Linyi Li, Sen Bing, Kui Liu, Jun Jiang, Jieqiong Zhang, Changsheng Chen, Yi Wan
{"title":"示波式血压监测仪对房颤筛查的诊断准确性。","authors":"Linyi Li, Sen Bing, Kui Liu, Jun Jiang, Jieqiong Zhang, Changsheng Chen, Yi Wan","doi":"10.1097/MBP.0000000000000645","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Atrial fibrillation is the most common arrhythmia of clinical significance and hypertension is one of its major risk factors. This study aimed to evaluate the diagnostic accuracy of an automated oscillometric blood pressure (BP) monitor with the function of atrial fibrillation detection for atrial fibrillation screening.</p><p><strong>Materials and methods: </strong>Patients attending outpatient cardiology clinics were recruited for atrial fibrillation screening by the BP monitor with triplicate BP measurements for atrial fibrillation detection. Furthermore, a single-lead ECG was recorded simultaneously for comparison as the reference standard. The diagnostic test's evaluation index were analyzed, including sensitivity, specificity, and receiver operator characteristic (ROC) analysis.</p><p><strong>Results: </strong>A total of 295 participants were analyzed including 166 males and 129 females, with an average age of 72.5 ± 5.9 years. The sensitivity and specificity for atrial fibrillation detection by the device were 1.000 and 0.904, respectively, with the area under the ROC curve of 0.952 (95% confidence interval: 0.929-0.975, P < 0.001). Furthermore, the device had a Kappa-value of 0.781 ( P < 0.001) with the single-lead ECG in detecting atrial fibrillation.</p><p><strong>Conclusion: </strong>The automated oscillometric BP monitor (G.LAB MD41A0) with atrial fibrillation detection function has high sensitivity and specificity with good accuracy for atrial fibrillation screening, which could be used as a reliable screening tool for the early detection of atrial fibrillation with potential benefits.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":"28 3","pages":"144-148"},"PeriodicalIF":1.2000,"publicationDate":"2023-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Diagnostic accuracy of an oscillometric blood pressure monitor for atrial fibrillation screening.\",\"authors\":\"Linyi Li, Sen Bing, Kui Liu, Jun Jiang, Jieqiong Zhang, Changsheng Chen, Yi Wan\",\"doi\":\"10.1097/MBP.0000000000000645\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>Atrial fibrillation is the most common arrhythmia of clinical significance and hypertension is one of its major risk factors. This study aimed to evaluate the diagnostic accuracy of an automated oscillometric blood pressure (BP) monitor with the function of atrial fibrillation detection for atrial fibrillation screening.</p><p><strong>Materials and methods: </strong>Patients attending outpatient cardiology clinics were recruited for atrial fibrillation screening by the BP monitor with triplicate BP measurements for atrial fibrillation detection. Furthermore, a single-lead ECG was recorded simultaneously for comparison as the reference standard. The diagnostic test's evaluation index were analyzed, including sensitivity, specificity, and receiver operator characteristic (ROC) analysis.</p><p><strong>Results: </strong>A total of 295 participants were analyzed including 166 males and 129 females, with an average age of 72.5 ± 5.9 years. The sensitivity and specificity for atrial fibrillation detection by the device were 1.000 and 0.904, respectively, with the area under the ROC curve of 0.952 (95% confidence interval: 0.929-0.975, P < 0.001). Furthermore, the device had a Kappa-value of 0.781 ( P < 0.001) with the single-lead ECG in detecting atrial fibrillation.</p><p><strong>Conclusion: </strong>The automated oscillometric BP monitor (G.LAB MD41A0) with atrial fibrillation detection function has high sensitivity and specificity with good accuracy for atrial fibrillation screening, which could be used as a reliable screening tool for the early detection of atrial fibrillation with potential benefits.</p>\",\"PeriodicalId\":8950,\"journal\":{\"name\":\"Blood Pressure Monitoring\",\"volume\":\"28 3\",\"pages\":\"144-148\"},\"PeriodicalIF\":1.2000,\"publicationDate\":\"2023-06-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Blood Pressure Monitoring\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1097/MBP.0000000000000645\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"PERIPHERAL VASCULAR DISEASE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000645","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
Diagnostic accuracy of an oscillometric blood pressure monitor for atrial fibrillation screening.
Objective: Atrial fibrillation is the most common arrhythmia of clinical significance and hypertension is one of its major risk factors. This study aimed to evaluate the diagnostic accuracy of an automated oscillometric blood pressure (BP) monitor with the function of atrial fibrillation detection for atrial fibrillation screening.
Materials and methods: Patients attending outpatient cardiology clinics were recruited for atrial fibrillation screening by the BP monitor with triplicate BP measurements for atrial fibrillation detection. Furthermore, a single-lead ECG was recorded simultaneously for comparison as the reference standard. The diagnostic test's evaluation index were analyzed, including sensitivity, specificity, and receiver operator characteristic (ROC) analysis.
Results: A total of 295 participants were analyzed including 166 males and 129 females, with an average age of 72.5 ± 5.9 years. The sensitivity and specificity for atrial fibrillation detection by the device were 1.000 and 0.904, respectively, with the area under the ROC curve of 0.952 (95% confidence interval: 0.929-0.975, P < 0.001). Furthermore, the device had a Kappa-value of 0.781 ( P < 0.001) with the single-lead ECG in detecting atrial fibrillation.
Conclusion: The automated oscillometric BP monitor (G.LAB MD41A0) with atrial fibrillation detection function has high sensitivity and specificity with good accuracy for atrial fibrillation screening, which could be used as a reliable screening tool for the early detection of atrial fibrillation with potential benefits.
期刊介绍:
Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research.
This innovative journal contains papers dealing with all aspects of manual, automated, and ambulatory monitoring. Basic and clinical science papers are considered although the emphasis is on clinical medicine.
Submitted articles undergo a preliminary review by the editor. Some articles may be returned to authors without further consideration. Those being considered for publication will undergo further assessment and peer-review by the editors and those invited to do so from a reviewer pool.