用于视网膜血流诊断的生物散斑流量计

Yoshihisa Aizu, Toshimitsu Asakura, Kouji Ogino, Toshiaki Sugita, Yasuyuki Suzuki, Kanjiro Masuda
{"title":"用于视网膜血流诊断的生物散斑流量计","authors":"Yoshihisa Aizu,&nbsp;Toshimitsu Asakura,&nbsp;Kouji Ogino,&nbsp;Toshiaki Sugita,&nbsp;Yasuyuki Suzuki,&nbsp;Kanjiro Masuda","doi":"10.1002/1361-6374(199612)4:4<254::AID-BIO4>3.0.CO;2-7","DOIUrl":null,"url":null,"abstract":"<p>Bio-speckle flowmetry for measuring the retinal blood flow velocity is described. The measuring principle is briefly discussed in comparison with the laser Doppler method. The basic properties of the photon correlation measurements including reproducibility were experimentally investigated with a rotating ground glass disk and for the normal human retina. The error was estimated to be less than 20% for <i>in vivo</i> measurements. By using a glass capillary model, the reciprocal of correlation time was calibrated to the mean flow velocity with a consideration of the effects of the vessel diameter and the background reflectance. The blood flow volume rate in the human retina was estimated by using the calibrated velocity and the vessel diameter. The results compared well with those reported in the literature, and show the usefulness of this flowmetry for clinical diagnostic purposes.</p>","PeriodicalId":100176,"journal":{"name":"Bioimaging","volume":"4 4","pages":"254-267"},"PeriodicalIF":0.0000,"publicationDate":"1996-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1002/1361-6374(199612)4:4<254::AID-BIO4>3.0.CO;2-7","citationCount":"19","resultStr":"{\"title\":\"Bio-speckle flowmetry for retinal blood flow diagnostics\",\"authors\":\"Yoshihisa Aizu,&nbsp;Toshimitsu Asakura,&nbsp;Kouji Ogino,&nbsp;Toshiaki Sugita,&nbsp;Yasuyuki Suzuki,&nbsp;Kanjiro Masuda\",\"doi\":\"10.1002/1361-6374(199612)4:4<254::AID-BIO4>3.0.CO;2-7\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p>Bio-speckle flowmetry for measuring the retinal blood flow velocity is described. The measuring principle is briefly discussed in comparison with the laser Doppler method. The basic properties of the photon correlation measurements including reproducibility were experimentally investigated with a rotating ground glass disk and for the normal human retina. The error was estimated to be less than 20% for <i>in vivo</i> measurements. By using a glass capillary model, the reciprocal of correlation time was calibrated to the mean flow velocity with a consideration of the effects of the vessel diameter and the background reflectance. The blood flow volume rate in the human retina was estimated by using the calibrated velocity and the vessel diameter. The results compared well with those reported in the literature, and show the usefulness of this flowmetry for clinical diagnostic purposes.</p>\",\"PeriodicalId\":100176,\"journal\":{\"name\":\"Bioimaging\",\"volume\":\"4 4\",\"pages\":\"254-267\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1996-12-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1002/1361-6374(199612)4:4<254::AID-BIO4>3.0.CO;2-7\",\"citationCount\":\"19\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Bioimaging\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1002/1361-6374%28199612%294%3A4%3C254%3A%3AAID-BIO4%3E3.0.CO%3B2-7\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Bioimaging","FirstCategoryId":"1085","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/1361-6374%28199612%294%3A4%3C254%3A%3AAID-BIO4%3E3.0.CO%3B2-7","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 19

摘要

描述了用于测量视网膜血流速度的生物散斑流量计。通过与激光多普勒法的比较,简要讨论了测量原理。光子相关性测量的基本特性,包括再现性,是用旋转的磨玻璃圆盘和正常人视网膜进行实验研究的。体内测量的误差估计小于20%。通过使用玻璃毛细管模型,考虑到血管直径和背景反射率的影响,将相关时间的倒数校准为平均流速。通过使用校准的速度和血管直径来估计人类视网膜中的血流量。结果与文献中报道的结果进行了很好的比较,并表明了这种流量计在临床诊断中的有用性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Bio-speckle flowmetry for retinal blood flow diagnostics

Bio-speckle flowmetry for measuring the retinal blood flow velocity is described. The measuring principle is briefly discussed in comparison with the laser Doppler method. The basic properties of the photon correlation measurements including reproducibility were experimentally investigated with a rotating ground glass disk and for the normal human retina. The error was estimated to be less than 20% for in vivo measurements. By using a glass capillary model, the reciprocal of correlation time was calibrated to the mean flow velocity with a consideration of the effects of the vessel diameter and the background reflectance. The blood flow volume rate in the human retina was estimated by using the calibrated velocity and the vessel diameter. The results compared well with those reported in the literature, and show the usefulness of this flowmetry for clinical diagnostic purposes.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信