{"title":"使用肝素锁定的隧道式透析导管测量血液透析患者国际正常化比率的新方法。","authors":"Céline B Seghers, Kristien Ver Elst, Jolien Claessens, Steven Weekx, Sigrid Vermeiren, Manu Henckes","doi":"10.1155/2020/7586437","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>To measure International Normalized Ratio (INR) in hemodialysis patients with tunneled dialysis catheters (TDCs), blood sampling is frequently obtained via the catheter at the start of the session. INR measurements via finger-prick point of care testing (POCT) and via blood sampling taken from the dialysis circuit are evaluated as alternatives.</p><p><strong>Methods: </strong>In 14 hemodialysis patients with TDCs, treated with vitamin K antagonists (VKA), INR measurements via POCT were compared with plasma INR samples taken via the catheter at the start of dialysis and via the dialysis circuit after 30 and 60 minutes during 3 nonconsecutive dialysis sessions.</p><p><strong>Results: </strong>Blood samples taken at the start of dialysis at the catheter site were frequently contaminated with heparin originating from the locking solution (unfractionated heparin concentration (UFH) >1.0 IU/ml in 13.2%). POCT INR at the start of dialysis was not different from plasma INR after 30 and 60 minutes (Wilcoxon test <i>p</i>=0.113, <i>n</i> = 37, and <i>p</i>=0.631, <i>n</i> = 36, respectively). Moreover, there was no difference between POCT INR at the start of dialysis and POCT INR after 30 and 60 minutes (Wilcoxon test <i>p</i>=0.797 and <i>p</i> = 0.801, respectively; <i>n</i> = 36). Passing and Bablok regression equation was used, <i>y</i> = 0.460 + 0.733x; <i>n</i> = 105. Treatment decisions based on these 2 methods showed a very good overall agreement (kappa = 0.810; 95% CI: 0.732-0.889; <i>n</i> = 105).</p><p><strong>Conclusions: </strong>Measuring plasma INR via the TDC at the start of dialysis should be abandoned. Measuring POCT INR via a finger prick at the start or even after 30 to 60 minutes is an alternative. The most elegant alternative is to take plasma INR samples via the dialysis circuit 30 minutes or later after the start of the dialysis.</p>","PeriodicalId":14177,"journal":{"name":"International Journal of Nephrology","volume":"2020 ","pages":"7586437"},"PeriodicalIF":1.7000,"publicationDate":"2020-12-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7772028/pdf/","citationCount":"0","resultStr":"{\"title\":\"A New Method for the Measurement of International Normalized Ratio in Hemodialysis Patients with Heparin-Locked Tunneled Dialysis Catheters.\",\"authors\":\"Céline B Seghers, Kristien Ver Elst, Jolien Claessens, Steven Weekx, Sigrid Vermeiren, Manu Henckes\",\"doi\":\"10.1155/2020/7586437\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>To measure International Normalized Ratio (INR) in hemodialysis patients with tunneled dialysis catheters (TDCs), blood sampling is frequently obtained via the catheter at the start of the session. INR measurements via finger-prick point of care testing (POCT) and via blood sampling taken from the dialysis circuit are evaluated as alternatives.</p><p><strong>Methods: </strong>In 14 hemodialysis patients with TDCs, treated with vitamin K antagonists (VKA), INR measurements via POCT were compared with plasma INR samples taken via the catheter at the start of dialysis and via the dialysis circuit after 30 and 60 minutes during 3 nonconsecutive dialysis sessions.</p><p><strong>Results: </strong>Blood samples taken at the start of dialysis at the catheter site were frequently contaminated with heparin originating from the locking solution (unfractionated heparin concentration (UFH) >1.0 IU/ml in 13.2%). POCT INR at the start of dialysis was not different from plasma INR after 30 and 60 minutes (Wilcoxon test <i>p</i>=0.113, <i>n</i> = 37, and <i>p</i>=0.631, <i>n</i> = 36, respectively). Moreover, there was no difference between POCT INR at the start of dialysis and POCT INR after 30 and 60 minutes (Wilcoxon test <i>p</i>=0.797 and <i>p</i> = 0.801, respectively; <i>n</i> = 36). Passing and Bablok regression equation was used, <i>y</i> = 0.460 + 0.733x; <i>n</i> = 105. Treatment decisions based on these 2 methods showed a very good overall agreement (kappa = 0.810; 95% CI: 0.732-0.889; <i>n</i> = 105).</p><p><strong>Conclusions: </strong>Measuring plasma INR via the TDC at the start of dialysis should be abandoned. Measuring POCT INR via a finger prick at the start or even after 30 to 60 minutes is an alternative. The most elegant alternative is to take plasma INR samples via the dialysis circuit 30 minutes or later after the start of the dialysis.</p>\",\"PeriodicalId\":14177,\"journal\":{\"name\":\"International Journal of Nephrology\",\"volume\":\"2020 \",\"pages\":\"7586437\"},\"PeriodicalIF\":1.7000,\"publicationDate\":\"2020-12-22\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7772028/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"International Journal of Nephrology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1155/2020/7586437\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2020/1/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"Q3\",\"JCRName\":\"UROLOGY & NEPHROLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Nephrology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1155/2020/7586437","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2020/1/1 0:00:00","PubModel":"eCollection","JCR":"Q3","JCRName":"UROLOGY & NEPHROLOGY","Score":null,"Total":0}
A New Method for the Measurement of International Normalized Ratio in Hemodialysis Patients with Heparin-Locked Tunneled Dialysis Catheters.
Background: To measure International Normalized Ratio (INR) in hemodialysis patients with tunneled dialysis catheters (TDCs), blood sampling is frequently obtained via the catheter at the start of the session. INR measurements via finger-prick point of care testing (POCT) and via blood sampling taken from the dialysis circuit are evaluated as alternatives.
Methods: In 14 hemodialysis patients with TDCs, treated with vitamin K antagonists (VKA), INR measurements via POCT were compared with plasma INR samples taken via the catheter at the start of dialysis and via the dialysis circuit after 30 and 60 minutes during 3 nonconsecutive dialysis sessions.
Results: Blood samples taken at the start of dialysis at the catheter site were frequently contaminated with heparin originating from the locking solution (unfractionated heparin concentration (UFH) >1.0 IU/ml in 13.2%). POCT INR at the start of dialysis was not different from plasma INR after 30 and 60 minutes (Wilcoxon test p=0.113, n = 37, and p=0.631, n = 36, respectively). Moreover, there was no difference between POCT INR at the start of dialysis and POCT INR after 30 and 60 minutes (Wilcoxon test p=0.797 and p = 0.801, respectively; n = 36). Passing and Bablok regression equation was used, y = 0.460 + 0.733x; n = 105. Treatment decisions based on these 2 methods showed a very good overall agreement (kappa = 0.810; 95% CI: 0.732-0.889; n = 105).
Conclusions: Measuring plasma INR via the TDC at the start of dialysis should be abandoned. Measuring POCT INR via a finger prick at the start or even after 30 to 60 minutes is an alternative. The most elegant alternative is to take plasma INR samples via the dialysis circuit 30 minutes or later after the start of the dialysis.
期刊介绍:
International Journal of Nephrology is a peer-reviewed, Open Access journal that publishes original research articles, review articles, and clinical studies focusing on the prevention, diagnosis, and management of kidney diseases and associated disorders. The journal welcomes submissions related to cell biology, developmental biology, genetics, immunology, pathology, pathophysiology of renal disease and progression, clinical nephrology, dialysis, and transplantation.