{"title":"cressmmicro™便携式地形仪的实验室评估:对临床研究的影响","authors":"M. Oldham, S. Plunkett, Michael T. Fisher, K. H. Shafer, M. Morton","doi":"10.2478/cttr-2014-0004","DOIUrl":null,"url":null,"abstract":"SUMMARY Performance of the CReSSmicro™ v.2.0.0 portable smoking topography device was evaluated in a controlled laboratory environment using a calibrated smoking machine to generate known, reproducible smoking topographies. Device performance was evaluated using unlit and lit cigarettes over a range of puff volumes (25 to 100 mL) at 2 s puff duration using a sine wave profile. The device missed 93% and 100% of the 25 mL/2 s puffs for unlit and lit cigarettes, respectively. The device underreported puff durations for the entire volume range, i.e., from 40 mL (0.9 ± 0.17 s) to 100 mL (1.6 ± 0.11 s) for both unlit and lit cigarettes. Puff volumes were under reported for lower volumes from 40 mL (25.6 ± 5.5 mL) to 70 mL (68.8 ± 3.7 mL) for both unlit and lit cigarettes and slightly over reported for higher volumes from 85 mL (89.4 ± 7.6 mL) to 100 mL (103.4 ± 9.8 mL) for lit cigarettes. Experimental results indicated that a 12.5-20mL/s (750-1,200 mL/min) flow threshold for puff recognition was responsible for the observed results. This study reinforces the importance of fully understanding device performance prior to performing clinical studies of human smoking topography. [Beitr. Tabakforsch. Int. 26 (2014) 19-25] ZUSAMMENFASSUNG Die Leistungsfähigkeit des mobilen Rauchtopographiegerätes CReSSmicro™ v.2.0.0 wurde in einer kontrollierten Laborumgebung untersucht. Dabei wurde eine kalibrierte Abrauchmaschine eingesetzt, um bekannte, reproduzierbare Rauchtopographien zu erzeugen. Die Maschinenleistung wurde mit Hilfe von angezündeten und nicht angezündeten Zigaretten mit unterschiedlichen Zugvolumina (25 bis 100 mL) mit einer Zugdauer von 2 s und einem Sinusprofil untersucht. Bei den nicht angezündeten bzw. angezündeten Zigaretten wurden jeweils 93 % bzw. 100 % der 25 mL/2 s- Züge vom Gerät nicht gezählt. Das Gerät wies in allen gewählten Volumina, d.h. von 40 mL (0,9 ± 0,17 s) bis 100 mL (1,6 ± 0,11 s), für sowohl nicht angezündete als auch angezündete Zigaretten eine zu geringe Zugdauer aus. Die Zugvolumina wurden bei den niedrigeren Volumina von 40 mL (25,6 ± 5,5 mL) bis 70 mL (68,8 ± 3,7 mL) für sowohl nicht angezündete als auch angezündete Zigaretten geringer als tatsächlich angegeben und bei den höheren Volumina von 85 mL (89,4 ± 7,6 mL) bis 100 mL (103,4 ± 9,8 mL) für angezündete Zigaretten etwas zu hoch angegeben. Die Versuchsergebnisse zeigten, dass ein Durchflussschwellenwert von 12,5-20 mL/s (750-1.200 mL/min) bei der Bestimmung der Züge für die beobachteten Ergebnisse verantwortlich war. Diese Studie unterstreicht, wie wichtig es ist, die Leistungsfähigkeit der Maschinen genau zu kennen, bevor klinische Studien zur Rauchtopographie beim Menschen durchgeführt werden können. [Beitr. Tabakforsch. Int. 26 (2014) 19-25] RESUME La performance de l’appareil portable de topographie du fumage CReSSmicro™ v.2.0.0 a été évaluée en milieu contrôlé dans un laboratoire, en utilisant une machine à fumer étalonnée pour générer des topographies du fumage reproductibles connues. La performance de l’appareil a été évaluée au moyen de cigarettes non-allumées et allumées dans une plage de volumes de bouffées (de 25 à 100 mL) pour une durée de bouffée de 2 s en utilisant un profil d’onde sinusoïdale. L’appareil s’est trompé pour 93% et 100% des bouffées de 25 mL/2 s avec les cigarettes nonallumées et allumées