Achim Langenbucher, Peter Hoffmann, Alan Cayless, Matthias Bolz, Jascha Wendelstein, Nóra Szentmáry
{"title":"使用蒙特卡洛模拟法分析生物测量参数的不确定性对眼内透镜功率公式预测屈光度的影响","authors":"Achim Langenbucher, Peter Hoffmann, Alan Cayless, Matthias Bolz, Jascha Wendelstein, Nóra Szentmáry","doi":"10.1111/aos.15726","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Purpose</h3>\n \n <p>The purpose of this study was to investigate the uncertainty in the formula predicted refractive outcome REFU after cataract surgery resulting from measurement uncertainties in modern optical biometers using literature data for within-subject standard deviation <i>S</i><sub>w</sub>.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>This Monte–Carlo simulation study used a large dataset containing 16 667 preoperative IOLMaster 700 biometric measurements. Based on literature <i>S</i><sub>w</sub> values, REFU was derived for both the Haigis and Castrop formulae using error propagation strategies. Using the Hoya Vivinex lens (IOL) as an example, REFU was calculated both with (WLT) and without (WoLT) consideration of IOL power labelling tolerances.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>WoLT the median REFU was 0.10/0.12 dpt for the Haigis/Castrop formula, and WLT it was 0.13/0.15 dpt. WoLT REFU increased systematically for short eyes (or high power IOLs), and WLT this effect was even more pronounced because of increased labelling tolerances. WoLT the uncertainty in the measurement of the corneal front surface radius showed the largest contribution to REFU, especially in long eyes (and low power IOLs). WLT the IOL power uncertainty dominated in short eyes (or high power IOLs) and the uncertainty of the corneal front surface in long eyes (or low power IOLs).</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>Compared with published data on the formula prediction error of refractive outcome after cataract surgery, the uncertainty of biometric measures seems to contribute with ⅓ to ½ to the entire standard deviation. REFU systematically increases with IOL power and decreases with axial length.</p>\n </section>\n </div>","PeriodicalId":6915,"journal":{"name":"Acta Ophthalmologica","volume":"102 3","pages":"e285-e295"},"PeriodicalIF":3.0000,"publicationDate":"2023-06-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1111/aos.15726","citationCount":"0","resultStr":"{\"title\":\"Impact of uncertainties in biometric parameters on intraocular lens power formula predicted refraction using a Monte-Carlo simulation\",\"authors\":\"Achim Langenbucher, Peter Hoffmann, Alan Cayless, Matthias Bolz, Jascha Wendelstein, Nóra Szentmáry\",\"doi\":\"10.1111/aos.15726\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div>\\n \\n \\n <section>\\n \\n <h3> Purpose</h3>\\n \\n <p>The purpose of this study was to investigate the uncertainty in the formula predicted refractive outcome REFU after cataract surgery resulting from measurement uncertainties in modern optical biometers using literature data for within-subject standard deviation <i>S</i><sub>w</sub>.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Methods</h3>\\n \\n <p>This Monte–Carlo simulation study used a large dataset containing 16 667 preoperative IOLMaster 700 biometric measurements. Based on literature <i>S</i><sub>w</sub> values, REFU was derived for both the Haigis and Castrop formulae using error propagation strategies. Using the Hoya Vivinex lens (IOL) as an example, REFU was calculated both with (WLT) and without (WoLT) consideration of IOL power labelling tolerances.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Results</h3>\\n \\n <p>WoLT the median REFU was 0.10/0.12 dpt for the Haigis/Castrop formula, and WLT it was 0.13/0.15 dpt. WoLT REFU increased systematically for short eyes (or high power IOLs), and WLT this effect was even more pronounced because of increased labelling tolerances. WoLT the uncertainty in the measurement of the corneal front surface radius showed the largest contribution to REFU, especially in long eyes (and low power IOLs). WLT the IOL power uncertainty dominated in short eyes (or high power IOLs) and the uncertainty of the corneal front surface in long eyes (or low power IOLs).</p>\\n </section>\\n \\n <section>\\n \\n <h3> Conclusions</h3>\\n \\n <p>Compared with published data on the formula prediction error of refractive outcome after cataract surgery, the uncertainty of biometric measures seems to contribute with ⅓ to ½ to the entire standard deviation. REFU systematically increases with IOL power and decreases with axial length.</p>\\n </section>\\n </div>\",\"PeriodicalId\":6915,\"journal\":{\"name\":\"Acta Ophthalmologica\",\"volume\":\"102 3\",\"pages\":\"e285-e295\"},\"PeriodicalIF\":3.0000,\"publicationDate\":\"2023-06-23\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://onlinelibrary.wiley.com/doi/epdf/10.1111/aos.15726\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Acta Ophthalmologica\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1111/aos.15726\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"OPHTHALMOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta Ophthalmologica","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/aos.15726","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
Impact of uncertainties in biometric parameters on intraocular lens power formula predicted refraction using a Monte-Carlo simulation
Purpose
The purpose of this study was to investigate the uncertainty in the formula predicted refractive outcome REFU after cataract surgery resulting from measurement uncertainties in modern optical biometers using literature data for within-subject standard deviation Sw.
Methods
This Monte–Carlo simulation study used a large dataset containing 16 667 preoperative IOLMaster 700 biometric measurements. Based on literature Sw values, REFU was derived for both the Haigis and Castrop formulae using error propagation strategies. Using the Hoya Vivinex lens (IOL) as an example, REFU was calculated both with (WLT) and without (WoLT) consideration of IOL power labelling tolerances.
Results
WoLT the median REFU was 0.10/0.12 dpt for the Haigis/Castrop formula, and WLT it was 0.13/0.15 dpt. WoLT REFU increased systematically for short eyes (or high power IOLs), and WLT this effect was even more pronounced because of increased labelling tolerances. WoLT the uncertainty in the measurement of the corneal front surface radius showed the largest contribution to REFU, especially in long eyes (and low power IOLs). WLT the IOL power uncertainty dominated in short eyes (or high power IOLs) and the uncertainty of the corneal front surface in long eyes (or low power IOLs).
Conclusions
Compared with published data on the formula prediction error of refractive outcome after cataract surgery, the uncertainty of biometric measures seems to contribute with ⅓ to ½ to the entire standard deviation. REFU systematically increases with IOL power and decreases with axial length.
期刊介绍:
Acta Ophthalmologica is published on behalf of the Acta Ophthalmologica Scandinavica Foundation and is the official scientific publication of the following societies: The Danish Ophthalmological Society, The Finnish Ophthalmological Society, The Icelandic Ophthalmological Society, The Norwegian Ophthalmological Society and The Swedish Ophthalmological Society, and also the European Association for Vision and Eye Research (EVER).
Acta Ophthalmologica publishes clinical and experimental original articles, reviews, editorials, educational photo essays (Diagnosis and Therapy in Ophthalmology), case reports and case series, letters to the editor and doctoral theses.