{"title":"通过流式细胞术进行人类白细胞抗原-B27 分型:三种 CE-IVD 方法的比较","authors":"Louis Waeckel, Anne Kennel, Anne-Emmanuelle Berger, Claude Lambert","doi":"10.1002/cyto.b.22102","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>The human leukocyte antigen B27 (HLA-B27), associated with chronic inflammatory diseases is thus widely performed for diagnostic purposes. Genotyping by molecular biology is the current gold standard for HLA-B27 typing, but cheaper and faster flow cytometry methods have been developed.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>In this report, we compare analytical performances of three CE-IVD flow cytometry kits: IOTest™ and DuraClone B27™ from Beckman Coulter and BD HLA-B27™ from BD Biosciences on a Navios™ cytometer as compared to molecular biology.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Routine analyses could conclude for HLA-B27 in197 patients (23.2%) and was not conclusive for 66 patients (7%, 8%). The experience revealed the needs to complete IOTest™ with a lineage marker (like CD3-APC) and a standardization procedure in detection of fluorescence. Comparative analysis considering 23/44 non-conclusive samples as negative, pointed out a 100% sensitivity and specificity of IOTest™ in determining genetically proved HLA-B27. Specificity of the BD HLA-B27<sup>TM</sup> kit was 94% (two false positives) sticking to the manufacturer instructions but raised to 100% and 94% sensitivity with a cutoff at 8.5 (225 on FACSdiva's scale).</p>\n </section>\n \n <section>\n \n <h3> Discussion</h3>\n \n <p>The DuraClone B27™ specificity was poor using the manufacturer cutoff but raised to 100% with a 8.0 cutoff instead of 15.</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>The three flow cytometry kits displayed satisfying performances but need adjustments. The DuraClone B27™ kit seems to be the best while the IOTest™ kit is not conclusive in 8% of cases. In most cases the use of molecular biology can be spared, but genotyping remains essential for patients whose HLA-B27 status cannot be determined with confidence by flow cytometry.</p>\n </section>\n </div>","PeriodicalId":2,"journal":{"name":"ACS Applied Bio Materials","volume":null,"pages":null},"PeriodicalIF":4.6000,"publicationDate":"2022-11-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/cyto.b.22102","citationCount":"1","resultStr":"{\"title\":\"Human leukocyte antigen-B27 typing by flow cytometry: Comparison of three CE-IVD methods\",\"authors\":\"Louis Waeckel, Anne Kennel, Anne-Emmanuelle Berger, Claude Lambert\",\"doi\":\"10.1002/cyto.b.22102\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div>\\n \\n \\n <section>\\n \\n <h3> Background</h3>\\n \\n <p>The human leukocyte antigen B27 (HLA-B27), associated with chronic inflammatory diseases is thus widely performed for diagnostic purposes. Genotyping by molecular biology is the current gold standard for HLA-B27 typing, but cheaper and faster flow cytometry methods have been developed.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Methods</h3>\\n \\n <p>In this report, we compare analytical performances of three CE-IVD flow cytometry kits: IOTest™ and DuraClone B27™ from Beckman Coulter and BD HLA-B27™ from BD Biosciences on a Navios™ cytometer as compared to molecular biology.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Results</h3>\\n \\n <p>Routine analyses could conclude for HLA-B27 in197 patients (23.2%) and was not conclusive for 66 patients (7%, 8%). The experience revealed the needs to complete IOTest™ with a lineage marker (like CD3-APC) and a standardization procedure in detection of fluorescence. Comparative analysis considering 23/44 non-conclusive samples as negative, pointed out a 100% sensitivity and specificity of IOTest™ in determining genetically proved HLA-B27. Specificity of the BD HLA-B27<sup>TM</sup> kit was 94% (two false positives) sticking to the manufacturer instructions but raised to 100% and 94% sensitivity with a cutoff at 8.5 (225 on FACSdiva's scale).</p>\\n </section>\\n \\n <section>\\n \\n <h3> Discussion</h3>\\n \\n <p>The DuraClone B27™ specificity was poor using the manufacturer cutoff but raised to 100% with a 8.0 cutoff instead of 15.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Conclusion</h3>\\n \\n <p>The three flow cytometry kits displayed satisfying performances but need adjustments. The DuraClone B27™ kit seems to be the best while the IOTest™ kit is not conclusive in 8% of cases. In most cases the use of molecular biology can be spared, but genotyping remains essential for patients whose HLA-B27 status cannot be determined with confidence by flow cytometry.</p>\\n </section>\\n </div>\",\"PeriodicalId\":2,\"journal\":{\"name\":\"ACS Applied Bio Materials\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":4.6000,\"publicationDate\":\"2022-11-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://onlinelibrary.wiley.com/doi/epdf/10.1002/cyto.b.22102\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"ACS Applied Bio Materials\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1002/cyto.b.22102\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"MATERIALS SCIENCE, BIOMATERIALS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"ACS Applied Bio Materials","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/cyto.b.22102","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MATERIALS SCIENCE, BIOMATERIALS","Score":null,"Total":0}
Human leukocyte antigen-B27 typing by flow cytometry: Comparison of three CE-IVD methods
Background
The human leukocyte antigen B27 (HLA-B27), associated with chronic inflammatory diseases is thus widely performed for diagnostic purposes. Genotyping by molecular biology is the current gold standard for HLA-B27 typing, but cheaper and faster flow cytometry methods have been developed.
Methods
In this report, we compare analytical performances of three CE-IVD flow cytometry kits: IOTest™ and DuraClone B27™ from Beckman Coulter and BD HLA-B27™ from BD Biosciences on a Navios™ cytometer as compared to molecular biology.
Results
Routine analyses could conclude for HLA-B27 in197 patients (23.2%) and was not conclusive for 66 patients (7%, 8%). The experience revealed the needs to complete IOTest™ with a lineage marker (like CD3-APC) and a standardization procedure in detection of fluorescence. Comparative analysis considering 23/44 non-conclusive samples as negative, pointed out a 100% sensitivity and specificity of IOTest™ in determining genetically proved HLA-B27. Specificity of the BD HLA-B27TM kit was 94% (two false positives) sticking to the manufacturer instructions but raised to 100% and 94% sensitivity with a cutoff at 8.5 (225 on FACSdiva's scale).
Discussion
The DuraClone B27™ specificity was poor using the manufacturer cutoff but raised to 100% with a 8.0 cutoff instead of 15.
Conclusion
The three flow cytometry kits displayed satisfying performances but need adjustments. The DuraClone B27™ kit seems to be the best while the IOTest™ kit is not conclusive in 8% of cases. In most cases the use of molecular biology can be spared, but genotyping remains essential for patients whose HLA-B27 status cannot be determined with confidence by flow cytometry.