[内科重症监护病房与专科重症监护病房有创呼吸机报警常见原因比较分析]。

Wei Tan, Long-feng Sun, Zheng Qin, Bing Dai, Hong-wen Zhao, Jian Kang
{"title":"[内科重症监护病房与专科重症监护病房有创呼吸机报警常见原因比较分析]。","authors":"Wei Tan,&nbsp;Long-feng Sun,&nbsp;Zheng Qin,&nbsp;Bing Dai,&nbsp;Hong-wen Zhao,&nbsp;Jian Kang","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To analyze the common reasons of invasive ventilator alarms between medical intensive care unit (ICU) and specialist ICU, and its related management methods.</p><p><strong>Methods: </strong>Patients admitted to medical ICU and specialist ICU from January to December in 2011 of the First Hospital of China Medical University were studied. Ventilator alarms and their reasons need to be handle by the front-line doctors, respiratory therapists, attending physicians or medical ICU doctors were analyzed and compared.</p><p><strong>Results: </strong>There were 375 ventilator alarms of the 59 patients in the medical ICU, incidence of the top three alarms parameters were high airway pressure alarms for 21.87%, high tide volume alarms for 15.73% and high minute ventilation alarms for 14.13%. In specialist ICU there were a total of 403 ventilator alarms with 249 patients, incidence of the top three alarms parameters were high airway pressure alarms for 32.51%, low airway pressure alarms for 15.38%, high respiratory rate alarms for 10.42%. The incidence of high airway pressure and low airway pressure alarms in medical ICU were significantly lower than the specialist ICU (21.87% vs. 32.51%, 8.53% vs. 15.38%, both P<0.01), and the incidence of high minute ventilation and high tidal volume alarms in medical ICU were higher than specialist ICU (14.13% vs. 7.20%, 15.73% vs. 9.68%, P<0.01 and P<0.05). The top three causes of the alarms were aerosol inhalation, sputum blockage, and oxygen battery expired in medical ICU, and sputum blockage, respiratory distress, and pipeline leak and oxygen expired battery in specialist ICU. The reasons of sputum blockage, tubes factors (intubation position change, pipeline water) and improper alarm parameters setting in medical ICU was significantly lower than those in specialist ICU (10.93% vs. 17.12%, 1.87% vs. 4.47%, 1.33% vs. 3.72%, 1.60% vs. 3.97%, all P<0.05). High tidal volume, high minute ventilation and serious breath-side filter blockage because of aerosol inhalation in medical ICU were significantly higher than those in specialist ICU (18.93% vs. 3.97%, P<0.01).</p><p><strong>Conclusion: </strong>Doctors in medical ICU and specialist ICU should understand the ventilator alarms characteristics, prevention, detect and timely problems management.</p>","PeriodicalId":23992,"journal":{"name":"Zhongguo wei zhong bing ji jiu yi xue = Chinese critical care medicine = Zhongguo weizhongbing jijiuyixue","volume":"24 10","pages":"582-6"},"PeriodicalIF":0.0000,"publicationDate":"2012-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[The comparative analysis of the common reasons of invasive ventilator alarms between medical and specialist intensive care unit].\",\"authors\":\"Wei Tan,&nbsp;Long-feng Sun,&nbsp;Zheng Qin,&nbsp;Bing Dai,&nbsp;Hong-wen Zhao,&nbsp;Jian Kang\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>To analyze the common reasons of invasive ventilator alarms between medical intensive care unit (ICU) and specialist ICU, and its related management methods.</p><p><strong>Methods: </strong>Patients admitted to medical ICU and specialist ICU from January to December in 2011 of the First Hospital of China Medical University were studied. Ventilator alarms and their reasons need to be handle by the front-line doctors, respiratory therapists, attending physicians or medical ICU doctors were analyzed and compared.</p><p><strong>Results: </strong>There were 375 ventilator alarms of the 59 patients in the medical ICU, incidence of the top three alarms parameters were high airway pressure alarms for 21.87%, high tide volume alarms for 15.73% and high minute ventilation alarms for 14.13%. In specialist ICU there were a total of 403 ventilator alarms with 249 patients, incidence of the top three alarms parameters were high airway pressure alarms for 32.51%, low airway pressure alarms for 15.38%, high respiratory rate alarms for 10.42%. The incidence of high airway pressure and low airway pressure alarms in medical ICU were significantly lower than the specialist ICU (21.87% vs. 32.51%, 8.53% vs. 15.38%, both P<0.01), and the incidence of high minute ventilation and high tidal volume alarms in medical ICU were higher than specialist ICU (14.13% vs. 7.20%, 15.73% vs. 9.68%, P<0.01 and P<0.05). The top three causes of the alarms were aerosol inhalation, sputum blockage, and oxygen battery expired in medical ICU, and sputum blockage, respiratory distress, and pipeline leak and oxygen expired battery in specialist ICU. The reasons of sputum blockage, tubes factors (intubation position change, pipeline water) and improper alarm parameters setting in medical ICU was significantly lower than those in specialist ICU (10.93% vs. 17.12%, 1.87% vs. 4.47%, 1.33% vs. 3.72%, 1.60% vs. 3.97%, all P<0.05). High tidal volume, high minute ventilation and serious breath-side filter blockage because of aerosol inhalation in medical ICU were significantly higher than those in specialist ICU (18.93% vs. 3.97%, P<0.01).</p><p><strong>Conclusion: </strong>Doctors in medical ICU and specialist ICU should understand the ventilator alarms characteristics, prevention, detect and timely problems management.</p>\",\"PeriodicalId\":23992,\"journal\":{\"name\":\"Zhongguo wei zhong bing ji jiu yi xue = Chinese critical care medicine = Zhongguo weizhongbing jijiuyixue\",\"volume\":\"24 10\",\"pages\":\"582-6\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2012-10-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Zhongguo wei zhong bing ji jiu yi xue = Chinese critical care medicine = Zhongguo weizhongbing jijiuyixue\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Zhongguo wei zhong bing ji jiu yi xue = Chinese critical care medicine = Zhongguo weizhongbing jijiuyixue","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目的:分析内科重症监护病房(ICU)与专科重症监护病房(ICU)有创呼吸机报警的常见原因,并探讨相应的处理方法。方法:对2011年1 - 12月中国医科大学第一医院内科ICU和专科ICU收治的患者进行分析。对一线医生、呼吸治疗师、主治医生或内科ICU医生需要处理的呼吸机报警及其原因进行分析比较。结果:59例内科ICU患者共发生375次呼吸机报警,前3位报警参数的发生率分别为高气道压报警(21.87%)、高潮气量报警(15.73%)和高分钟通气报警(14.13%)。专科ICU共有249例患者发生403次呼吸机报警,前3位报警参数的发生率分别为高气道压报警(32.51%)、低气道压报警(15.38%)、高呼吸率报警(10.42%)。内科ICU高、低气道压报警发生率明显低于专科ICU (21.87% vs. 32.51%, 8.53% vs. 15.38%)。结论:内科ICU和专科ICU医生应了解呼吸机报警特点,预防、发现并及时处理问题。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
[The comparative analysis of the common reasons of invasive ventilator alarms between medical and specialist intensive care unit].

