{"title":"圆锥角膜交联后角膜的生物力学变化。","authors":"Adel Hamid,&nbsp;Hamidreza Jahadi-Hosseini,&nbsp;Mohammad Reza Khalili,&nbsp;Hamidreza Jahanbani-Ardakani","doi":"10.4103/joco.joco_177_21","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate the changes in biomechanical properties of the cornea using the Corvis ST device after the treatment of keratoconus patients with ultraviolet-A/riboflavin corneal cross-linking (CXL).</p><p><strong>Methods: </strong>Thirty-seven eyes from 37 consecutive patients with progressive keratoconus were included in this prospective observational case series. Corneal biomechanical parameters including the length of the applanated cornea (L1 and L2), corneal movement velocity during applanation (V1 and V2) at the moment of the first and second applanation, deformation amplitude (DA), distance between bending points of the cornea points of the cornea (PD), and concave radius of curvature (R) at the point of the highest concavity were recorded using the Corvis ST at baseline, 3 months, and 1 year after CXL.</p><p><strong>Results: </strong>The mean age of the patients was 23.27 years (range, 19-31 years). Among CorVis ST corneal biomechanical parameters, L1, DA, PD, and R at the point of the highest concavity did not change significantly. The length of the applanated cornea at the moment of second applanation (L2) showed a significant change 3 months after CXL, but no significant difference was found between the 3-month and 1-year values of this parameter. Corneal movement velocity during applanation (V1 and V2) did not change 3 months after doing CXL, but the changes in these parameters were significant 1 year after CXL.</p><p><strong>Conclusions: </strong>Although the CorVis ST device may detect changes in some biomechanical properties of cornea after the treatment of keratoconus patients with CXL, many parameters remain unchanged, and this device cannot readily be used to find the effects of CXL.</p>","PeriodicalId":15423,"journal":{"name":"Journal of Current Ophthalmology","volume":null,"pages":null},"PeriodicalIF":1.2000,"publicationDate":"2022-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/4a/bb/JCO-34-409.PMC10170983.pdf","citationCount":"0","resultStr":"{\"title\":\"Corneal Biomechanical Changes after Corneal Cross-Linking in Patients with Keratoconus.\",\"authors\":\"Adel Hamid,&nbsp;Hamidreza Jahadi-Hosseini,&nbsp;Mohammad Reza Khalili,&nbsp;Hamidreza Jahanbani-Ardakani\",\"doi\":\"10.4103/joco.joco_177_21\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>To evaluate the changes in biomechanical properties of the cornea using the Corvis ST device after the treatment of keratoconus patients with ultraviolet-A/riboflavin corneal cross-linking (CXL).</p><p><strong>Methods: </strong>Thirty-seven eyes from 37 consecutive patients with progressive keratoconus were included in this prospective observational case series. Corneal biomechanical parameters including the length of the applanated cornea (L1 and L2), corneal movement velocity during applanation (V1 and V2) at the moment of the first and second applanation, deformation amplitude (DA), distance between bending points of the cornea points of the cornea (PD), and concave radius of curvature (R) at the point of the highest concavity were recorded using the Corvis ST at baseline, 3 months, and 1 year after CXL.</p><p><strong>Results: </strong>The mean age of the patients was 23.27 years (range, 19-31 years). Among CorVis ST corneal biomechanical parameters, L1, DA, PD, and R at the point of the highest concavity did not change significantly. The length of the applanated cornea at the moment of second applanation (L2) showed a significant change 3 months after CXL, but no significant difference was found between the 3-month and 1-year values of this parameter. Corneal movement velocity during applanation (V1 and V2) did not change 3 months after doing CXL, but the changes in these parameters were significant 1 year after CXL.</p><p><strong>Conclusions: </strong>Although the CorVis ST device may detect changes in some biomechanical properties of cornea after the treatment of keratoconus patients with CXL, many parameters remain unchanged, and this device cannot readily be used to find the effects of CXL.</p>\",\"PeriodicalId\":15423,\"journal\":{\"name\":\"Journal of Current Ophthalmology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.2000,\"publicationDate\":\"2022-10-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/4a/bb/JCO-34-409.PMC10170983.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Current Ophthalmology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.4103/joco.joco_177_21\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"OPHTHALMOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Current Ophthalmology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/joco.joco_177_21","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

目的:探讨uv - a /核黄素角膜交联(CXL)治疗圆锥角膜患者后,Corvis ST装置对角膜生物力学性能的影响。方法:37例连续进行性圆锥角膜患者的37只眼睛被纳入前瞻性观察病例系列。使用Corvis ST记录角膜生物力学参数,包括被压平角膜的长度(L1和L2)、第一次和第二次压平时角膜在压平过程中的运动速度(V1和V2)、变形幅度(DA)、角膜点弯曲点之间的距离(PD)和最高凹度点的凹曲率半径(R)。结果:患者平均年龄23.27岁(范围19 ~ 31岁)。CorVis ST角膜生物力学参数中,最高凹点L1、DA、PD、R无明显变化。第二次压平时刻(L2)的被压角膜长度在CXL后3个月有显著变化,但该参数在3个月和1年之间无显著差异。角膜压平运动速度(V1和V2)在CXL术后3个月无变化,但这些参数在CXL术后1年变化显著。结论:虽然CorVis ST装置可以检测到圆锥角膜患者行CXL治疗后角膜部分生物力学特性的变化,但许多参数仍未改变,该装置不易用于发现CXL的作用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Corneal Biomechanical Changes after Corneal Cross-Linking in Patients with Keratoconus.

Purpose: To evaluate the changes in biomechanical properties of the cornea using the Corvis ST device after the treatment of keratoconus patients with ultraviolet-A/riboflavin corneal cross-linking (CXL).

Methods: Thirty-seven eyes from 37 consecutive patients with progressive keratoconus were included in this prospective observational case series. Corneal biomechanical parameters including the length of the applanated cornea (L1 and L2), corneal movement velocity during applanation (V1 and V2) at the moment of the first and second applanation, deformation amplitude (DA), distance between bending points of the cornea points of the cornea (PD), and concave radius of curvature (R) at the point of the highest concavity were recorded using the Corvis ST at baseline, 3 months, and 1 year after CXL.

Results: The mean age of the patients was 23.27 years (range, 19-31 years). Among CorVis ST corneal biomechanical parameters, L1, DA, PD, and R at the point of the highest concavity did not change significantly. The length of the applanated cornea at the moment of second applanation (L2) showed a significant change 3 months after CXL, but no significant difference was found between the 3-month and 1-year values of this parameter. Corneal movement velocity during applanation (V1 and V2) did not change 3 months after doing CXL, but the changes in these parameters were significant 1 year after CXL.

Conclusions: Although the CorVis ST device may detect changes in some biomechanical properties of cornea after the treatment of keratoconus patients with CXL, many parameters remain unchanged, and this device cannot readily be used to find the effects of CXL.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
2.50
自引率
6.70%
发文量
45
审稿时长
8 weeks
期刊介绍: Peer Review under the responsibility of Iranian Society of Ophthalmology Journal of Current Ophthalmology, the official publication of the Iranian Society of Ophthalmology, is a peer-reviewed, open-access, scientific journal that welcomes high quality original articles related to vision science and all fields of ophthalmology. Journal of Current Ophthalmology is the continuum of Iranian Journal of Ophthalmology published since 1969.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信