{"title":"综合顶点定位系统的体外评估:工作长度精度、操作参数和对NiTi仪器的机械影响。","authors":"Roberto Barreto Osaki, Raimundo Sales de Oliveira-Neto, Gabriel Nakamura Ramos de Luca, Ricardo Gariba da Silva, Guilherme Ferreira da Silva, Bruno Carvalho de Vasconcelos, Rodrigo Ricci Vivan, Murilo Priori Alcalde, Marco Antonio Hungaro Duarte","doi":"10.4317/jced.64103","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>This study aimed to evaluate the accuracy of four hybrid endodontic motors-VDW Gold (VDW), iRoot Pro (IRP), E-Connect S (ECS), and TriAuto ZX2 (TZX)-in maintaining length, speed, and angulation during alternating and continuous rotation kinematics and the influence of these parameters on instrument fatigue.</p><p><strong>Materials and methods: </strong>Forty-eight mandibular incisors (n=12 per group) were used to assess apex locator accuracy during glide path procedures with alternating rotation and auto apical stop (AAS) set to the apical foramen (AF: 0.0). Positioning precision was evaluated using micro-CT, while torsional resistance of instruments was tested per ISO 3630-1. Actual speed and angulation were measured using a tachometer and Arduino system. Cyclic fatigue testing was conducted with Only One File 25.08 instruments in an artificial curved canal (60°, 5 mm radius). Data were statistically analyzed with a significance level of 5%.</p><p><strong>Results: </strong>The evaluated devices exhibited similar precision (p>0.05). The TZX showed the lowest number of determinations beyond the AF (p<0.05). Regarding torsional testing, no significant differences were observed between groups (p>0.05). The equipment showed substantial differences compared to the values displayed on the screen (p<0.05). The Optimum Glide Path (OGP) motion provided significantly higher resistance to cyclic fatigue (p<0.05).</p><p><strong>Conclusions: </strong>All hybrid motors were accurate in length control, with TZX being the most conservative. Kinematics did not influence torsional resistance, but significant differences were found between programmed and real parameters.</p>","PeriodicalId":15376,"journal":{"name":"Journal of Clinical and Experimental Dentistry","volume":"18 7","pages":"e921-e928"},"PeriodicalIF":0.0000,"publicationDate":"2026-06-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13386448/pdf/","citationCount":"0","resultStr":"{\"title\":\"In vitro assessment of integrated apex locator systems: working length accuracy, operational parameters, and mechanical effects on NiTi instruments.\",\"authors\":\"Roberto Barreto Osaki, Raimundo Sales de Oliveira-Neto, Gabriel Nakamura Ramos de Luca, Ricardo Gariba da Silva, Guilherme Ferreira da Silva, Bruno Carvalho de Vasconcelos, Rodrigo Ricci Vivan, Murilo Priori Alcalde, Marco Antonio Hungaro Duarte\",\"doi\":\"10.4317/jced.64103\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>This study aimed to evaluate the accuracy of four hybrid endodontic motors-VDW Gold (VDW), iRoot Pro (IRP), E-Connect S (ECS), and TriAuto ZX2 (TZX)-in maintaining length, speed, and angulation during alternating and continuous rotation kinematics and the influence of these parameters on instrument fatigue.</p><p><strong>Materials and methods: </strong>Forty-eight mandibular incisors (n=12 per group) were used to assess apex locator accuracy during glide path procedures with alternating rotation and auto apical stop (AAS) set to the apical foramen (AF: 0.0). Positioning precision was evaluated using micro-CT, while torsional resistance of instruments was tested per ISO 3630-1. Actual speed and angulation were measured using a tachometer and Arduino system. Cyclic fatigue testing was conducted with Only One File 25.08 instruments in an artificial curved canal (60°, 5 mm radius). Data were statistically analyzed with a significance level of 5%.</p><p><strong>Results: </strong>The evaluated devices exhibited similar precision (p>0.05). The TZX showed the lowest number of determinations beyond the AF (p<0.05). Regarding torsional testing, no significant differences were observed between groups (p>0.05). The equipment showed substantial differences compared to the values displayed on the screen (p<0.05). The Optimum Glide Path (OGP) motion provided significantly higher resistance to cyclic fatigue (p<0.05).