Objective: To analyze the common reasons of invasive ventilator alarms between medical intensive care unit (ICU) and specialist ICU, and its related management methods.

Methods: Patients admitted to medical ICU and specialist ICU from January to December in 2011 of the First Hospital of China Medical University were studied. Ventilator alarms and their reasons need to be handle by the front-line doctors, respiratory therapists, attending physicians or medical ICU doctors were analyzed and compared.

Results: There were 375 ventilator alarms of the 59 patients in the medical ICU, incidence of the top three alarms parameters were high airway pressure alarms for 21.87%, high tide volume alarms for 15.73% and high minute ventilation alarms for 14.13%. In specialist ICU there were a total of 403 ventilator alarms with 249 patients, incidence of the top three alarms parameters were high airway pressure alarms for 32.51%, low airway pressure alarms for 15.38%, high respiratory rate alarms for 10.42%. The incidence of high airway pressure and low airway pressure alarms in medical ICU were significantly lower than the specialist ICU (21.87% vs. 32.51%, 8.53% vs. 15.38%, both P<0.01), and the incidence of high minute ventilation and high tidal volume alarms in medical ICU were higher than specialist ICU (14.13% vs. 7.20%, 15.73% vs. 9.68%, P<0.01 and P<0.05). The top three causes of the alarms were aerosol inhalation, sputum blockage, and oxygen battery expired in medical ICU, and sputum blockage, respiratory distress, and pipeline leak and oxygen expired battery in specialist ICU. The reasons of sputum blockage, tubes factors (intubation position change, pipeline water) and improper alarm parameters setting in medical ICU was significantly lower than those in specialist ICU (10.93% vs. 17.12%, 1.87% vs. 4.47%, 1.33% vs. 3.72%, 1.60% vs. 3.97%, all P<0.05). High tidal volume, high minute ventilation and serious breath-side filter blockage because of aerosol inhalation in medical ICU were significantly higher than those in specialist ICU (18.93% vs. 3.97%, P<0.01).

Conclusion: Doctors in medical ICU and specialist ICU should understand the ventilator alarms characteristics, prevention, detect and timely problems management.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信