respectivement. L’appareil a sousestimé les durées de bouffées sur toute la plage de volumes, c’est-à-dire de 40 mL (0,9 ± 0,17 s) à 100 mL (1,6 ± 0,11 s), tant pour les cigarettes non-allumées que pour les cigarettes allumées. Les volumes de bouffées ont été sousestimés pour les volumes inférieurs de 40 mL (25,6 ± 5,5 mL) à 70 mL (68,8 ± 3,7 mL), à la fois pour les cigarettes non-allumées et allumées et ils ont été légèrement surestimés pour des volumes plus élevés de 85 mL (89,4 ± 7,6 mL) à 100 mL (103,4 ± 9,8 mL) pour les cigarettes allumées. Les résultats expérimentaux ont indiqué qu’un seuil de flux de 12,5-20 mL/s (750-1200 mL/min) pour la reconnaissance des bouffées était responsable des résultats observés. Cette étude confirme l’importance de parfaitement bien comprendre la performance de l’appareil avant d’effectuer des essais cliniques de topographie du fumage humain. [Beitr. Tabakforsch. Int. 26 (2014) 19-25]","PeriodicalId":35431,"journal":{"name":"Beitrage zur Tabakforschung International/ Contributions to Tobacco Research","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2014-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"6","resultStr":"{\"title\":\"Laboratory Evaluation of the CReSSmicro™ Portable Topography Device: Implications for Clinical Research\",\"authors\":\"M. Oldham, S. Plunkett, Michael T. Fisher, K. H. Shafer, M. Morton\",\"doi\":\"10.2478/cttr-2014-0004\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"SUMMARY Performance of the CReSSmicro™ v.2.0.0 portable smoking topography device was evaluated in a controlled laboratory environment using a calibrated smoking machine to generate known, reproducible smoking topographies. Device performance was evaluated using unlit and lit cigarettes over a range of puff volumes (25 to 100 mL) at 2 s puff duration using a sine wave profile. The device missed 93% and 100% of the 25 mL/2 s puffs for unlit and lit cigarettes, respectively. The device underreported puff durations for the entire volume range, i.e., from 40 mL (0.9 ± 0.17 s) to 100 mL (1.6 ± 0.11 s) for both unlit and lit cigarettes. Puff volumes were under reported for lower volumes from 40 mL (25.6 ± 5.5 mL) to 70 mL (68.8 ± 3.7 mL) for both unlit and lit cigarettes and slightly over reported for higher volumes from 85 mL (89.4 ± 7.6 mL) to 100 mL (103.4 ± 9.8 mL) for lit cigarettes. Experimental results indicated that a 12.5-20mL/s (750-1,200 mL/min) flow threshold for puff recognition was responsible for the observed results. This study reinforces the importance of fully understanding device performance prior to performing clinical studies of human smoking topography. [Beitr. Tabakforsch. Int. 26 (2014) 19-25] ZUSAMMENFASSUNG Die Leistungsfähigkeit des mobilen Rauchtopographiegerätes CReSSmicro™ v.2.0.0 wurde in einer kontrollierten Laborumgebung untersucht. Dabei wurde eine kalibrierte Abrauchmaschine eingesetzt, um bekannte, reproduzierbare Rauchtopographien zu erzeugen. Die Maschinenleistung wurde mit Hilfe von angezündeten und nicht angezündeten Zigaretten mit unterschiedlichen Zugvolumina (25 bis 100 mL) mit einer Zugdauer von 2 s und einem Sinusprofil untersucht. Bei den nicht angezündeten bzw. angezündeten Zigaretten wurden jeweils 93 % bzw. 100 % der 25 mL/2 s- Züge vom Gerät nicht gezählt. Das Gerät wies in allen gewählten Volumina, d.h. von 40 mL (0,9 ± 0,17 s) bis 100 mL (1,6 ± 0,11 s), für sowohl nicht angezündete als auch angezündete Zigaretten eine zu geringe Zugdauer aus. Die Zugvolumina wurden bei den niedrigeren Volumina