</p><p><strong>Conclusions: </strong>All hybrid motors were accurate in length control, with TZX being the most conservative. Kinematics did not influence torsional resistance, but significant differences were found between programmed and real parameters.</p>\",\"PeriodicalId\":15376,\"journal\":{\"name\":\"Journal of Clinical and Experimental Dentistry\",\"volume\":\"18 7\",\"pages\":\"e921-e928\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2026-06-29\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13386448/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Clinical and Experimental Dentistry\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.4317/jced.64103\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2026/6/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"Q2\",\"JCRName\":\"Dentistry\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Clinical and Experimental Dentistry","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4317/jced.64103","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/6/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"Dentistry","Score":null,"Total":0}
引用次数: 0
摘要
背景:本研究旨在评估四种混合牙根电机-VDW Gold (VDW), iRoot Pro (IRP), E-Connect S (ECS)和TriAuto ZX2 (TZX)-在交替和连续旋转运动中保持长度,速度和角度的准确性以及这些参数对仪器疲劳的影响。材料和方法:采用48个下颌骨切牙(每组12个),在交替旋转和自动根尖停止(AAS)设置于根尖孔(AF: 0.0)的滑动路径手术中评估根尖定位器的准确性。使用micro-CT评估定位精度,同时根据ISO 3630-1测试仪器的抗扭性。实际速度和角度测量使用转速计和Arduino系统。使用One File 25.08仪器在人工弯曲管(60°,5 mm半径)中进行循环疲劳试验。对数据进行统计学分析,显著性水平为5%。结果:所评估的设备具有相似的精度(p>0.05)。TZX在AF之外的检测次数最少(p<0.05)。扭转试验方面,两组间无显著差异(p>0.05)。与屏幕上显示的数值相比,设备显示出实质性差异(p<0.05)。最佳滑行路径(OGP)运动对循环疲劳的抵抗能力显著提高(p < 0.05)。结论:所有混合动力马达长度控制准确,其中TZX最保守。运动学不影响扭转阻力,但程序参数和实际参数之间存在显著差异。
In vitro assessment of integrated apex locator systems: working length accuracy, operational parameters, and mechanical effects on NiTi instruments.
Background: This study aimed to evaluate the accuracy of four hybrid endodontic motors-VDW Gold (VDW), iRoot Pro (IRP), E-Connect S (ECS), and TriAuto ZX2 (TZX)-in maintaining length, speed, and angulation during alternating and continuous rotation kinematics and the influence of these parameters on instrument fatigue.
Materials and methods: Forty-eight mandibular incisors (n=12 per group) were used to assess apex locator accuracy during glide path procedures with alternating rotation and auto apical stop (AAS) set to the apical foramen (AF: 0.0). Positioning precision was evaluated using micro-CT, while torsional resistance of instruments was tested per ISO 3630-1. Actual speed and angulation were measured using a tachometer and Arduino system. Cyclic fatigue testing was conducted with Only One File 25.08 instruments in an artificial curved canal (60°, 5 mm radius). Data were statistically analyzed with a significance level of 5%.
Results: The evaluated devices exhibited similar precision (p>0.05). The TZX showed the lowest number of determinations beyond the AF (p<0.05). Regarding torsional testing, no significant differences were observed between groups (p>0.05). The equipment showed substantial differences compared to the values displayed on the screen (p<0.05). The Optimum Glide Path (OGP) motion provided significantly higher resistance to cyclic fatigue (p<0.05).
Conclusions: All hybrid motors were accurate in length control, with TZX being the most conservative. Kinematics did not influence torsional resistance, but significant differences were found between programmed and real parameters.
期刊介绍:
Indexed in PUBMED, PubMed Central® (PMC) since 2012 and SCOPUSJournal of Clinical and Experimental Dentistry is an Open Access (free access on-line) - http://www.medicinaoral.com/odo/indice.htm. The aim of the Journal of Clinical and Experimental Dentistry is: - Periodontology - Community and Preventive Dentistry - Esthetic Dentistry - Biomaterials and Bioengineering in Dentistry - Operative Dentistry and Endodontics - Prosthetic Dentistry - Orthodontics - Oral Medicine and Pathology - Odontostomatology for the disabled or special patients - Oral Surgery