤ 0.001) with UHR images being preferred by readers in the majority of comparisons. Irrespective of scan mode, a substantial indirect correlation was observed between image noise and subjective image quality ranking ( r = -0.97; P ≤ 0.001).</p><p><strong>Conclusions: </strong>In PCD-CT of the lung, UHR scan mode reduces image noise considerably over standard resolution acquisition. Originating from the smaller detector element size in fan direction, the small pixel effect allows for superior image quality in ultra-low dose examinations with considerable potential for radiation dose reduction.</p>","PeriodicalId":14486,"journal":{"name":"Investigative Radiology","volume":" ","pages":"293-297"},"PeriodicalIF":7.0000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Investigating the Small Pixel Effect in Ultra-High Resolution Photon-Counting CT of the Lung.\",\"authors\":\"Henner Huflage, Robin Hendel, Andreas Steven Kunz, Süleyman Ergün, Saif Afat, Nils Petri, Viktor Hartung, Philipp Gruschwitz, Thorsten Alexander Bley, Jan-Peter Grunz\",\"doi\":\"10.1097/RLI.0000000000001013\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objectives: </strong>The aim of this study was to investigate potential benefits of ultra-high resolution (UHR) over standard resolution scan mode in ultra-low dose photon-counting detector CT (PCD-CT) of the lung.</p><p><strong>Materials and methods: </strong>Six cadaveric specimens were examined with 5 dose settings using tin prefiltration, each in UHR (120 × 0.2 mm) and standard mode (144 × 0.4 mm), on a first-generation PCD-CT scanner. Image quality was evaluated quantitatively by noise comparisons in the trachea and both main bronchi. In addition, 16 readers (14 radiologists and 2 internal medicine physicians) independently completed a browser-based pairwise forced-choice comparison task for assessment of subjective image quality. The Kendall rank coefficient ( W ) was calculated to assess interrater agreement, and Pearson's correlation coefficient ( r ) was used to analyze the relationship between noise measurements and image quality rankings.</p><p><strong>Results: </strong>Across all dose levels, image noise in UHR mode was lower than in standard mode for scan protocols matched by CTDI vol ( P < 0.001). UHR examinations exhibited noise levels comparable to the next higher dose setting in standard mode ( P ≥ 0.275). Subjective ranking of protocols based on 5760 pairwise tests showed high interrater agreement ( W = 0.99; P ≤ 0.001) with UHR images being preferred by readers in the majority of comparisons. Irrespective of scan mode, a substantial indirect correlation was observed between image noise and subjective image quality ranking ( r = -0.97; P ≤ 0.001).</p><p><strong>Conclusions: </strong>In PCD-CT of the lung, UHR scan mode reduces image noise considerably over standard resolution acquisition. Originating from the smaller detector element size in fan direction, the small pixel effect allows for superior image quality in ultra-low dose examinations with considerable potential for radiation dose reduction.</p>\",\"PeriodicalId\":14486,\"journal\":{\"name\":\"Investigative Radiology\",\"volume\":\" \",\"pages\":\"293-297\"},\"PeriodicalIF\":7.0000,\"publicationDate\":\"2024-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Investigative Radiology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1097/RLI.0000000000001013\",\"RegionNum\":1,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Investigative Radiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/RLI.0000000000001013","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
Investigating the Small Pixel Effect in Ultra-High Resolution Photon-Counting CT of the Lung.
Objectives: The aim of this study was to investigate potential benefits of ultra-high resolution (UHR) over standard resolution scan mode in ultra-low dose photon-counting detector CT (PCD-CT) of the lung.
Materials and methods: Six cadaveric specimens were examined with 5 dose settings using tin prefiltration, each in UHR (120 × 0.2 mm) and standard mode (144 × 0.4 mm), on a first-generation PCD-CT scanner. Image quality was evaluated quantitatively by noise comparisons in the trachea and both main bronchi. In addition, 16 readers (14 radiologists and 2 internal medicine physicians) independently completed a browser-based pairwise forced-choice comparison task for assessment of subjective image quality. The Kendall rank coefficient ( W ) was calculated to assess interrater agreement, and Pearson's correlation coefficient ( r ) was used to analyze the relationship between noise measurements and image quality rankings.
Results: Across all dose levels, image noise in UHR mode was lower than in standard mode for scan protocols matched by CTDI vol ( P < 0.001). UHR examinations exhibited noise levels comparable to the next higher dose setting in standard mode ( P ≥ 0.275). Subjective ranking of protocols based on 5760 pairwise tests showed high interrater agreement ( W = 0.99; P ≤ 0.001) with UHR images being preferred by readers in the majority of comparisons. Irrespective of scan mode, a substantial indirect correlation was observed between image noise and subjective image quality ranking ( r = -0.97; P ≤ 0.001).
Conclusions: In PCD-CT of the lung, UHR scan mode reduces image noise considerably over standard resolution acquisition. Originating from the smaller detector element size in fan direction, the small pixel effect allows for superior image quality in ultra-low dose examinations with considerable potential for radiation dose reduction.
期刊介绍:
Investigative Radiology publishes original, peer-reviewed reports on clinical and laboratory investigations in diagnostic imaging, the diagnostic use of radioactive isotopes, computed tomography, positron emission tomography, magnetic resonance imaging, ultrasound, digital subtraction angiography, and related modalities. Emphasis is on early and timely publication. Primarily research-oriented, the journal also includes a wide variety of features of interest to clinical radiologists.