{"title":"用于提高铀灵敏度和质量分馏的高效腔离子源设计改进。","authors":"Varun Kant Yadav,&nbsp;R. K. Bhatia,&nbsp;A. Singh,&nbsp;Rabi Datta,&nbsp;M. M. Gulhane,&nbsp;E. Ravisankar,&nbsp;R. D. Phulsundar,&nbsp;K. G. Bhushan","doi":"10.1002/rcm.70184","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Rationale</h3>\n \n <p>Thermal ionization mass spectrometry (TIMS) integrated with a high-efficiency cavity ion source (CIS) has emerged as a promising analytical tool for precise isotope ratio measurements and trace elemental analysis in nuclear science. Operation with high ionization potential (HIP) elements such as uranium requires higher cavity temperatures and emission currents. This caused glow discharge, thermal instability, localized overheating, and degradation of ion beam stability during prolonged operation. This necessitates a detailed approach for further enhancement of source performance and operational stability.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>The CIS was systematically redesigned for HIP elements through detailed engineering optimization of the cavity geometry, thermionic filament dimension, ion extraction configuration, and shielding electrode design to achieve stable ion beam generation. The optimized CIS assembly was then integrated with indigenous magnetic sector TIMS platform. After that, its analytical performance was systematically evaluated relative to a conventional thermal ionization source (TIS) configuration operated under same experimental conditions. Comparative measurement was carried out in terms of analytical sensitivity (the total detected ion signal per unit sample loading). Also, the time-dependent instrumental mass fractionation behavior and long-term ion beam stability study were carried out.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>The design improvements and systematic optimization of the CIS showed nearly an order of magnitude increase in analytical sensitivity for uranium together with a ~1%–2% improvement in mass fractionation stability relative to the conventional TIS configuration.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>The design improvements carried out in the CIS configuration resulted in higher analytical sensitivity and better resistance to systematic isotopic drift during uranium measurements. This established the CIS TIMS platform as an effective and high-performance analytical tool. It useful for precise actinide isotope ratio measurements relevant to nuclear safeguards, nuclear forensics, and trace isotopic analysis.</p>\n </section>\n </div>","PeriodicalId":225,"journal":{"name":"Rapid Communications in Mass Spectrometry","volume":"40 24","pages":""},"PeriodicalIF":1.8000,"publicationDate":"2026-09-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Design Improvements of a High-Efficiency Cavity Ion Source for Enhanced Uranium Sensitivity and Mass Fractionation Studies Using TIMS\",\"authors\":\"Varun Kant Yadav,&nbsp;R. K. Bhatia,&nbsp;A. Singh,&nbsp;Rabi Datta,&nbsp;M. M. Gulhane,&nbsp;E. Ravisankar,&nbsp;R. D. Phulsundar,&nbsp;K. G. Bhushan\",\"doi\":\"10.1002/rcm.70184\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div>\\n \\n \\n <section>\\n \\n <h3> Rationale</h3>\\n \\n <p>Thermal ionization mass spectrometry (TIMS) integrated with a high-efficiency cavity ion source (CIS) has emerged as a promising analytical tool for precise isotope ratio measurements and trace elemental analysis in nuclear science. Operation with high ionization potential (HIP) elements such as uranium requires higher cavity temperatures and emission currents. This caused glow discharge, thermal instability, localized overheating, and degradation of ion beam stability during prolonged operation. This necessitates a detailed approach for further enhancement of source performance and operational stability.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Methods</h3>\\n \\n <p>The CIS was systematically redesigned for HIP elements through detailed engineering optimization of the cavity geometry, thermionic filament dimension, ion extraction configuration, and shielding electrode design to achieve stable ion beam generation. The optimized CIS assembly was then integrated with indigenous magnetic sector TIMS platform. After that, its analytical performance was systematically evaluated relative to a conventional thermal ionization source (TIS) configuration operated under same experimental conditions. Comparative measurement was carried out in terms of analytical sensitivity (the total detected ion signal per unit sample loading). Also, the time-dependent instrumental mass fractionation behavior and long-term ion beam stability study were carried out.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Results</h3>\\n \\n <p>The design improvements and systematic optimization of the CIS showed nearly an order of magnitude increase in analytical sensitivity for uranium together with a ~1%–2% improvement in mass fractionation stability relative to the conventional TIS configuration.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Conclusions</h3>\\n \\n <p>The design improvements carried out in the CIS configuration resulted in higher analytical sensitivity and better resistance to systematic isotopic drift during uranium measurements. This established the CIS TIMS platform as an effective and high-performance analytical tool. It useful for precise actinide isotope ratio measurements relevant to nuclear safeguards, nuclear forensics, and trace isotopic analysis.</p>\\n </section>\\n </div>\",\"PeriodicalId\":225,\"journal\":{\"name\":\"Rapid Communications in Mass Spectrometry\",\"volume\":\"40 24\",\"pages\":\"\"},\"PeriodicalIF\":1.8000,\"publicationDate\":\"2026-09-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Rapid Communications in Mass Spectrometry\",\"FirstCategoryId\":\"92\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1002/rcm.70184\",\"RegionNum\":3,\"RegionCategory\":\"化学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"BIOCHEMICAL RESEARCH METHODS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Rapid Communications in Mass Spectrometry","FirstCategoryId":"92","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/rcm.70184","RegionNum":3,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"BIOCHEMICAL RESEARCH METHODS","Score":null,"Total":0}
引用次数: 0