von 40 mL (25,6 ± 5,5 mL) bis 70 mL (68,8 ± 3,7 mL) für sowohl nicht angezündete als auch angezündete Zigaretten geringer als tatsächlich angegeben und bei den höheren Volumina von 85 mL (89,4 ± 7,6 mL) bis 100 mL (103,4 ± 9,8 mL) für angezündete Zigaretten etwas zu hoch angegeben. Die Versuchsergebnisse zeigten, dass ein Durchflussschwellenwert von 12,5-20 mL/s (750-1.200 mL/min) bei der Bestimmung der Züge für die beobachteten Ergebnisse verantwortlich war. Diese Studie unterstreicht, wie wichtig es ist, die Leistungsfähigkeit der Maschinen genau zu kennen, bevor klinische Studien zur Rauchtopographie beim Menschen durchgeführt werden können. [Beitr. Tabakforsch. Int. 26 (2014) 19-25] RESUME La performance de l’appareil portable de topographie du fumage CReSSmicro™ v.2.0.0 a été évaluée en milieu contrôlé dans un laboratoire, en utilisant une machine à fumer étalonnée pour générer des topographies du fumage reproductibles connues. La performance de l’appareil a été évaluée au moyen de cigarettes non-allumées et allumées dans une plage de volumes de bouffées (de 25 à 100 mL) pour une durée de bouffée de 2 s en utilisant un profil d’onde sinusoïdale. L’appareil s’est trompé pour 93% et 100% des bouffées de 25 mL/2 s avec les cigarettes nonallumées et allumées respectivement. L’appareil a sousestimé les durées de bouffées sur toute la plage de volumes, c’est-à-dire de 40 mL (0,9 ± 0,17 s) à 100 mL (1,6 ± 0,11 s), tant pour les cigarettes non-allumées que pour les cigarettes allumées. Les volumes de bouffées ont été sousestimés pour les volumes inférieurs de 40 mL (25,6 ± 5,5 mL) à 70 mL (68,8 ± 3,7 mL), à la fois pour les cigarettes non-allumées et allumées et ils ont été légèrement surestimés pour des volumes plus élevés de 85 mL (89,4 ± 7,6 mL) à 100 mL (103,4 ± 9,8 mL) pour les cigarettes allumées. Les résultats expérimentaux ont indiqué qu’un seuil de flux de 12,5-20 mL/s (750-1200 mL/min) pour la reconnaissance des bouffées était responsable des résultats observés. Cette étude confirme l’importance de parfaitement bien comprendre la performance de l’appareil avant d’effectuer des essais cliniques de topographie du fumage humain. [Beitr. Tabakforsch. Int. 26 (2014) 19-25]\",\"PeriodicalId\":35431,\"journal\":{\"name\":\"Beitrage zur Tabakforschung International/ Contributions to Tobacco Research\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2014-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"6\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Beitrage zur Tabakforschung International/ Contributions to Tobacco Research\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.2478/cttr-2014-0004\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"Agricultural and Biological Sciences\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Beitrage zur Tabakforschung International/ Contributions to Tobacco Research","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2478/cttr-2014-0004","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Agricultural and Biological Sciences","Score":null,"Total":0}
引用次数: 6

摘要

