用于骨矿物质测量的GE医疗保健Lunar Prodigy和iDXA双能x射线密度仪的交叉校准。

IF 1.1 Q3 ORTHOPEDICS
Journal of Osteoporosis Pub Date : 2016-01-01 Epub Date: 2016-04-27 DOI:10.1155/2016/1424582
J Saarelainen, M Hakulinen, T Rikkonen, H Kröger, M Tuppurainen, H Koivumaa-Honkanen, R Honkanen, M Hujo, J S Jurvelin
{"title":"用于骨矿物质测量的GE医疗保健Lunar Prodigy和iDXA双能x射线密度仪的交叉校准。","authors":"J Saarelainen,&nbsp;M Hakulinen,&nbsp;T Rikkonen,&nbsp;H Kröger,&nbsp;M Tuppurainen,&nbsp;H Koivumaa-Honkanen,&nbsp;R Honkanen,&nbsp;M Hujo,&nbsp;J S Jurvelin","doi":"10.1155/2016/1424582","DOIUrl":null,"url":null,"abstract":"<p><p>In long-term prospective studies, dual-energy X-ray absorptiometry (DXA) devices need to be inevitably changed. It is essential to assess whether systematic differences will exist between measurements with the new and old device. A group of female volunteers (21-72 years) underwent anteroposterior lumbar spine L2-L4 (n = 72), proximal femur (n = 72), and total body (n = 62) measurements with the Prodigy and the iDXA scanners at the same visit. The bone mineral density (BMD) measurements with these two scanners showed a high linear association at all tested sites (r = 0.962-0.995; p < 0.0001). The average iDXA BMD values were 1.5%, 0.5%, and 0.9% higher than those of Prodigy for lumbar spine (L2-L4) (p < 0.0001), femoral neck (p = 0.048), and total hip (p < 0.0001), respectively. Total body BMD values measured with the iDXA were -1.3% lower (p < 0.0001) than those measured with the Prodigy. For total body, lumbar spine, and femoral neck, the BMD differences as measured with these two devices were independent of subject height and weight. Linear correction equations were developed to ensure comparability of BMD measurements obtained with both DXA scanners. Importantly, use of equations from previous studies would have increased the discrepancy between these particular DXA scanners, especially at hip and at spine. </p>","PeriodicalId":45384,"journal":{"name":"Journal of Osteoporosis","volume":"2016 ","pages":"1424582"},"PeriodicalIF":1.1000,"publicationDate":"2016-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1155/2016/1424582","citationCount":"18","resultStr":"{\"title\":\"Cross-Calibration of GE Healthcare Lunar Prodigy and iDXA Dual-Energy X-Ray Densitometers for Bone Mineral Measurements.\",\"authors\":\"J Saarelainen,&nbsp;M Hakulinen,&nbsp;T Rikkonen,&nbsp;H Kröger,&nbsp;M Tuppurainen,&nbsp;H Koivumaa-Honkanen,&nbsp;R Honkanen,&nbsp;M Hujo,&nbsp;J S Jurvelin\",\"doi\":\"10.1155/2016/1424582\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>In long-term prospective studies, dual-energy X-ray absorptiometry (DXA) devices need to be inevitably changed. It is essential to assess whether systematic differences will exist between measurements with the new and old device. A group of female volunteers (21-72 years) underwent anteroposterior lumbar spine L2-L4 (n = 72), proximal femur (n = 72), and total body (n = 62) measurements with the Prodigy and the iDXA scanners at the same visit. The bone mineral density (BMD) measurements with these two scanners showed a high linear association at all tested sites (r = 0.962-0.995; p < 0.0001). The average iDXA BMD values were 1.5%, 0.5%, and 0.9% higher than those of Prodigy for lumbar spine (L2-L4) (p < 0.0001), femoral neck (p = 0.048), and total hip (p < 0.0001), respectively. Total body BMD values measured with the iDXA were -1.3% lower (p < 0.0001) than those measured with the Prodigy. For total body, lumbar spine, and femoral neck, the BMD differences as measured with these two devices were independent of subject height and weight. Linear correction equations were developed to ensure comparability of BMD measurements obtained with both DXA scanners. Importantly, use of equations from previous studies would have increased the discrepancy between these particular DXA scanners, especially at hip and at spine. </p>\",\"PeriodicalId\":45384,\"journal\":{\"name\":\"Journal of Osteoporosis\",\"volume\":\"2016 \",\"pages\":\"1424582\"},\"PeriodicalIF\":1.1000,\"publicationDate\":\"2016-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1155/2016/1424582\",\"citationCount\":\"18\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Osteoporosis\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1155/2016/1424582\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2016/4/27 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q3\",\"JCRName\":\"ORTHOPEDICS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Osteoporosis","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1155/2016/1424582","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2016/4/27 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 18

摘要

在长期前瞻性研究中,双能x射线吸收仪(DXA)设备不可避免地需要改变。评估新旧仪器测量结果之间是否存在系统差异是至关重要的。一组女性志愿者(21-72岁)在同一次访问中使用Prodigy和iDXA扫描仪进行腰椎L2-L4前后位(n = 72)、股骨近端(n = 72)和全身(n = 62)测量。用这两种扫描仪测量的骨密度(BMD)在所有测试部位都显示出高度的线性关联(r = 0.962-0.995;P < 0.0001)。与Prodigy相比,iDXA BMD在腰椎(L2-L4)、股骨颈(p = 0.048)和全髋关节(p < 0.0001)的平均值分别高出1.5%、0.5%和0.9%。用iDXA测量的总体骨密度值比用Prodigy测量的低-1.3% (p < 0.0001)。对于全身、腰椎和股骨颈,用这两种装置测量的骨密度差异与受试者的身高和体重无关。建立了线性校正方程,以确保两种DXA扫描仪获得的骨密度测量值的可比性。重要的是,使用先前研究中的公式会增加这些特定DXA扫描仪之间的差异,特别是在髋关节和脊柱。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

Cross-Calibration of GE Healthcare Lunar Prodigy and iDXA Dual-Energy X-Ray Densitometers for Bone Mineral Measurements.

Cross-Calibration of GE Healthcare Lunar Prodigy and iDXA Dual-Energy X-Ray Densitometers for Bone Mineral Measurements.

Cross-Calibration of GE Healthcare Lunar Prodigy and iDXA Dual-Energy X-Ray Densitometers for Bone Mineral Measurements.

In long-term prospective studies, dual-energy X-ray absorptiometry (DXA) devices need to be inevitably changed. It is essential to assess whether systematic differences will exist between measurements with the new and old device. A group of female volunteers (21-72 years) underwent anteroposterior lumbar spine L2-L4 (n = 72), proximal femur (n = 72), and total body (n = 62) measurements with the Prodigy and the iDXA scanners at the same visit. The bone mineral density (BMD) measurements with these two scanners showed a high linear association at all tested sites (r = 0.962-0.995; p < 0.0001). The average iDXA BMD values were 1.5%, 0.5%, and 0.9% higher than those of Prodigy for lumbar spine (L2-L4) (p < 0.0001), femoral neck (p = 0.048), and total hip (p < 0.0001), respectively. Total body BMD values measured with the iDXA were -1.3% lower (p < 0.0001) than those measured with the Prodigy. For total body, lumbar spine, and femoral neck, the BMD differences as measured with these two devices were independent of subject height and weight. Linear correction equations were developed to ensure comparability of BMD measurements obtained with both DXA scanners. Importantly, use of equations from previous studies would have increased the discrepancy between these particular DXA scanners, especially at hip and at spine.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
3.60
自引率
0.00%
发文量
6
审稿时长
20 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信