{"title":"狗的Bion髋关节假体失败后的组织反应。","authors":"S D Cook, R C Anderson, H G French, R J Haddad","doi":"10.3109/10731198409118818","DOIUrl":null,"url":null,"abstract":"<p><p>The tissue reaction in a dog to a failed constrained total hip replacement system with a Bion elastomer component was evaluated. The device consisted of Ti-6A1-4V alloy acetabular and femoral components connected with a Bion component. At necropsy the Bion component was found to have fractured. The tissue surrounding the Bion was granulomatous with numerous blackened regions. Histologically a severe chronic inflammatory response was present in the surrounding tissue with numerous giant cells and Bion laden macrophages.</p>","PeriodicalId":75597,"journal":{"name":"Biomaterials, medical devices, and artificial organs","volume":"12 1-2","pages":"15-23"},"PeriodicalIF":0.0000,"publicationDate":"1984-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.3109/10731198409118818","citationCount":"1","resultStr":"{\"title\":\"Tissue reaction to a failed Bion hip prosthesis in a dog.\",\"authors\":\"S D Cook, R C Anderson, H G French, R J Haddad\",\"doi\":\"10.3109/10731198409118818\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The tissue reaction in a dog to a failed constrained total hip replacement system with a Bion elastomer component was evaluated. The device consisted of Ti-6A1-4V alloy acetabular and femoral components connected with a Bion component. At necropsy the Bion component was found to have fractured. The tissue surrounding the Bion was granulomatous with numerous blackened regions. Histologically a severe chronic inflammatory response was present in the surrounding tissue with numerous giant cells and Bion laden macrophages.</p>\",\"PeriodicalId\":75597,\"journal\":{\"name\":\"Biomaterials, medical devices, and artificial organs\",\"volume\":\"12 1-2\",\"pages\":\"15-23\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1984-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.3109/10731198409118818\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Biomaterials, medical devices, and artificial organs\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.3109/10731198409118818\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biomaterials, medical devices, and artificial organs","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3109/10731198409118818","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Tissue reaction to a failed Bion hip prosthesis in a dog.
The tissue reaction in a dog to a failed constrained total hip replacement system with a Bion elastomer component was evaluated. The device consisted of Ti-6A1-4V alloy acetabular and femoral components connected with a Bion component. At necropsy the Bion component was found to have fractured. The tissue surrounding the Bion was granulomatous with numerous blackened regions. Histologically a severe chronic inflammatory response was present in the surrounding tissue with numerous giant cells and Bion laden macrophages.