Brent A Lanting, Olawale A Sogbein, Steven J MacDonald, Nirmit Shah, Tea-Lyn Kok, Ryan Willing, Matthew G Teeter
{"title":"一系列完整的全髋关节置换术股骨柄中耳轴损伤的量化先前被确定为具有灾难性失败风险。","authors":"Brent A Lanting, Olawale A Sogbein, Steven J MacDonald, Nirmit Shah, Tea-Lyn Kok, Ryan Willing, Matthew G Teeter","doi":"10.1177/11207000231199941","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Corrosion at the head-neck junction of femoral stems is a rare complication of total hip arthroplasty (THA) with manifestations ranging from subclinical wear to failure. Prior studies have identified a single femoral component design with an increased propensity for catastrophic trunnion failure. The purpose of the present study was to quantify trunnion damage of this femoral component retrieved from patients undergoing revision THA for non-trunnionosis indications.</p><p><strong>Methods: </strong>24 femoral components from a single manufacturer were identified for study inclusion. Each prosthesis underwent stereomicroscopic inspection. Corrosion and fretting scores were assigned per the Goldberg criteria to quadrants of the trunnion. Material loss was calculated based on cone angles across trunnion quadrants. This was carried out using a coordinate measuring machine that digitised each trunnion surface. Stems were compared to a series of femoral stems with the same trunnion design.</p><p><strong>Results: </strong>20 of the 24 (83%) trunnions demonstrated corrosion, all 24 trunnions demonstrated fretting. Corrosion scores did not statistically differ with respect to trunnion zone (<i>p</i> = 0.53), while fretting scores were higher in the inferior compared to the superior zones (<i>p</i> <i><</i> 0.001). There was no significant difference in cone angles assessing material loss between stems (<i>p</i> = 0.25).</p><p><strong>Conclusions: </strong>Evidence of trunnion damage was observed in each stem retrieved for non-trunnionosis revision. Fretting occurred more frequently about the inferior quadrants. However, digitised trunnion shapes were similar between compared stems exhibiting no material loss. Therefore, it is possible that previous reports of trunnion failures for this implant are not a systemic issue, and that further investigation is required.</p>","PeriodicalId":12911,"journal":{"name":"HIP International","volume":null,"pages":null},"PeriodicalIF":1.3000,"publicationDate":"2024-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Quantification of trunnion damage in a series of intact total hip arthroplasty femoral stems previously identified to be at risk of catastrophic failure.\",\"authors\":\"Brent A Lanting, Olawale A Sogbein, Steven J MacDonald, Nirmit Shah, Tea-Lyn Kok, Ryan Willing, Matthew G Teeter\",\"doi\":\"10.1177/11207000231199941\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Corrosion at the head-neck junction of femoral stems is a rare complication of total hip arthroplasty (THA) with manifestations ranging from subclinical wear to failure. Prior studies have identified a single femoral component design with an increased propensity for catastrophic trunnion failure. The purpose of the present study was to quantify trunnion damage of this femoral component retrieved from patients undergoing revision THA for non-trunnionosis indications.</p><p><strong>Methods: </strong>24 femoral components from a single manufacturer were identified for study inclusion. Each prosthesis underwent stereomicroscopic inspection. Corrosion and fretting scores were assigned per the Goldberg criteria to quadrants of the trunnion. Material loss was calculated based on cone angles across trunnion quadrants. This was carried out using a coordinate measuring machine that digitised each trunnion surface. Stems were compared to a series of femoral stems with the same trunnion design.</p><p><strong>Results: </strong>20 of the 24 (83%) trunnions demonstrated corrosion, all 24 trunnions demonstrated fretting. Corrosion scores did not statistically differ with respect to trunnion zone (<i>p</i> = 0.53), while fretting scores were higher in the inferior compared to the superior zones (<i>p</i> <i><</i> 0.001). There was no significant difference in cone angles assessing material loss between stems (<i>p</i> = 0.25).</p><p><strong>Conclusions: </strong>Evidence of trunnion damage was observed in each stem retrieved for non-trunnionosis revision. Fretting occurred more frequently about the inferior quadrants. However, digitised trunnion shapes were similar between compared stems exhibiting no material loss. Therefore, it is possible that previous reports of trunnion failures for this implant are not a systemic issue, and that further investigation is required.</p>\",\"PeriodicalId\":12911,\"journal\":{\"name\":\"HIP International\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2024-05-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"HIP International\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1177/11207000231199941\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2023/10/2 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q3\",\"JCRName\":\"ORTHOPEDICS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"HIP International","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/11207000231199941","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/10/2 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
Quantification of trunnion damage in a series of intact total hip arthroplasty femoral stems previously identified to be at risk of catastrophic failure.
Background: Corrosion at the head-neck junction of femoral stems is a rare complication of total hip arthroplasty (THA) with manifestations ranging from subclinical wear to failure. Prior studies have identified a single femoral component design with an increased propensity for catastrophic trunnion failure. The purpose of the present study was to quantify trunnion damage of this femoral component retrieved from patients undergoing revision THA for non-trunnionosis indications.
Methods: 24 femoral components from a single manufacturer were identified for study inclusion. Each prosthesis underwent stereomicroscopic inspection. Corrosion and fretting scores were assigned per the Goldberg criteria to quadrants of the trunnion. Material loss was calculated based on cone angles across trunnion quadrants. This was carried out using a coordinate measuring machine that digitised each trunnion surface. Stems were compared to a series of femoral stems with the same trunnion design.
Results: 20 of the 24 (83%) trunnions demonstrated corrosion, all 24 trunnions demonstrated fretting. Corrosion scores did not statistically differ with respect to trunnion zone (p = 0.53), while fretting scores were higher in the inferior compared to the superior zones (p< 0.001). There was no significant difference in cone angles assessing material loss between stems (p = 0.25).
Conclusions: Evidence of trunnion damage was observed in each stem retrieved for non-trunnionosis revision. Fretting occurred more frequently about the inferior quadrants. However, digitised trunnion shapes were similar between compared stems exhibiting no material loss. Therefore, it is possible that previous reports of trunnion failures for this implant are not a systemic issue, and that further investigation is required.
期刊介绍:
HIP International is the official journal of the European Hip Society. It is the only international, peer-reviewed, bi-monthly journal dedicated to diseases of the hip. HIP International considers contributions relating to hip surgery, traumatology of the hip, prosthetic surgery, biomechanics, and basic sciences relating to the hip. HIP International invites reviews from leading specialists with the aim of informing its readers of current evidence-based best practice.
The journal also publishes supplements containing proceedings of symposia, special meetings or articles of special educational merit.
HIP International is divided into six independent sections led by editors of the highest scientific merit. These sections are:
• Biomaterials
• Biomechanics
• Conservative Hip Surgery
• Paediatrics
• Primary and Revision Hip Arthroplasty
• Traumatology