{"title":"[颗粒增强免疫比浊法测定高敏感c反应蛋白]。","authors":"Min Hu, Yan-wu Li, Yong-hong Hu","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To evaluate the particle enhanced immunoturbidimetic assay of high sensitive C-reactive protein by the automated chemistry analyzer (Hitachi 7170A).</p><p><strong>Methods: </strong>The detection limit of quantification, measurement range, precision, interference and reference interval were assessed.</p><p><strong>Results: </strong>The measurement range of the method was from 0.125 to 12.5 mg/L. The detection limitation was 0.115 mg/L. The internal and external accuracy were 0.89% and 0.90% respectively. The reference range of the assay was from 0.20 mg/L to 2.60 mg/L.</p><p><strong>Conclusion: </strong>The particle enhanced immunoturbidimetic assay is sensitive and stable, and can be used in clinical diagnosis.</p>","PeriodicalId":13115,"journal":{"name":"Hunan yi ke da xue xue bao = Hunan yike daxue xuebao = Bulletin of Hunan Medical University","volume":"28 4","pages":"415-7"},"PeriodicalIF":0.0000,"publicationDate":"2003-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Particle enhanced immunoturbidimetic assay of high sensitive C-reactive protein].\",\"authors\":\"Min Hu, Yan-wu Li, Yong-hong Hu\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>To evaluate the particle enhanced immunoturbidimetic assay of high sensitive C-reactive protein by the automated chemistry analyzer (Hitachi 7170A).</p><p><strong>Methods: </strong>The detection limit of quantification, measurement range, precision, interference and reference interval were assessed.</p><p><strong>Results: </strong>The measurement range of the method was from 0.125 to 12.5 mg/L. The detection limitation was 0.115 mg/L. The internal and external accuracy were 0.89% and 0.90% respectively. The reference range of the assay was from 0.20 mg/L to 2.60 mg/L.</p><p><strong>Conclusion: </strong>The particle enhanced immunoturbidimetic assay is sensitive and stable, and can be used in clinical diagnosis.</p>\",\"PeriodicalId\":13115,\"journal\":{\"name\":\"Hunan yi ke da xue xue bao = Hunan yike daxue xuebao = Bulletin of Hunan Medical University\",\"volume\":\"28 4\",\"pages\":\"415-7\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2003-08-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Hunan yi ke da xue xue bao = Hunan yike daxue xuebao = Bulletin of Hunan Medical University\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Hunan yi ke da xue xue bao = Hunan yike daxue xuebao = Bulletin of Hunan Medical University","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
[Particle enhanced immunoturbidimetic assay of high sensitive C-reactive protein].
Objective: To evaluate the particle enhanced immunoturbidimetic assay of high sensitive C-reactive protein by the automated chemistry analyzer (Hitachi 7170A).
Methods: The detection limit of quantification, measurement range, precision, interference and reference interval were assessed.
Results: The measurement range of the method was from 0.125 to 12.5 mg/L. The detection limitation was 0.115 mg/L. The internal and external accuracy were 0.89% and 0.90% respectively. The reference range of the assay was from 0.20 mg/L to 2.60 mg/L.
Conclusion: The particle enhanced immunoturbidimetic assay is sensitive and stable, and can be used in clinical diagnosis.