摘要

原理:热电离质谱(TIMS)与高效腔离子源(CIS)相结合,已成为核科学中精确同位素比测量和痕量元素分析的一种有前途的分析工具。操作高电离势(HIP)元素,如铀,需要更高的腔温度和发射电流。这会导致辉光放电、热不稳定、局部过热以及在长时间操作期间离子束稳定性的退化。这就需要一种详细的方法来进一步提高源性能和操作稳定性。方法:通过对腔体几何形状、热离子灯丝尺寸、离子提取配置和屏蔽电极设计进行详细的工程优化,系统地重新设计了HIP元件的CIS,以实现稳定的离子束生成。然后将优化后的CIS组件与国产磁扇形TIMS平台集成。之后,在相同的实验条件下,相对于传统的热电离源(TIS)配置,系统地评估了其分析性能。在分析灵敏度(每单位样品负载的总检测离子信号)方面进行了比较测量。此外,还进行了随时间变化的仪器质量分馏行为和长期离子束稳定性研究。结果:CIS的设计改进和系统优化表明,与传统TIS配置相比,CIS对铀的分析灵敏度提高了近一个数量级,质量分馏稳定性提高了约1%-2%。结论:在CIS配置中进行的设计改进提高了分析灵敏度,并在铀测量过程中更好地抵抗系统同位素漂移。这使CIS TIMS平台成为一个有效的高性能分析工具。它可用于与核保障、核取证和痕量同位素分析有关的锕系元素同位素比率的精确测量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Design Improvements of a High-Efficiency Cavity Ion Source for Enhanced Uranium Sensitivity and Mass Fractionation Studies Using TIMS

Rationale

Thermal ionization mass spectrometry (TIMS) integrated with a high-efficiency cavity ion source (CIS) has emerged as a promising analytical tool for precise isotope ratio measurements and trace elemental analysis in nuclear science. Operation with high ionization potential (HIP) elements such as uranium requires higher cavity temperatures and emission currents. This caused glow discharge, thermal instability, localized overheating, and degradation of ion beam stability during prolonged operation. This necessitates a detailed approach for further enhancement of source performance and operational stability.

Methods

The CIS was systematically redesigned for HIP elements through detailed engineering optimization of the cavity geometry, thermionic filament dimension, ion extraction configuration, and shielding electrode design to achieve stable ion beam generation. The optimized CIS assembly was then integrated with indigenous magnetic sector TIMS platform. After that, its analytical performance was systematically evaluated relative to a conventional thermal ionization source (TIS) configuration operated under same experimental conditions. Comparative measurement was carried out in terms of analytical sensitivity (the total detected ion signal per unit sample loading). Also, the time-dependent instrumental mass fractionation behavior and long-term ion beam stability study were carried out.

Results

The design improvements and systematic optimization of the CIS showed nearly an order of magnitude increase in analytical sensitivity for uranium together with a ~1%–2% improvement in mass fractionation stability relative to the conventional TIS configuration.

Conclusions

The design improvements carried out in the CIS configuration resulted in higher analytical sensitivity and better resistance to systematic isotopic drift during uranium measurements. This established the CIS TIMS platform as an effective and high-performance analytical tool. It useful for precise actinide isotope ratio measurements relevant to nuclear safeguards, nuclear forensics, and trace isotopic analysis.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
4.10
自引率
5.00%
发文量
219
审稿时长
2.6 months
期刊介绍: Rapid Communications in Mass Spectrometry is a journal whose aim is the rapid publication of original research results and ideas on all aspects of the science of gas-phase ions; it covers all the associated scientific disciplines. There is no formal limit on paper length ("rapid" is not synonymous with "brief"), but papers should be of a length that is commensurate with the importance and complexity of the results being reported. Contributions may be theoretical or practical in nature; they may deal with methods, techniques and applications, or with the interpretation of results; they may cover any area in science that depends directly on measurements made upon gaseous ions or that is associated with such measurements.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书