Michael A McGeehan, Peter G Adamczyk, Kieran M Nichols, Michael E Hahn
{"title":"基于仿真的可变义肢刚度对义肢窝与残肢界面动力学影响分析。","authors":"Michael A McGeehan, Peter G Adamczyk, Kieran M Nichols, Michael E Hahn","doi":"10.1177/20556683221111986","DOIUrl":null,"url":null,"abstract":"<p><p><b>Introduction:</b> Loading of a residual limb within a prosthetic socket can cause tissue damage such as ulceration. Computational simulations may be useful tools for estimating tissue loading within the socket, and thus provide insights into how prosthesis designs affect residual limb-socket interface dynamics. The purpose of this study was to model and simulate residual limb-socket interface dynamics and evaluate the effects of varied prosthesis stiffness on interface dynamics during gait. <b>Methods:</b> A spatial contact model of a residual limb-socket interface was developed and integrated into a gait model with a below-knee amputation. Gait trials were simulated for four subjects walking with low, medium, and high prosthesis stiffness settings. The effects of prosthesis stiffness on interface kinematics, normal pressure, and shear stresses were evaluated. <b>Results:</b> Model-predicted values were similar to those reported previously in sensor-based experiments; increased stiffness resulted in greater average normal pressure and shear stress (<i>p</i> < 0.05). <b>Conclusions:</b> These methods may be useful to aid experimental studies by providing insights into the effects of varied prosthesis design parameters or gait conditions on residual limb-socket interface dynamics. The current results suggest that these effects may be subject-specific.</p>","PeriodicalId":2,"journal":{"name":"ACS Applied Bio Materials","volume":" ","pages":"20556683221111986"},"PeriodicalIF":4.6000,"publicationDate":"2022-07-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/c3/98/10.1177_20556683221111986.PMC9289901.pdf","citationCount":"0","resultStr":"{\"title\":\"A simulation-based analysis of the effects of variable prosthesis stiffness on interface dynamics between the prosthetic socket and residual limb.\",\"authors\":\"Michael A McGeehan, Peter G Adamczyk, Kieran M Nichols, Michael E Hahn\",\"doi\":\"10.1177/20556683221111986\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p><b>Introduction:</b> Loading of a residual limb within a prosthetic socket can cause tissue damage such as ulceration. Computational simulations may be useful tools for estimating tissue loading within the socket, and thus provide insights into how prosthesis designs affect residual limb-socket interface dynamics. The purpose of this study was to model and simulate residual limb-socket interface dynamics and evaluate the effects of varied prosthesis stiffness on interface dynamics during gait. <b>Methods:</b> A spatial contact model of a residual limb-socket interface was developed and integrated into a gait model with a below-knee amputation. Gait trials were simulated for four subjects walking with low, medium, and high prosthesis stiffness settings. The effects of prosthesis stiffness on interface kinematics, normal pressure, and shear stresses were evaluated. <b>Results:</b> Model-predicted values were similar to those reported previously in sensor-based experiments; increased stiffness resulted in greater average normal pressure and shear stress (<i>p</i> < 0.05). <b>Conclusions:</b> These methods may be useful to aid experimental studies by providing insights into the effects of varied prosthesis design parameters or gait conditions on residual limb-socket interface dynamics. The current results suggest that these effects may be subject-specific.</p>\",\"PeriodicalId\":2,\"journal\":{\"name\":\"ACS Applied Bio Materials\",\"volume\":\" \",\"pages\":\"20556683221111986\"},\"PeriodicalIF\":4.6000,\"publicationDate\":\"2022-07-15\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/c3/98/10.1177_20556683221111986.PMC9289901.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"ACS Applied Bio Materials\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1177/20556683221111986\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2022/1/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"Q2\",\"JCRName\":\"MATERIALS SCIENCE, BIOMATERIALS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"ACS Applied Bio Materials","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1177/20556683221111986","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2022/1/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"MATERIALS SCIENCE, BIOMATERIALS","Score":null,"Total":0}
A simulation-based analysis of the effects of variable prosthesis stiffness on interface dynamics between the prosthetic socket and residual limb.
Introduction: Loading of a residual limb within a prosthetic socket can cause tissue damage such as ulceration. Computational simulations may be useful tools for estimating tissue loading within the socket, and thus provide insights into how prosthesis designs affect residual limb-socket interface dynamics. The purpose of this study was to model and simulate residual limb-socket interface dynamics and evaluate the effects of varied prosthesis stiffness on interface dynamics during gait. Methods: A spatial contact model of a residual limb-socket interface was developed and integrated into a gait model with a below-knee amputation. Gait trials were simulated for four subjects walking with low, medium, and high prosthesis stiffness settings. The effects of prosthesis stiffness on interface kinematics, normal pressure, and shear stresses were evaluated. Results: Model-predicted values were similar to those reported previously in sensor-based experiments; increased stiffness resulted in greater average normal pressure and shear stress (p < 0.05). Conclusions: These methods may be useful to aid experimental studies by providing insights into the effects of varied prosthesis design parameters or gait conditions on residual limb-socket interface dynamics. The current results suggest that these effects may be subject-specific.
期刊介绍:
ACS Applied Bio Materials is an interdisciplinary journal publishing original research covering all aspects of biomaterials and biointerfaces including and beyond the traditional biosensing, biomedical and therapeutic applications.
The journal is devoted to reports of new and original experimental and theoretical research of an applied nature that integrates knowledge in the areas of materials, engineering, physics, bioscience, and chemistry into important bio applications. The journal is specifically interested in work that addresses the relationship between structure and function and assesses the stability and degradation of materials under relevant environmental and biological conditions.