Marli van Heerden, Jaya A George, Siyabonga Khoza
{"title":"西格玛指标在实验室的应用,以评估南非的质量控制过程。","authors":"Marli van Heerden,&nbsp;Jaya A George,&nbsp;Siyabonga Khoza","doi":"10.4102/ajlm.v11i1.1344","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Laboratories use quality control processes to monitor and evaluate analytical performance in terms of precision and bias. Sigma metrics provide an objective assessment of laboratory quality using the total allowable error as an additional parameter.</p><p><strong>Objective: </strong>This study aimed to determine the sigma metrics of analytes when using different total allowable error guidelines.</p><p><strong>Methods: </strong>A retrospective analysis was performed on 19 general chemistry analytes at Charlotte Maxeke Johannesburg Academic Hospital in South Africa between January 2017 and December 2017. Sigma metrics were calculated on two identical analysers, using internal quality control data and total allowable error guidelines from the Ricos biological variation database and three alternative sources (the Royal College of Pathologists of Australasia, the Clinical Laboratory Improvements Amendment, and the European Federation of Clinical Chemistry and Laboratory Medicine).</p><p><strong>Results: </strong>The sigma performance was similar on both analysers but varied based on the guideline used, with the Clinical Laboratory Improvements Amendment guidelines resulting in the best sigma metrics (53% of analytes on one analyser and 46% on the other had acceptable sigma metrics) and the Royal College of Pathologists of Australia guidelines being the most stringent (21% and 23%). Sodium and chloride performed poorly across all guidelines (sigma < 3). There were also month-to-month variations that may result in acceptable sigma despite poor performance during certain months.</p><p><strong>Conclusion: </strong>The sigma varies greatly depending on the total allowable error, but could be a valuable tool to save time and decrease costs in high-volume laboratories. Sigma metrics calculations need to be standardised.</p>","PeriodicalId":45412,"journal":{"name":"African Journal of Laboratory Medicine","volume":"11 1","pages":"1344"},"PeriodicalIF":1.0000,"publicationDate":"2022-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9257767/pdf/","citationCount":"1","resultStr":"{\"title\":\"The application of sigma metrics in the laboratory to assess quality control processes in South Africa.\",\"authors\":\"Marli van Heerden,&nbsp;Jaya A George,&nbsp;Siyabonga Khoza\",\"doi\":\"10.4102/ajlm.v11i1.1344\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Laboratories use quality control processes to monitor and evaluate analytical performance in terms of precision and bias. Sigma metrics provide an objective assessment of laboratory quality using the total allowable error as an additional parameter.</p><p><strong>Objective: </strong>This study aimed to determine the sigma metrics of analytes when using different total allowable error guidelines.</p><p><strong>Methods: </strong>A retrospective analysis was performed on 19 general chemistry analytes at Charlotte Maxeke Johannesburg Academic Hospital in South Africa between January 2017 and December 2017. Sigma metrics were calculated on two identical analysers, using internal quality control data and total allowable error guidelines from the Ricos biological variation database and three alternative sources (the Royal College of Pathologists of Australasia, the Clinical Laboratory Improvements Amendment, and the European Federation of Clinical Chemistry and Laboratory Medicine).</p><p><strong>Results: </strong>The sigma performance was similar on both analysers but varied based on the guideline used, with the Clinical Laboratory Improvements Amendment guidelines resulting in the best sigma metrics (53% of analytes on one analyser and 46% on the other had acceptable sigma metrics) and the Royal College of Pathologists of Australia guidelines being the most stringent (21% and 23%). Sodium and chloride performed poorly across all guidelines (sigma < 3). There were also month-to-month variations that may result in acceptable sigma despite poor performance during certain months.</p><p><strong>Conclusion: </strong>The sigma varies greatly depending on the total allowable error, but could be a valuable tool to save time and decrease costs in high-volume laboratories. Sigma metrics calculations need to be standardised.</p>\",\"PeriodicalId\":45412,\"journal\":{\"name\":\"African Journal of Laboratory Medicine\",\"volume\":\"11 1\",\"pages\":\"1344\"},\"PeriodicalIF\":1.0000,\"publicationDate\":\"2022-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9257767/pdf/\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"African Journal of Laboratory Medicine\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.4102/ajlm.v11i1.1344\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"MEDICINE, RESEARCH & EXPERIMENTAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"African Journal of Laboratory Medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4102/ajlm.v11i1.1344","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"MEDICINE, RESEARCH & EXPERIMENTAL","Score":null,"Total":0}
引用次数: 1

