鼻咽癌放射治疗质量控制实践中建立可重复性的统计过程控制工具

IF 0.4 4区 医学 Q4 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
S. Xu, Z. Li, J. Hu, J. Zhang
{"title":"鼻咽癌放射治疗质量控制实践中建立可重复性的统计过程控制工具","authors":"S. Xu, Z. Li, J. Hu, J. Zhang","doi":"10.52547/ijrr.20.2.7","DOIUrl":null,"url":null,"abstract":"Background : This study aims to perform quality control (QC) practices for setup reproducibility during radiotherapy for nasopharyngeal carcinoma (NPC) using statistical process control (SPC) tools. Materials and Methods : A total of 480 fractional images from 48 NPC patients with the first 10 fractions of the treatment were collected. In QC practices, setup errors were described using the histogram and normal curve, cumulative frequencies of absolute setup errors and 3D Euclidean Distance (E u ) were analyzed; the X ̅ - S chart and process capability index (Cpk) with the variable E u were utilized to identify whether the outlier occurred and to evaluate the QC process. Results : The translational setup error distributions were almost normal in Lateral, Longitudinal and Vertical directions and were narrower in Lateral and Vertical directions. Vertical translational errors and E u with a larger magnitude sag appeared the most frequently. Between the couch sag and no sag, the E u mean of 7 to 7 NPC patients with the same 3 patients was out of control and the standard deviation of E u of nil to 2 patients was outlier based on the X ̅ - S chart, and the Cpk was 1.05 and 1.36 respectively, when the specification limit of translational errors was ±3 mm. Conclusion : Daily imaging is necessary to increase setup reproducibility for NPC patients and more measures should be taken to facilitate quality assurance procedures. SPC is better applied to QC practices depending on the reliable data and the acceptable tolerance levels in","PeriodicalId":48815,"journal":{"name":"International Journal of Radiation Research","volume":"68 1","pages":""},"PeriodicalIF":0.4000,"publicationDate":"2022-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Statistical process control tools for setup reproducibility in quality control practices during nasopharyngeal carcinoma radiotherapy\",\"authors\":\"S. Xu, Z. Li, J. Hu, J. Zhang\",\"doi\":\"10.52547/ijrr.20.2.7\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Background : This study aims to perform quality control (QC) practices for setup reproducibility during radiotherapy for nasopharyngeal carcinoma (NPC) using statistical process control (SPC) tools. Materials and Methods : A total of 480 fractional images from 48 NPC patients with the first 10 fractions of the treatment were collected. In QC practices, setup errors were described using the histogram and normal curve, cumulative frequencies of absolute setup errors and 3D Euclidean Distance (E u ) were analyzed; the X ̅ - S chart and process capability index (Cpk) with the variable E u were utilized to identify whether the outlier occurred and to evaluate the QC process. Results : The translational setup error distributions were almost normal in Lateral, Longitudinal and Vertical directions and were narrower in Lateral and Vertical directions. Vertical translational errors and E u with a larger magnitude sag appeared the most frequently. Between the couch sag and no sag, the E u mean of 7 to 7 NPC patients with the same 3 patients was out of control and the standard deviation of E u of nil to 2 patients was outlier based on the X ̅ - S chart, and the Cpk was 1.05 and 1.36 respectively, when the specification limit of translational errors was ±3 mm. Conclusion : Daily imaging is necessary to increase setup reproducibility for NPC patients and more measures should be taken to facilitate quality assurance procedures. SPC is better applied to QC practices depending on the reliable data and the acceptable tolerance levels in\",\"PeriodicalId\":48815,\"journal\":{\"name\":\"International Journal of Radiation Research\",\"volume\":\"68 1\",\"pages\":\"\"},\"PeriodicalIF\":0.4000,\"publicationDate\":\"2022-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"International Journal of Radiation Research\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.52547/ijrr.20.2.7\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Radiation Research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.52547/ijrr.20.2.7","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

背景:本研究旨在利用统计过程控制(SPC)工具对鼻咽癌(NPC)放疗过程中设置的可重复性进行质量控制(QC)实践。材料与方法:收集48例鼻咽癌患者前10次治疗的480张分数图像。在质量控制实践中,用直方图和正态曲线描述设置误差,分析绝对设置误差的累积频率和三维欧氏距离(E u);利用X′S图和过程能力指数(Cpk)(变量E u)来识别异常值是否发生,并对QC过程进行评价。结果:横向、纵向和纵向的平移设置误差分布基本符合正态分布,横向和纵向的平移设置误差分布较窄。垂直平移误差和较大幅度凹陷的E - u出现频率最高。在沙发凹陷与无凹陷之间,同样3例患者的7 ~ 7例鼻咽癌患者的E均值失控,根据X′S图,0 ~ 2例患者的E均值标准差为异常值,当平移误差规范限为±3 mm时,Cpk分别为1.05和1.36。结论:日常影像学检查对提高鼻咽癌患者设置的可重复性是必要的,应采取更多的措施来促进质量保证程序。根据可靠的数据和可接受的公差水平,SPC更好地应用于QC实践
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Statistical process control tools for setup reproducibility in quality control practices during nasopharyngeal carcinoma radiotherapy
Background : This study aims to perform quality control (QC) practices for setup reproducibility during radiotherapy for nasopharyngeal carcinoma (NPC) using statistical process control (SPC) tools. Materials and Methods : A total of 480 fractional images from 48 NPC patients with the first 10 fractions of the treatment were collected. In QC practices, setup errors were described using the histogram and normal curve, cumulative frequencies of absolute setup errors and 3D Euclidean Distance (E u ) were analyzed; the X ̅ - S chart and process capability index (Cpk) with the variable E u were utilized to identify whether the outlier occurred and to evaluate the QC process. Results : The translational setup error distributions were almost normal in Lateral, Longitudinal and Vertical directions and were narrower in Lateral and Vertical directions. Vertical translational errors and E u with a larger magnitude sag appeared the most frequently. Between the couch sag and no sag, the E u mean of 7 to 7 NPC patients with the same 3 patients was out of control and the standard deviation of E u of nil to 2 patients was outlier based on the X ̅ - S chart, and the Cpk was 1.05 and 1.36 respectively, when the specification limit of translational errors was ±3 mm. Conclusion : Daily imaging is necessary to increase setup reproducibility for NPC patients and more measures should be taken to facilitate quality assurance procedures. SPC is better applied to QC practices depending on the reliable data and the acceptable tolerance levels in
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
International Journal of Radiation Research
International Journal of Radiation Research RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
1.10
自引率
33.30%
发文量
42
期刊介绍: International Journal of Radiation Research (IJRR) publishes original scientific research and clinical investigations related to radiation oncology, radiation biology, and Medical and health physics. The clinical studies submitted for publication include experimental studies of combined modality treatment, especially chemoradiotherapy approaches, and relevant innovations in hyperthermia, brachytherapy, high LET irradiation, nuclear medicine, dosimetry, tumor imaging, radiation treatment planning, radiosensitizers, and radioprotectors. All manuscripts must pass stringent peer-review and only papers that are rated of high scientific quality are accepted.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信