Paolo Palatini, Claudio Fania, Elisabetta Benetti, Francesca Saladini, Francesca Battista, Andrea Ermolao
{"title":"臂围在43到50厘米之间的人的血压测量装置验证研究的合适参考袖带标准","authors":"Paolo Palatini, Claudio Fania, Elisabetta Benetti, Francesca Saladini, Francesca Battista, Andrea Ermolao","doi":"10.1097/MBP.0000000000000630","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Recommendations about the dimensions of the reference cuff for device validations in people with arm size >42 cm are still unavailable. The aim of this study was to identify the criteria for an appropriate reference cuff for validation studies in people with upper arm circumference between 43 and 50 cm.</p><p><strong>Methods: </strong>In 20 adults with upper arm circumference between 43 and 50 cm (X-large group), 34 subjects with arm circumference between 37 and 42 cm and 78 subjects with arm circumference <37 cm cylindrical and tronco-conical cuffs were compared. In all participants, the pressure transmitted to the arm under the two cuffs was measured using a paper-thin pressure sensor.</p><p><strong>Results: </strong>In the X-large group, all participants had an arm slant angle <86.0°. In this group, the difference between the pressure detected on the arm surface with the sensor using the cylindrical versus the tronco-conical cuff (13.5 mmHg) was larger than in the group with an arm circumference of 37-to 42 cm and the group with a circumference <37 cm (3.7 and 0.6 mmHg, respectively, P < 0.001 versus both). In the whole sample, the between-cuff pressure difference was proportional to the conical shape of the arm ( P < 0.001).</p><p><strong>Conclusions: </strong>These data suggest that in people with arm size between 43 and 50 cm the reference cuff for validation studies should have a conical shape with an 84-85° slant angle. To comply with current guidelines, an 18.5 × 37.0 cm bladder should be used which would allow proper cuffing in the large majority of subjects.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":"28 1","pages":"67-72"},"PeriodicalIF":1.2000,"publicationDate":"2023-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Criteria for a suitable reference cuff for validation studies of blood pressure measuring devices in people with arm circumference between 43 and 50 cm.\",\"authors\":\"Paolo Palatini, Claudio Fania, Elisabetta Benetti, Francesca Saladini, Francesca Battista, Andrea Ermolao\",\"doi\":\"10.1097/MBP.0000000000000630\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>Recommendations about the dimensions of the reference cuff for device validations in people with arm size >42 cm are still unavailable. The aim of this study was to identify the criteria for an appropriate reference cuff for validation studies in people with upper arm circumference between 43 and 50 cm.</p><p><strong>Methods: </strong>In 20 adults with upper arm circumference between 43 and 50 cm (X-large group), 34 subjects with arm circumference between 37 and 42 cm and 78 subjects with arm circumference <37 cm cylindrical and tronco-conical cuffs were compared. In all participants, the pressure transmitted to the arm under the two cuffs was measured using a paper-thin pressure sensor.</p><p><strong>Results: </strong>In the X-large group, all participants had an arm slant angle <86.0°. In this group, the difference between the pressure detected on the arm surface with the sensor using the cylindrical versus the tronco-conical cuff (13.5 mmHg) was larger than in the group with an arm circumference of 37-to 42 cm and the group with a circumference <37 cm (3.7 and 0.6 mmHg, respectively, P < 0.001 versus both). In the whole sample, the between-cuff pressure difference was proportional to the conical shape of the arm ( P < 0.001).</p><p><strong>Conclusions: </strong>These data suggest that in people with arm size between 43 and 50 cm the reference cuff for validation studies should have a conical shape with an 84-85° slant angle. To comply with current guidelines, an 18.5 × 37.0 cm bladder should be used which would allow proper cuffing in the large majority of subjects.</p>\",\"PeriodicalId\":8950,\"journal\":{\"name\":\"Blood Pressure Monitoring\",\"volume\":\"28 1\",\"pages\":\"67-72\"},\"PeriodicalIF\":1.2000,\"publicationDate\":\"2023-02-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Blood Pressure Monitoring\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1097/MBP.0000000000000630\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"PERIPHERAL VASCULAR DISEASE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000630","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
Criteria for a suitable reference cuff for validation studies of blood pressure measuring devices in people with arm circumference between 43 and 50 cm.
Objective: Recommendations about the dimensions of the reference cuff for device validations in people with arm size >42 cm are still unavailable. The aim of this study was to identify the criteria for an appropriate reference cuff for validation studies in people with upper arm circumference between 43 and 50 cm.
Methods: In 20 adults with upper arm circumference between 43 and 50 cm (X-large group), 34 subjects with arm circumference between 37 and 42 cm and 78 subjects with arm circumference <37 cm cylindrical and tronco-conical cuffs were compared. In all participants, the pressure transmitted to the arm under the two cuffs was measured using a paper-thin pressure sensor.
Results: In the X-large group, all participants had an arm slant angle <86.0°. In this group, the difference between the pressure detected on the arm surface with the sensor using the cylindrical versus the tronco-conical cuff (13.5 mmHg) was larger than in the group with an arm circumference of 37-to 42 cm and the group with a circumference <37 cm (3.7 and 0.6 mmHg, respectively, P < 0.001 versus both). In the whole sample, the between-cuff pressure difference was proportional to the conical shape of the arm ( P < 0.001).
Conclusions: These data suggest that in people with arm size between 43 and 50 cm the reference cuff for validation studies should have a conical shape with an 84-85° slant angle. To comply with current guidelines, an 18.5 × 37.0 cm bladder should be used which would allow proper cuffing in the large majority of subjects.
期刊介绍:
Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research.
This innovative journal contains papers dealing with all aspects of manual, automated, and ambulatory monitoring. Basic and clinical science papers are considered although the emphasis is on clinical medicine.
Submitted articles undergo a preliminary review by the editor. Some articles may be returned to authors without further consideration. Those being considered for publication will undergo further assessment and peer-review by the editors and those invited to do so from a reviewer pool.