摘要

背景:实验室使用质量控制程序来监测和评估分析的精度和偏差。西格玛指标使用总允许误差作为附加参数,提供对实验室质量的客观评估。目的:本研究旨在确定在使用不同的总允许误差准则时分析物的西格玛指标。方法:对2017年1月至2017年12月在南非Charlotte Maxeke约翰内斯堡学术医院进行的19例普通化学分析进行回顾性分析。在两台相同的分析仪上计算Sigma指标,使用内部质量控制数据和来自Ricos生物变异数据库的总允许误差指南以及三个替代来源(澳大利亚皇家病理学家学院、临床实验室改进修正案和欧洲临床化学和实验室医学联合会)。结果:两种分析仪的西格玛表现相似,但根据所使用的指南而有所不同,临床实验室改进修订指南产生了最好的西格玛指标(一台分析仪的分析物的53%和另一台分析仪的46%具有可接受的西格玛指标),澳大利亚皇家病理学家学院指南是最严格的(21%和23%)。钠和氯化物在所有指南中的表现都很差(sigma < 3)。尽管在某些月份表现不佳,但每月的变化也可能导致可接受的sigma。结论:sigma随总允许误差的不同而变化很大,但在大批量实验室中,它是一种节省时间和降低成本的有价值的工具。西格玛度量的计算需要标准化。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

The application of sigma metrics in the laboratory to assess quality control processes in South Africa.

The application of sigma metrics in the laboratory to assess quality control processes in South Africa.

The application of sigma metrics in the laboratory to assess quality control processes in South Africa.

The application of sigma metrics in the laboratory to assess quality control processes in South Africa.

Background: Laboratories use quality control processes to monitor and evaluate analytical performance in terms of precision and bias. Sigma metrics provide an objective assessment of laboratory quality using the total allowable error as an additional parameter.

Objective: This study aimed to determine the sigma metrics of analytes when using different total allowable error guidelines.

Methods: A retrospective analysis was performed on 19 general chemistry analytes at Charlotte Maxeke Johannesburg Academic Hospital in South Africa between January 2017 and December 2017. Sigma metrics were calculated on two identical analysers, using internal quality control data and total allowable error guidelines from the Ricos biological variation database and three alternative sources (the Royal College of Pathologists of Australasia, the Clinical Laboratory Improvements Amendment, and the European Federation of Clinical Chemistry and Laboratory Medicine).

Results: The sigma performance was similar on both analysers but varied based on the guideline used, with the Clinical Laboratory Improvements Amendment guidelines resulting in the best sigma metrics (53% of analytes on one analyser and 46% on the other had acceptable sigma metrics) and the Royal College of Pathologists of Australia guidelines being the most stringent (21% and 23%). Sodium and chloride performed poorly across all guidelines (sigma < 3). There were also month-to-month variations that may result in acceptable sigma despite poor performance during certain months.

Conclusion: The sigma varies greatly depending on the total allowable error, but could be a valuable tool to save time and decrease costs in high-volume laboratories. Sigma metrics calculations need to be standardised.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
African Journal of Laboratory Medicine
African Journal of Laboratory Medicine MEDICINE, RESEARCH & EXPERIMENTAL-
CiteScore
1.70
自引率
9.10%
发文量
53
审稿时长
12 weeks
期刊介绍: The African Journal of Laboratory Medicine, the official journal of ASLM, focuses on the role of the laboratory and its professionals in the clinical and public healthcare sectors,and is specifically based on an African frame of reference. Emphasis is on all aspects that promote and contribute to the laboratory medicine practices of Africa. This includes, amongst others: laboratories, biomedical scientists and clinicians, medical community, public health officials and policy makers, laboratory systems and policies (translation of laboratory knowledge, practices and technologies in clinical care), interfaces of laboratory with medical science, laboratory-based epidemiology, laboratory investigations, evidence-based effectiveness in real world (actual) settings.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信