cresmicroo™v.2.0.0便携式吸烟地形仪的性能在受控的实验室环境中使用校准的吸烟机进行评估,以生成已知的,可重复的吸烟地形。使用未点燃和点燃的香烟在吸烟量范围内(25至100毫升),在2秒的吸烟持续时间内使用正弦波剖面评估设备性能。对于未点燃的香烟和点燃的香烟,该装置分别错过了93%和100%的25毫升/2秒。该装置少报了整个体积范围内的吞吐持续时间,即从40毫升(0.9±0.17秒)到100毫升(1.6±0.11秒),无论是未点燃的还是点燃的香烟。未点燃和点燃香烟的吞云吐雾量均低于40 mL(25.6±5.5 mL)至70 mL(68.8±3.7 mL),点燃香烟的吞云吐雾量高于85 mL(89.4±7.6 mL)至100 mL(103.4±9.8 mL)。实验结果表明,烟雾识别的12.5-20mL/s (750- 1200ml /min)流量阈值是观察到的结果的原因。这项研究强调了在进行人体吸烟地形临床研究之前充分了解设备性能的重要性。[Beitr。Tabakforsch。[j] .机械工程学报,26 (2014):19-25 [j] . ZUSAMMENFASSUNG Die Leistungsfähigkeit des mobilen Rauchtopographiegerätes cresmicroro™v.2.0.0 wde in iner control . lierink . Laborumgebung unsuke。Dabei wurde einekalibrierte abauchmachine eingesetzt, um bekante, ereduziertographien zerzegen。Die Maschinenleistung wurde mit Hilfe von angez ndeten und nicht angez ndeten Zigaretten mit unterschiedlichen Zugvolumina (25 bis 100 mL) mit einer Zugdauer von 2 s and einem sinusprofilum untersucht。北京夜间天使 ndeten bzw。angez ndeten Zigaretten wurden珠宝93% bzw。100% der 25ml / 2s - z ge vom Gerät nicht gezählt。Das Gerät wies in allen gewählten Volumina, d.h. von 40 mL(0,9±0,17 s)和100 mL(1,6±0,11 s), fr sowhhl nicht angez ndete als auch angez ndete Zigaretten eine zu geringe Zugdauer aus。Die Zugvolumina wurden bei den niedrigeren Volumina von 40 mL(25,6±5,5 mL) bis 70 mL(68,8±3,7 mL) f r sowohl nicht angez ndete Zigaretten geringer als tatsächlich angegeben und bei den höheren Volumina von 85 mL(89,4±7,6 mL) bis 100 mL(103,4±9,8 mL) f r angez ndete Zigaretten etwas zu hoch angegeben。Die Versuchsergebnisse zeigten, dass ein Durchflussschwellenwert von 12,5-20 mL/s (750-1.200 mL/min) beder Bestimmung der zge f r Die beobachteten Ergebnisse verantwortlich war。德国医学研究,德国医学研究协会,德国医学研究协会,德国医学研究协会,德国医学研究协会,德国医学研究协会,德国医学研究协会,德国医学研究协会können。[Beitr。Tabakforsch。[j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j] [j]。拉性能de l 'appareil疾病平均evaluee盟德香烟non-allumees et allumees de卷在一个海滨de bouffees (25 100 mL)倒一个duree de bouffee de 2 s现在联合国的资料d 'onde sinusoidale。L’apapil’s从93%和100%的烟草中分别提取25ml / s的烟草,非烟草和烟草中提取25ml / s的烟草。L ' apareil a sousestimest . L ' apareil a sousestimest . L ' d40 mL(0.9±0.17 s) × 100 mL(1.6±0.11 s), L ' apareil a sousestimest . L ' apareil a sousestimest . L ' estest . o 40 mL(0.9±0.17 s), L ' apareil a sresresest . L ' apareil a sresresest .)里面的卷de bouffees高频sousestimes倒Les卷inferieurs 40 mL(25 6±5 5毫升)70毫升(68、8±3、7毫升),拉回倒Les香烟non-allumees et allumees等观察高频legerement surestimes倒des卷+ eleves 85 mL(89 4±7 6毫升)100毫升(103 4±9 8毫升)倒allumees Les香烟。5- 20ml /s (750- 1200ml /min), pour la reconnaissance des bouffsames, achsames, achsamys, achsamys, achsamys, achsamys, achsamys, achsamys, achsamys, achsamys。Cette的研究证实了在理解服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装、服装等方面的重要性。[Beitr。Tabakforsch。Int. 26 (2014) 19-25]
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Laboratory Evaluation of the CReSSmicro™ Portable Topography Device: Implications for Clinical Research
SUMMARY Performance of the CReSSmicro™ v.2.0.0 portable smoking topography device was evaluated in a controlled laboratory environment using a calibrated smoking machine to generate known, reproducible smoking topographies. Device performance was evaluated using unlit and lit cigarettes over a range of puff volumes (25 to 100 mL) at 2 s puff duration using a sine wave profile. The device missed 93% and 100% of the 25 mL/2 s puffs for unlit and lit cigarettes, respectively. The device underreported puff durations for the entire volume range, i.e., from 40 mL (0.9 ± 0.17 s) to 100 mL (1.6 ± 0.11 s) for both unlit and lit cigarettes. Puff volumes were under reported for lower volumes from 40 mL (25.6 ± 5.5 mL) to 70 mL (68.8 ± 3.7 mL) for both unlit and lit cigarettes and slightly over reported for higher volumes from 85 mL (89.4 ± 7.6 mL) to 100 mL (103.4 ± 9.8 mL) for lit cigarettes. Experimental results indicated that a 12.5-20mL/s (750-1,200 mL/min) flow threshold for puff recognition was responsible for the observed results. This study reinforces