{"title":"双波长视网膜血氧计中血管反射和多条光路的潜在测量误差。","authors":"James M. Beach, Benjamin Shoemaker, George J. Eckert, Alon Harris, Brent Siesky, Julia C. Arciero","doi":"10.1111/aos.15776","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Purpose</h3>\n \n <p>This study aims to characterize the dependence of measured retinal arterial and venous saturation on vessel diameter and central reflex in retinal oximetry, with an ultimate goal of identifying potential causes and suggesting approaches to improve measurement accuracy.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>In 10 subjects, oxygen saturation, vessel diameter and optical density are obtained using Oxymap Analyzer software without diameter correction. Diameter dependence of saturation is characterized using linear regression between measured values of saturation and diameter. Occurrences of negative values of vessel optical densities (ODs) associated with central vessel reflex are acquired from Oxymap Analyzer. A conceptual model is used to calculate the ratio of optical densities (ODRs) according to retinal reflectance properties and single and double-pass light transmission across fixed path lengths. Model-predicted values are compared with measured oximetry values at different vessel diameters.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Venous saturation shows an inverse relationship with vessel diameter (D) across subjects, with a mean slope of −0.180 (SE = 0.022) %/μm (20 < D < 180 μm) and a more rapid saturation increase at small vessel diameters reaching to over 80%. Arterial saturation yields smaller positive and negative slopes in individual subjects, with an average of −0.007 (SE = 0.021) %/μm (20 < D < 200 μm) across all subjects. Measurements where vessel brightness exceeds that of the retinal background result in negative values of optical density, causing an artifactual increase in saturation. Optimization of model reflectance values produces a good fit of the conceptual model to measured ODRs.</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>Measurement artefacts in retinal oximetry are caused by strong central vessel reflections, and apparent diameter sensitivity may result from single and double-pass transmission in vessels. Improvement in correction for vessel diameter is indicated for arteries however further study is necessary for venous corrections.</p>\n </section>\n </div>","PeriodicalId":6915,"journal":{"name":"Acta Ophthalmologica","volume":"102 3","pages":"e367-e380"},"PeriodicalIF":3.0000,"publicationDate":"2023-10-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1111/aos.15776","citationCount":"0","resultStr":"{\"title\":\"Potential measurement error from vessel reflex and multiple light paths in dual-wavelength retinal oximetry\",\"authors\":\"James M. Beach, Benjamin Shoemaker, George J. Eckert, Alon Harris, Brent Siesky, Julia C. Arciero\",\"doi\":\"10.1111/aos.15776\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div>\\n \\n \\n <section>\\n \\n <h3> Purpose</h3>\\n \\n <p>This study aims to characterize the dependence of measured retinal arterial and venous saturation on vessel diameter and central reflex in retinal oximetry, with an ultimate goal of identifying potential causes and suggesting approaches to improve measurement accuracy.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Methods</h3>\\n \\n <p>In 10 subjects, oxygen saturation, vessel diameter and optical density are obtained using Oxymap Analyzer software without diameter correction. Diameter dependence of saturation is characterized using linear regression between measured values of saturation and diameter. Occurrences of negative values of vessel optical densities (ODs) associated with central vessel reflex are acquired from Oxymap Analyzer. A conceptual model is used to calculate the ratio of optical densities (ODRs) according to retinal reflectance properties and single and double-pass light transmission across fixed path lengths. Model-predicted values are compared with measured oximetry values at different vessel diameters.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Results</h3>\\n \\n <p>Venous saturation shows an inverse relationship with vessel diameter (D) across subjects, with a mean slope of −0.180 (SE = 0.022) %/μm (20 < D < 180 μm) and a more rapid saturation increase at small vessel diameters reaching to over 80%. Arterial saturation yields smaller positive and negative slopes in individual subjects, with an average of −0.007 (SE = 0.021) %/μm (20 < D < 200 μm) across all subjects. Measurements where vessel brightness exceeds that of the retinal background result in negative values of optical density, causing an artifactual increase in saturation. Optimization of model reflectance values produces a good fit of the conceptual model to measured ODRs.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Conclusion</h3>\\n \\n <p>Measurement artefacts in retinal oximetry are caused by strong central vessel reflections, and apparent diameter sensitivity may result from single and double-pass transmission in vessels. Improvement in correction for vessel diameter is indicated for arteries however further study is necessary for venous corrections.</p>\\n </section>\\n </div>\",\"PeriodicalId\":6915,\"journal\":{\"name\":\"Acta Ophthalmologica\",\"volume\":\"102 3\",\"pages\":\"e367-e380\"},\"PeriodicalIF\":3.0000,\"publicationDate\":\"2023-10-03\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://onlinelibrary.wiley.com/doi/epdf/10.1111/aos.15776\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Acta Ophthalmologica\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1111/aos.15776\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"OPHTHALMOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta Ophthalmologica","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/aos.15776","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
Potential measurement error from vessel reflex and multiple light paths in dual-wavelength retinal oximetry
Purpose
This study aims to characterize the dependence of measured retinal arterial and venous saturation on vessel diameter and central reflex in retinal oximetry, with an ultimate goal of identifying potential causes and suggesting approaches to improve measurement accuracy.
Methods
In 10 subjects, oxygen saturation, vessel diameter and optical density are obtained using Oxymap Analyzer software without diameter correction. Diameter dependence of saturation is characterized using linear regression between measured values of saturation and diameter. Occurrences of negative values of vessel optical densities (ODs) associated with central vessel reflex are acquired from Oxymap Analyzer. A conceptual model is used to calculate the ratio of optical densities (ODRs) according to retinal reflectance properties and single and double-pass light transmission across fixed path lengths. Model-predicted values are compared with measured oximetry values at different vessel diameters.
Results
Venous saturation shows an inverse relationship with vessel diameter (D) across subjects, with a mean slope of −0.180 (SE = 0.022) %/μm (20 < D < 180 μm) and a more rapid saturation increase at small vessel diameters reaching to over 80%. Arterial saturation yields smaller positive and negative slopes in individual subjects, with an average of −0.007 (SE = 0.021) %/μm (20 < D < 200 μm) across all subjects. Measurements where vessel brightness exceeds that of the retinal background result in negative values of optical density, causing an artifactual increase in saturation. Optimization of model reflectance values produces a good fit of the conceptual model to measured ODRs.
Conclusion
Measurement artefacts in retinal oximetry are caused by strong central vessel reflections, and apparent diameter sensitivity may result from single and double-pass transmission in vessels. Improvement in correction for vessel diameter is indicated for arteries however further study is necessary for venous corrections.
期刊介绍:
Acta Ophthalmologica is published on behalf of the Acta Ophthalmologica Scandinavica Foundation and is the official scientific publication of the following societies: The Danish Ophthalmological Society, The Finnish Ophthalmological Society, The Icelandic Ophthalmological Society, The Norwegian Ophthalmological Society and The Swedish Ophthalmological Society, and also the European Association for Vision and Eye Research (EVER).