用于评估小动物SPECT和PET冷热区域图像质量和部分体积效应的模型的开发。

IF 1.5 Q4 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Takayuki Shibutani, Masahisa Onoguchi
{"title":"用于评估小动物SPECT和PET冷热区域图像质量和部分体积效应的模型的开发。","authors":"Takayuki Shibutani, Masahisa Onoguchi","doi":"10.2967/jnmt.125.271314","DOIUrl":null,"url":null,"abstract":"<p><p>The partial-volume effect (PVE) is a major factor affecting quantitative accuracy in small-animal SPECT and PET imaging, leading to underestimation of activity in hot-rod images and overestimation in cold-rod images. However, a phantom capable of assessing image quality, including the PVE in both hot and cold regions, has not been available. This study aimed to determine whether a newly developed phantom can accurately evaluate image quality, including the PVE of hot- and cold-rod images, in small-animal SPECT and PET systems. <b>Methods:</b> The phantom consisted of a cylindric pool and rod sections. The rod section included 6 hot rods and 6 cold rods with diameters of 1-6 mm. The rod and pool sections were filled with solutions containing <sup>99m</sup>Tc and <sup>18</sup>F, respectively. In the hot-rod section, the radioactive concentration ratio between rods and background was set to 4:1. Images were acquired for 30 min and reconstructed with a voxel size of 0.8 mm using a pixel-based ordered-subset expectation maximization algorithm. Image contrast, absolute and inverse recovery coefficients (RCs) (indicators of the PVE), and percent coefficient of variation (%CV) and signal-to-noise ratio (indicators of uniformity) were evaluated for both hot and cold rods. <b>Results:</b> The contrast of the hot rod for <sup>99m</sup>Tc and <sup>18</sup>F showed lower values with decreasing rod diameters. Furthermore, the <sup>99m</sup>Tc image demonstrated a higher contrast than the <sup>18</sup>F image and approached the true contrast. The cold-rod contrasts with <sup>99m</sup>Tc and <sup>18</sup>F followed a similar trend. The RCs of <sup>99m</sup>Tc and <sup>18</sup>F were lower as rod diameters decreased. The radioactive concentration in the 40-mm pool section was similar to that of the 6-mm hot rod for both radionuclides. The inverse RC was lower with a decreasing cold-rod diameter. Moreover, cold-rod images with <sup>18</sup>F demonstrated lower inverse RCs than with <sup>99m</sup>Tc. The %CVs for the <sup>99m</sup>Tc and <sup>18</sup>F images were 3.49% ± 0.03% and 6.29% ± 0.07%, respectively, with the <sup>99m</sup>Tc image displaying a lower %CV compared with the <sup>18</sup>F image. <b>Conclusion:</b> We developed a phantom that allows the evaluation of physical phenomena in small-animal SPECT and PET images, providing reliable assessment of image contrast, PVE, and uniformity in both hot and cold rods.</p>","PeriodicalId":16548,"journal":{"name":"Journal of nuclear medicine technology","volume":" ","pages":""},"PeriodicalIF":1.5000,"publicationDate":"2026-06-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Development of a Phantom for Evaluating Image Quality and Partial-Volume Effects in Hot and Cold Regions in Small-Animal SPECT and PET.\",\"authors\":\"Takayuki Shibutani, Masahisa Onoguchi\",\"doi\":\"10.2967/jnmt.125.271314\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The partial-volume effect (PVE) is a major factor affecting quantitative accuracy in small-animal SPECT and PET imaging, leading to underestimation of activity in hot-rod images and overestimation in cold-rod images. However, a phantom capable of assessing image quality, including the PVE in both hot and cold regions, has not been available. This study aimed to determine whether a newly developed phantom can accurately evaluate image quality, including the PVE of hot- and cold-rod images, in small-animal SPECT and PET systems. <b>Methods:</b> The phantom consisted of a cylindric pool and rod sections. The rod section included 6 hot rods and 6 cold rods with diameters of 1-6 mm. The rod and pool sections were filled with solutions containing <sup>99m</sup>Tc and <sup>18</sup>F, respectively. In the hot-rod section, the radioactive concentration ratio between rods and background was set to 4:1. Images were acquired for 30 min and reconstructed with a voxel size of 0.8 mm using a pixel-based ordered-subset expectation maximization algorithm. Image contrast, absolute and inverse recovery coefficients (RCs) (indicators of the PVE), and percent coefficient of variation (%CV) and signal-to-noise ratio (indicators of uniformity) were evaluated for both hot and cold rods. <b>Results:</b> The contrast of the hot rod for <sup>99m</sup>Tc and <sup>18</sup>F showed lower values with decreasing rod diameters. Furthermore, the <sup>99m</sup>Tc image demonstrated a higher contrast than the <sup>18</sup>F image and approached the true contrast. The cold-rod contrasts with <sup>99m</sup>Tc and <sup>18</sup>F followed a similar trend. The RCs of <sup>99m</sup>Tc and <sup>18</sup>F were lower as rod diameters decreased. The radioactive concentration in the 40-mm pool section was similar to that of the 6-mm hot rod for both radionuclides. The inverse RC was lower with a decreasing cold-rod diameter. Moreover, cold-rod images with <sup>18</sup>F demonstrated lower inverse RCs than with <sup>99m</sup>Tc. The %CVs for the <sup>99m</sup>Tc and <sup>18</sup>F images were 3.49% ± 0.03% and 6.29% ± 0.07%, respectively, with the <sup>99m</sup>Tc image displaying a lower %CV compared with the <sup>18</sup>F image. <b>Conclusion:</b> We developed a phantom that allows the evaluation of physical phenomena in small-animal SPECT and PET images, providing reliable assessment of image contrast, PVE, and uniformity in both hot and cold rods.</p>\",\"PeriodicalId\":16548,\"journal\":{\"name\":\"Journal of nuclear medicine technology\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":1.5000,\"publicationDate\":\"2026-06-23\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of nuclear medicine technology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.2967/jnmt.125.271314\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of nuclear medicine technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2967/jnmt.125.271314","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