the importance of fully understanding device performance prior to performing clinical studies of human smoking topography. [Beitr. Tabakforsch. Int. 26 (2014) 19-25] ZUSAMMENFASSUNG Die Leistungsfähigkeit des mobilen Rauchtopographiegerätes CReSSmicro™ v.2.0.0 wurde in einer kontrollierten Laborumgebung untersucht. Dabei wurde eine kalibrierte Abrauchmaschine eingesetzt, um bekannte, reproduzierbare Rauchtopographien zu erzeugen. Die Maschinenleistung wurde mit Hilfe von angezündeten und nicht angezündeten Zigaretten mit unterschiedlichen Zugvolumina (25 bis 100 mL) mit einer Zugdauer von 2 s und einem Sinusprofil untersucht. Bei den nicht angezündeten bzw. angezündeten Zigaretten wurden jeweils 93 % bzw. 100 % der 25 mL/2 s- Züge vom Gerät nicht gezählt. Das Gerät wies in allen gewählten Volumina, d.h. von 40 mL (0,9 ± 0,17 s) bis 100 mL (1,6 ± 0,11 s), für sowohl nicht angezündete als auch angezündete Zigaretten eine zu geringe Zugdauer aus. Die Zugvolumina wurden bei den niedrigeren Volumina von 40 mL (25,6 ± 5,5 mL) bis 70 mL (68,8 ± 3,7 mL) für sowohl nicht angezündete als auch angezündete Zigaretten geringer als tatsächlich angegeben und bei den höheren Volumina von 85 mL (89,4 ± 7,6 mL) bis 100 mL (103,4 ± 9,8 mL) für angezündete Zigaretten etwas zu hoch angegeben. Die Versuchsergebnisse zeigten, dass ein Durchflussschwellenwert von 12,5-20 mL/s (750-1.200 mL/min) bei der Bestimmung der Züge für die beobachteten Ergebnisse verantwortlich war. Diese Studie unterstreicht, wie wichtig es ist, die Leistungsfähigkeit der Maschinen genau zu kennen, bevor klinische Studien zur Rauchtopographie beim Menschen durchgeführt werden können. [Beitr. Tabakforsch. Int. 26 (2014) 19-25] RESUME La performance de l’appareil portable de topographie du fumage CReSSmicro™ v.2.0.0 a été évaluée en milieu contrôlé dans un laboratoire, en utilisant une machine à fumer étalonnée pour générer des topographies du fumage reproductibles connues. La performance de l’appareil a été évaluée au moyen de cigarettes non-allumées et allumées dans une plage de volumes de bouffées (de 25 à 100 mL) pour une durée de bouffée de 2 s en utilisant un profil d’onde sinusoïdale. L’appareil s’est trompé pour 93% et 100% des bouffées de 25 mL/2 s avec les cigarettes nonallumées et allumées respectivement. L’appareil a sousestimé les durées de bouffées sur toute la plage de volumes, c’est-à-dire de 40 mL (0,9 ± 0,17 s) à 100 mL (1,6 ± 0,11 s), tant pour les cigarettes non-allumées que pour les cigarettes allumées. Les volumes de bouffées ont été sousestimés pour les volumes inférieurs de 40 mL (25,6 ± 5,5 mL) à 70 mL (68,8 ± 3,7 mL), à la fois pour les cigarettes non-allumées et allumées et ils ont été légèrement surestimés pour des volumes plus élevés de 85 mL (89,4 ± 7,6 mL) à 100 mL (103,4 ± 9,8 mL) pour les cigarettes allumées. Les résultats expérimentaux ont indiqué qu’un seuil de flux de 12,5-20 mL/s (750-1200 mL/min) pour la reconnaissance des bouffées était responsable des résultats observés. Cette étude confirme l’importance de parfaitement bien comprendre la performance de l’appareil avant d’effectuer des essais cliniques de topographie du fumage humain. [Beitr. Tabakforsch. Int. 26 (2014) 19-25]
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Beitrage zur Tabakforschung International/ Contributions to Tobacco Research
Beitrage zur Tabakforschung International/ Contributions to Tobacco Research Agricultural and Biological Sciences-Agronomy and Crop Science
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信