{"title":"除颤对植入式迷走神经刺激系统性能的影响。","authors":"Imad Libbus, Scott R Stubbs, Scott T Mazar, Scott Mindrebo, Bruce H KenKnight, Lorenzo A DiCarlo","doi":"10.1186/s42234-021-00064-w","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Vagus Nerve Stimulation (VNS) delivers Autonomic Regulation Therapy (ART) for heart failure (HF), and has been associated with improvement in cardiac function and heart failure symptoms. VNS is delivered using an implantable pulse generator (IPG) and lead with electrodes placed around the cervical vagus nerve. Because HF patients may receive concomitant cardiac defibrillation therapy, testing was conducted to determine the effect of defibrillation (DF) on the VNS system.</p><p><strong>Methods: </strong>DF testing was conducted on three ART IPGs (LivaNova USA, Inc.) according to international standard ISO14708-1, which evaluated whether DF had any permanent effects on the system. Each IPG was connected to a defibrillation pulse generator and subjected to a series of high-energy pulses.</p><p><strong>Results: </strong>The specified series of pulses were successfully delivered to each of the three devices. All three IPGs passed factory electrical tests, and interrogation confirmed that software and data were unchanged from the pre-programmed values. No shifts in parameters or failures were observed.</p><p><strong>Conclusions: </strong>Implantable VNS systems were tested for immunity to defibrillation, and were found to be unaffected by a series of high-energy defibrillation pulses. These results suggest that this VNS system can be used safely and continue to function after patients have been defibrillated.</p>","PeriodicalId":72363,"journal":{"name":"Bioelectronic medicine","volume":" ","pages":"3"},"PeriodicalIF":0.0000,"publicationDate":"2021-03-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7962382/pdf/","citationCount":"0","resultStr":"{\"title\":\"Effect of defibrillation on the performance of an implantable vagus nerve stimulation system.\",\"authors\":\"Imad Libbus, Scott R Stubbs, Scott T Mazar, Scott Mindrebo, Bruce H KenKnight, Lorenzo A DiCarlo\",\"doi\":\"10.1186/s42234-021-00064-w\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Vagus Nerve Stimulation (VNS) delivers Autonomic Regulation Therapy (ART) for heart failure (HF), and has been associated with improvement in cardiac function and heart failure symptoms. VNS is delivered using an implantable pulse generator (IPG) and lead with electrodes placed around the cervical vagus nerve. Because HF patients may receive concomitant cardiac defibrillation therapy, testing was conducted to determine the effect of defibrillation (DF) on the VNS system.</p><p><strong>Methods: </strong>DF testing was conducted on three ART IPGs (LivaNova USA, Inc.) according to international standard ISO14708-1, which evaluated whether DF had any permanent effects on the system. Each IPG was connected to a defibrillation pulse generator and subjected to a series of high-energy pulses.</p><p><strong>Results: </strong>The specified series of pulses were successfully delivered to each of the three devices. All three IPGs passed factory electrical tests, and interrogation confirmed that software and data were unchanged from the pre-programmed values. No shifts in parameters or failures were observed.</p><p><strong>Conclusions: </strong>Implantable VNS systems were tested for immunity to defibrillation, and were found to be unaffected by a series of high-energy defibrillation pulses. These results suggest that this VNS system can be used safely and continue to function after patients have been defibrillated.</p>\",\"PeriodicalId\":72363,\"journal\":{\"name\":\"Bioelectronic medicine\",\"volume\":\" \",\"pages\":\"3\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2021-03-16\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7962382/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Bioelectronic medicine\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1186/s42234-021-00064-w\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Bioelectronic medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1186/s42234-021-00064-w","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
摘要
背景:迷走神经刺激疗法(VNS)是治疗心力衰竭(HF)的自律神经调节疗法(ART),可改善心脏功能和心力衰竭症状。VNS 采用植入式脉冲发生器 (IPG) 和导线,电极放置在颈迷走神经周围。由于高频患者可能会同时接受心脏除颤治疗,因此进行了测试以确定除颤(DF)对 VNS 系统的影响:根据国际标准 ISO14708-1,对三个 ART IPG(LivaNova USA, Inc.)每个 IPG 都与除颤脉冲发生器相连,并接受一系列高能量脉冲:结果:三个装置都成功地输出了指定的一系列脉冲。所有三个 IPG 都通过了出厂电气测试,检查确认软件和数据与预编程值相同。没有发现参数变化或故障:对植入式 VNS 系统进行了除颤抗扰度测试,发现它不受一系列高能除颤脉冲的影响。这些结果表明,这种 VNS 系统可以安全使用,并在患者除颤后继续发挥作用。
Effect of defibrillation on the performance of an implantable vagus nerve stimulation system.
Background: Vagus Nerve Stimulation (VNS) delivers Autonomic Regulation Therapy (ART) for heart failure (HF), and has been associated with improvement in cardiac function and heart failure symptoms. VNS is delivered using an implantable pulse generator (IPG) and lead with electrodes placed around the cervical vagus nerve. Because HF patients may receive concomitant cardiac defibrillation therapy, testing was conducted to determine the effect of defibrillation (DF) on the VNS system.
Methods: DF testing was conducted on three ART IPGs (LivaNova USA, Inc.) according to international standard ISO14708-1, which evaluated whether DF had any permanent effects on the system. Each IPG was connected to a defibrillation pulse generator and subjected to a series of high-energy pulses.
Results: The specified series of pulses were successfully delivered to each of the three devices. All three IPGs passed factory electrical tests, and interrogation confirmed that software and data were unchanged from the pre-programmed values. No shifts in parameters or failures were observed.
Conclusions: Implantable VNS systems were tested for immunity to defibrillation, and were found to be unaffected by a series of high-energy defibrillation pulses. These results suggest that this VNS system can be used safely and continue to function after patients have been defibrillated.