经认证的68Ge PET圆柱体模体中的不均匀性:标准化质量保证的含义。

IF 1.5 Q4 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Subhash Kheruka, Sana Al-Rashdi, Naema Al-Maymani, Anas Al-Balushi, Tasnim Al-Raii, Noura Al-Makhmari, Vipin Jayakrishnan, Anjali Jain, Khulood Al-Riyami, Rashid Al-Sukaiti
{"title":"经认证的68Ge PET圆柱体模体中的不均匀性:标准化质量保证的含义。","authors":"Subhash Kheruka, Sana Al-Rashdi, Naema Al-Maymani, Anas Al-Balushi, Tasnim Al-Raii, Noura Al-Makhmari, Vipin Jayakrishnan, Anjali Jain, Khulood Al-Riyami, Rashid Al-Sukaiti","doi":"10.2967/jnmt.125.271634","DOIUrl":null,"url":null,"abstract":"<p><p>Quantitative PET imaging depends critically on the accuracy of system normalization and uniformity calibration. We report a manufacturing defect in a certified RadQual <sup>68</sup>Ge PET cylinder phantom (BMCY68-2030A-03), a standard reference used for normalization and daily quality control in Siemens Biograph Vision scanners. A localized hot spot was identified within the phantom, producing integral uniformity greater than 8.5% (acceptance limit ≤5%). The nonuniformity was intrinsic to the phantom material and rendered it unsuitable for quantitative use. This finding underscores the importance of independent acceptance testing of long-lived <sup>68</sup>Ge standards before clinical deployment to prevent calibration bias and ensure reliable SUV quantification.</p>","PeriodicalId":16548,"journal":{"name":"Journal of nuclear medicine technology","volume":" ","pages":""},"PeriodicalIF":1.5000,"publicationDate":"2026-06-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Nonuniformity in a Certified <sup>68</sup>Ge PET Cylinder Phantom: Implications for Normalization Quality Assurance.\",\"authors\":\"Subhash Kheruka, Sana Al-Rashdi, Naema Al-Maymani, Anas Al-Balushi, Tasnim Al-Raii, Noura Al-Makhmari, Vipin Jayakrishnan, Anjali Jain, Khulood Al-Riyami, Rashid Al-Sukaiti\",\"doi\":\"10.2967/jnmt.125.271634\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Quantitative PET imaging depends critically on the accuracy of system normalization and uniformity calibration. We report a manufacturing defect in a certified RadQual <sup>68</sup>Ge PET cylinder phantom (BMCY68-2030A-03), a standard reference used for normalization and daily quality control in Siemens Biograph Vision scanners. A localized hot spot was identified within the phantom, producing integral uniformity greater than 8.5% (acceptance limit ≤5%). The nonuniformity was intrinsic to the phantom material and rendered it unsuitable for quantitative use. This finding underscores the importance of independent acceptance testing of long-lived <sup>68</sup>Ge standards before clinical deployment to prevent calibration bias and ensure reliable SUV quantification.</p>\",\"PeriodicalId\":16548,\"journal\":{\"name\":\"Journal of nuclear medicine technology\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":1.5000,\"publicationDate\":\"2026-06-23\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of nuclear medicine technology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.2967/jnmt.125.271634\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of nuclear medicine technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2967/jnmt.125.271634","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

定量PET成像关键取决于系统归一化和均匀性校准的准确性。我们报告了经认证的RadQual 68Ge PET圆柱体(BMCY68-2030A-03)的制造缺陷,该圆柱体是西门子Biograph Vision扫描仪规范化和日常质量控制的标准参考。在体模内发现局部热点,整体均匀度大于8.5%(接受极限≤5%)。这种不均匀性是模体材料固有的,不适合定量使用。这一发现强调了在临床部署之前对长寿命68Ge标准进行独立验收测试的重要性,以防止校准偏差并确保可靠的SUV量化。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Nonuniformity in a Certified 68Ge PET Cylinder Phantom: Implications for Normalization Quality Assurance.

Quantitative PET imaging depends critically on the accuracy of system normalization and uniformity calibration. We report a manufacturing defect in a certified RadQual 68Ge PET cylinder phantom (BMCY68-2030A-03), a standard reference used for normalization and daily quality control in Siemens Biograph Vision scanners. A localized hot spot was identified within the phantom, producing integral uniformity greater than 8.5% (acceptance limit ≤5%). The nonuniformity was intrinsic to the phantom material and rendered it unsuitable for quantitative use. This finding underscores the importance of independent acceptance testing of long-lived 68Ge standards before clinical deployment to prevent calibration bias and ensure reliable SUV quantification.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of nuclear medicine technology
Journal of nuclear medicine technology RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
1.90
自引率
15.40%
发文量
57
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书