Anna Lorenzin, Luca Sgarabotto, Maria Laura Bacci, Alberto Elmi, Domenico Ventrella, Camilla Aniballi, Monica Zanella, Alessandra Brendolan, Luca Di Lullo, Claudio Ronco
{"title":"人工利尿:新型微型体外超滤装置的有效性和安全性动物研究。","authors":"Anna Lorenzin, Luca Sgarabotto, Maria Laura Bacci, Alberto Elmi, Domenico Ventrella, Camilla Aniballi, Monica Zanella, Alessandra Brendolan, Luca Di Lullo, Claudio Ronco","doi":"10.1159/000530382","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>We have recently developed a new miniaturized device for extracorporeal ultrafiltration (UF) to be used in patients with fluid overload: Artificial Diuresis-1 (AD1) (Medica S.p.A., Medolla, Italy). The device has a reduced priming volume, operates at very low pressures and flow regimes, and is designed to perform extracorporeal UF at bedside. After accurate experiments were carried out in vitro, we report in this paper the results of in vivo UF sessions carried out in selected animals according to veterinary best practice.</p><p><strong>Materials and methods: </strong>The AD1 kit is pre-filled with sterile isotonic solution and operates with a polysulfone mini-filter, MediSulfone (polysulfone at 50,000 Dalton). A collection bag with a volumetric scale is connected to the UF line, and the ultrafiltrate is obtained by gravity based on the height at which the ultrafiltrate collection bag is placed. Animals were prepared and anesthetized. The jugular vein was cannulated with a double-lumen catheter. Three 6-h sessions of UF were scheduled with a target fluid removal of 1,500 mL. Heparin was used as anticoagulant.</p><p><strong>Results: </strong>In all treatments, the target value of UF was obtained in the absence of major clinical or technical problems with a maximum deviation from the scheduled UF rate lower than 10%. The device resulted to be safe, reliable, accurate, and easily usable thanks to a user-friendly interface and its very small dimensions.</p><p><strong>Conclusions: </strong>This study opens the way for clinical trials in different settings including departments with low intensity of care and even in ambulatory centers or patient's home.</p>","PeriodicalId":9584,"journal":{"name":"Cardiorenal Medicine","volume":" ","pages":"167-175"},"PeriodicalIF":2.4000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Artificial Diuresis: Animal Studies on Efficacy and Safety of a New Miniaturized Device for Extracorporeal Ultrafiltration.\",\"authors\":\"Anna Lorenzin, Luca Sgarabotto, Maria Laura Bacci, Alberto Elmi, Domenico Ventrella, Camilla Aniballi, Monica Zanella, Alessandra Brendolan, Luca Di Lullo, Claudio Ronco\",\"doi\":\"10.1159/000530382\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Introduction: </strong>We have recently developed a new miniaturized device for extracorporeal ultrafiltration (UF) to be used in patients with fluid overload: Artificial Diuresis-1 (AD1) (Medica S.p.A., Medolla, Italy). The device has a reduced priming volume, operates at very low pressures and flow regimes, and is designed to perform extracorporeal UF at bedside. After accurate experiments were carried out in vitro, we report in this paper the results of in vivo UF sessions carried out in selected animals according to veterinary best practice.</p><p><strong>Materials and methods: </strong>The AD1 kit is pre-filled with sterile isotonic solution and operates with a polysulfone mini-filter, MediSulfone (polysulfone at 50,000 Dalton). A collection bag with a volumetric scale is connected to the UF line, and the ultrafiltrate is obtained by gravity based on the height at which the ultrafiltrate collection bag is placed. Animals were prepared and anesthetized. The jugular vein was cannulated with a double-lumen catheter. Three 6-h sessions of UF were scheduled with a target fluid removal of 1,500 mL. Heparin was used as anticoagulant.</p><p><strong>Results: </strong>In all treatments, the target value of UF was obtained in the absence of major clinical or technical problems with a maximum deviation from the scheduled UF rate lower than 10%. The device resulted to be safe, reliable, accurate, and easily usable thanks to a user-friendly interface and its very small dimensions.</p><p><strong>Conclusions: </strong>This study opens the way for clinical trials in different settings including departments with low intensity of care and even in ambulatory centers or patient's home.</p>\",\"PeriodicalId\":9584,\"journal\":{\"name\":\"Cardiorenal Medicine\",\"volume\":\" \",\"pages\":\"167-175\"},\"PeriodicalIF\":2.4000,\"publicationDate\":\"2023-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Cardiorenal Medicine\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1159/000530382\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2023/3/28 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q2\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Cardiorenal Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1159/000530382","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/3/28 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
Artificial Diuresis: Animal Studies on Efficacy and Safety of a New Miniaturized Device for Extracorporeal Ultrafiltration.
Introduction: We have recently developed a new miniaturized device for extracorporeal ultrafiltration (UF) to be used in patients with fluid overload: Artificial Diuresis-1 (AD1) (Medica S.p.A., Medolla, Italy). The device has a reduced priming volume, operates at very low pressures and flow regimes, and is designed to perform extracorporeal UF at bedside. After accurate experiments were carried out in vitro, we report in this paper the results of in vivo UF sessions carried out in selected animals according to veterinary best practice.
Materials and methods: The AD1 kit is pre-filled with sterile isotonic solution and operates with a polysulfone mini-filter, MediSulfone (polysulfone at 50,000 Dalton). A collection bag with a volumetric scale is connected to the UF line, and the ultrafiltrate is obtained by gravity based on the height at which the ultrafiltrate collection bag is placed. Animals were prepared and anesthetized. The jugular vein was cannulated with a double-lumen catheter. Three 6-h sessions of UF were scheduled with a target fluid removal of 1,500 mL. Heparin was used as anticoagulant.
Results: In all treatments, the target value of UF was obtained in the absence of major clinical or technical problems with a maximum deviation from the scheduled UF rate lower than 10%. The device resulted to be safe, reliable, accurate, and easily usable thanks to a user-friendly interface and its very small dimensions.
Conclusions: This study opens the way for clinical trials in different settings including departments with low intensity of care and even in ambulatory centers or patient's home.
期刊介绍:
The journal ''Cardiorenal Medicine'' explores the mechanisms by which obesity and other metabolic abnormalities promote the pathogenesis and progression of heart and kidney disease (cardiorenal metabolic syndrome). It provides an interdisciplinary platform for the advancement of research and clinical practice, focussing on translational issues.