部分体积效应(PVE)是影响小动物SPECT和PET成像定量准确性的主要因素,导致热棒图像活动低估和冷棒图像活动高估。然而,目前还没有能够评估图像质量的幻影,包括在冷热地区的PVE。本研究旨在确定新开发的假体能否准确评估小动物SPECT和PET系统中的图像质量,包括热棒和冷棒图像的PVE。方法:假体由圆柱形池和杆状段组成。棒段包括6根热棒和6根冷棒,直径为1- 6mm。棒段和池段分别填充含有99mTc和18F的溶液。在热棒部分,棒与本底的放射性浓度比设置为4:1。图像采集时间为30 min,使用基于像素的有序子集期望最大化算法重构体素大小为0.8 mm的图像。对热棒和冷棒的图像对比度、绝对恢复系数和逆恢复系数(RCs) (PVE指标)、百分比变异系数(%CV)和信噪比(均匀性指标)进行了评估。结果:99mTc和18F热棒的对比值随热棒直径的减小而降低。99mTc图像的对比度高于18F图像,接近真实对比度。冷棒与99mTc和18F的对比也遵循类似的趋势。99mTc和18F的rc随杆径的减小而降低。两种放射性核素在40 mm池段的放射性浓度与6 mm热棒段的放射性浓度相似。逆RC随冷棒直径的减小而降低。此外,与99mTc相比,18F冷棒图像显示出更低的逆RCs。99mTc和18F图像的%CV分别为3.49%±0.03%和6.29%±0.07%,其中99mTc图像的%CV较18F图像低。结论:我们开发了一种可以评估小动物SPECT和PET图像物理现象的模体,为热棒和冷棒的图像对比度、PVE和均匀性提供可靠的评估。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Development of a Phantom for Evaluating Image Quality and Partial-Volume Effects in Hot and Cold Regions in Small-Animal SPECT and PET.

The partial-volume effect (PVE) is a major factor affecting quantitative accuracy in small-animal SPECT and PET imaging, leading to underestimation of activity in hot-rod images and overestimation in cold-rod images. However, a phantom capable of assessing image quality, including the PVE in both hot and cold regions, has not been available. This study aimed to determine whether a newly developed phantom can accurately evaluate image quality, including the PVE of hot- and cold-rod images, in small-animal SPECT and PET systems. Methods: The phantom consisted of a cylindric pool and rod sections. The rod section included 6 hot rods and 6 cold rods with diameters of 1-6 mm. The rod and pool sections were filled with solutions containing 99mTc and 18F, respectively. In the hot-rod section, the radioactive concentration ratio between rods and background was set to 4:1. Images were acquired for 30 min and reconstructed with a voxel size of 0.8 mm using a pixel-based ordered-subset expectation maximization algorithm. Image contrast, absolute and inverse recovery coefficients (RCs) (indicators of the PVE), and percent coefficient of variation (%CV) and signal-to-noise ratio (indicators of uniformity) were evaluated for both hot and cold rods. Results: The contrast of the hot rod for 99mTc and 18F showed lower values with decreasing rod diameters. Furthermore, the 99mTc image demonstrated a higher contrast than the 18F image and approached the true contrast. The cold-rod contrasts with 99mTc and 18F followed a similar trend. The RCs of 99mTc and 18F were lower as rod diameters decreased. The radioactive concentration in the 40-mm pool section was similar to that of the 6-mm hot rod for both radionuclides. The inverse RC was lower with a decreasing cold-rod diameter. Moreover, cold-rod images with 18F demonstrated lower inverse RCs than with 99mTc. The %CVs for the 99mTc and 18F images were 3.49% ± 0.03% and 6.29% ± 0.07%, respectively, with the 99mTc image displaying a lower %CV compared with the 18F image. Conclusion: We developed a phantom that allows the evaluation of physical phenomena in small-animal SPECT and PET images, providing reliable assessment of image contrast, PVE, and uniformity in both hot and cold rods.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of nuclear medicine technology
Journal of nuclear medicine technology RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
1.90
自引率
15.40%
发文量
57
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书