Jing Li, Li-Hong Han, Ru Feng, Li-Hao Wu, Jie Cai, Tian-Ying Zeng, Hui Liu, Ya-Wen Zeng, Li Zhou, Zhong-Qun Zhan
{"title":"根据医疗器械进步协会/欧洲高血压学会/ISO 81060-2:2018协议,对JAMR F1701T(臂型)压力监测仪进行验证。","authors":"Jing Li, Li-Hong Han, Ru Feng, Li-Hao Wu, Jie Cai, Tian-Ying Zeng, Hui Liu, Ya-Wen Zeng, Li Zhou, Zhong-Qun Zhan","doi":"10.1097/MBP.0000000000000643","DOIUrl":null,"url":null,"abstract":"<p><p>To validate the JAMR F1701T (arm type) blood pressure (BP) monitor according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018). A total of 90 subjects (male 60 and female 30) were recruited to fulfill the criteria of the AAMI/ESH/ISO Universal Standard (the number, gender, age, limber size, and BP distribution), and sequential measurements of BP, including both SBP and DBP were obtained using the test device and the standard mercury sphygmomanometer. A total of 270 sets of comparison data (three sets of each subject) were obtained and analyzed. According to the validation criterion 1 of ISO 81060-2:2018, the mean ± SD of the differences between the JAMR F1701T and mercury sphygmomanometer BP (systolic/diastolic) readings was 2.06 ± 6.83/-4.84 ± 5.23 mmHg. For criterion 2, the SD of the averaged BP (systolic/diastolic) differences between the JAMR F1701 and reference BP (systolic/diastolic) per participant was 5.62/4.39 mmHg (the requirement was ≤6.43/5.01 mmHg by calculation). The JAMR F1701T met all the requirements of the ISO 81060-2:2018, and can be recommended for clinical and self/home use.</p>","PeriodicalId":1,"journal":{"name":"Accounts of Chemical Research","volume":null,"pages":null},"PeriodicalIF":16.4000,"publicationDate":"2023-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/6c/d2/bpmj-28-163.PMC10132453.pdf","citationCount":"0","resultStr":"{\"title\":\"Validation of the JAMR F1701T (arm type) pressure monitor according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/ISO 81060-2:2018 protocol.\",\"authors\":\"Jing Li, Li-Hong Han, Ru Feng, Li-Hao Wu, Jie Cai, Tian-Ying Zeng, Hui Liu, Ya-Wen Zeng, Li Zhou, Zhong-Qun Zhan\",\"doi\":\"10.1097/MBP.0000000000000643\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>To validate the JAMR F1701T (arm type) blood pressure (BP) monitor according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018). A total of 90 subjects (male 60 and female 30) were recruited to fulfill the criteria of the AAMI/ESH/ISO Universal Standard (the number, gender, age, limber size, and BP distribution), and sequential measurements of BP, including both SBP and DBP were obtained using the test device and the standard mercury sphygmomanometer. A total of 270 sets of comparison data (three sets of each subject) were obtained and analyzed. According to the validation criterion 1 of ISO 81060-2:2018, the mean ± SD of the differences between the JAMR F1701T and mercury sphygmomanometer BP (systolic/diastolic) readings was 2.06 ± 6.83/-4.84 ± 5.23 mmHg. For criterion 2, the SD of the averaged BP (systolic/diastolic) differences between the JAMR F1701 and reference BP (systolic/diastolic) per participant was 5.62/4.39 mmHg (the requirement was ≤6.43/5.01 mmHg by calculation). The JAMR F1701T met all the requirements of the ISO 81060-2:2018, and can be recommended for clinical and self/home use.</p>\",\"PeriodicalId\":1,\"journal\":{\"name\":\"Accounts of Chemical Research\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":16.4000,\"publicationDate\":\"2023-06-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/6c/d2/bpmj-28-163.PMC10132453.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Accounts of Chemical Research\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1097/MBP.0000000000000643\",\"RegionNum\":1,\"RegionCategory\":\"化学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"CHEMISTRY, MULTIDISCIPLINARY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Accounts of Chemical Research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000643","RegionNum":1,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CHEMISTRY, MULTIDISCIPLINARY","Score":null,"Total":0}
Validation of the JAMR F1701T (arm type) pressure monitor according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/ISO 81060-2:2018 protocol.
To validate the JAMR F1701T (arm type) blood pressure (BP) monitor according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018). A total of 90 subjects (male 60 and female 30) were recruited to fulfill the criteria of the AAMI/ESH/ISO Universal Standard (the number, gender, age, limber size, and BP distribution), and sequential measurements of BP, including both SBP and DBP were obtained using the test device and the standard mercury sphygmomanometer. A total of 270 sets of comparison data (three sets of each subject) were obtained and analyzed. According to the validation criterion 1 of ISO 81060-2:2018, the mean ± SD of the differences between the JAMR F1701T and mercury sphygmomanometer BP (systolic/diastolic) readings was 2.06 ± 6.83/-4.84 ± 5.23 mmHg. For criterion 2, the SD of the averaged BP (systolic/diastolic) differences between the JAMR F1701 and reference BP (systolic/diastolic) per participant was 5.62/4.39 mmHg (the requirement was ≤6.43/5.01 mmHg by calculation). The JAMR F1701T met all the requirements of the ISO 81060-2:2018, and can be recommended for clinical and self/home use.
期刊介绍:
Accounts of Chemical Research presents short, concise and critical articles offering easy-to-read overviews of basic research and applications in all areas of chemistry and biochemistry. These short reviews focus on research from the author’s own laboratory and are designed to teach the reader about a research project. In addition, Accounts of Chemical Research publishes commentaries that give an informed opinion on a current research problem. Special Issues online are devoted to a single topic of unusual activity and significance.
Accounts of Chemical Research replaces the traditional article abstract with an article "Conspectus." These entries synopsize the research affording the reader a closer look at the content and significance of an article. Through this provision of a more detailed description of the article contents, the Conspectus enhances the article's discoverability by search engines and the exposure for the research.