根据国际标准化组织81060-2:2018对Wellvii BP Go血压监测仪进行临床评估。

IF 1 4区 医学 Q4 PERIPHERAL VASCULAR DISEASE
Blood Pressure Monitoring Pub Date : 2026-10-01 Epub Date: 2026-09-02 DOI:10.1097/MBP.0000000000000805
Jialiang Ma, Yao Xiao, Yage Wang, Jieqiong Zhang, Yi Wan
{"title":"根据国际标准化组织81060-2:2018对Wellvii BP Go血压监测仪进行临床评估。","authors":"Jialiang Ma, Yao Xiao, Yage Wang, Jieqiong Zhang, Yi Wan","doi":"10.1097/MBP.0000000000000805","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>This study aimed to evaluate the accuracy of the Wellvii BP Go automated oscillometric finger blood pressure (BP) monitor for self/home BP measurement according to the International Standardization Organization (ISO) 81060-2 : 2018.</p><p><strong>Methods: </strong>A total of 85 eligible participants was included in the study to fulfill the age, sex, and BP distribution criteria of the ISO standard in the general population. The same arm/finger sequential measurement method was used for comparison with a standard mercury sphygmomanometer, and the assessment were in accordance with the ISO 81060-2 : 2018 protocol.</p><p><strong>Results: </strong>A total of 255 comparison pairs were obtained for analysis. For validation Criterion 1, the mean ± SD of the differences between the test device and reference systolic BP/diastolic BP readings was 2.12 ± 7.33 and 1.30 ± 6.35 mmHg. For validation Criterion 2, the SD of the averaged systolic BP/diastolic BP differences between the test device and reference BP per participant was 6.58/5.60 mmHg, respectively.</p><p><strong>Conclusion: </strong>The Wellvii BP Go automated finger BP monitor has passed the requirements of the ISO 81060-2 : 2018, and can be recommended for self/home BP measurement in general population.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":" ","pages":"243-247"},"PeriodicalIF":1.0000,"publicationDate":"2026-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Clinical evaluation of the Wellvii BP Go blood pressure monitor according to the International Standardization Organization 81060-2:2018.\",\"authors\":\"Jialiang Ma, Yao Xiao, Yage Wang, Jieqiong Zhang, Yi Wan\",\"doi\":\"10.1097/MBP.0000000000000805\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>This study aimed to evaluate the accuracy of the Wellvii BP Go automated oscillometric finger blood pressure (BP) monitor for self/home BP measurement according to the International Standardization Organization (ISO) 81060-2 : 2018.</p><p><strong>Methods: </strong>A total of 85 eligible participants was included in the study to fulfill the age, sex, and BP distribution criteria of the ISO standard in the general population. The same arm/finger sequential measurement method was used for comparison with a standard mercury sphygmomanometer, and the assessment were in accordance with the ISO 81060-2 : 2018 protocol.</p><p><strong>Results: </strong>A total of 255 comparison pairs were obtained for analysis. For validation Criterion 1, the mean ± SD of the differences between the test device and reference systolic BP/diastolic BP readings was 2.12 ± 7.33 and 1.30 ± 6.35 mmHg. For validation Criterion 2, the SD of the averaged systolic BP/diastolic BP differences between the test device and reference BP per participant was 6.58/5.60 mmHg, respectively.</p><p><strong>Conclusion: </strong>The Wellvii BP Go automated finger BP monitor has passed the requirements of the ISO 81060-2 : 2018, and can be recommended for self/home BP measurement in general population.</p>\",\"PeriodicalId\":8950,\"journal\":{\"name\":\"Blood Pressure Monitoring\",\"volume\":\" \",\"pages\":\"243-247\"},\"PeriodicalIF\":1.0000,\"publicationDate\":\"2026-10-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Blood Pressure Monitoring\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1097/MBP.0000000000000805\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2026/9/2 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q4\",\"JCRName\":\"PERIPHERAL VASCULAR DISEASE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000805","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/9/2 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0

摘要

目的:本研究旨在根据国际标准化组织(ISO) 81060- 2:20 18,评估用于自我/家庭血压测量的Wellvii BP Go自动振荡手指血压(BP)监测仪的准确性。方法:共纳入85名符合年龄、性别、血压分布标准的普通人群。采用相同的手臂/手指顺序测量方法与标准水银血压计进行比较,评估符合ISO 81060- 2:20 18方案。结果:共获得255对对照对进行分析。对于验证标准1,测试装置与参考收缩压/舒张压读数之间差异的平均值±SD为2.12±7.33和1.30±6.35 mmHg。对于验证标准2,每位参与者的测试装置和参考血压之间的平均收缩压/舒张压差的SD分别为6.58/5.60 mmHg。结论:Wellvii BP Go自动手指血压监测仪已通过ISO 81060- 2:20 18的要求,可推荐用于普通人群的自我/家庭血压测量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Clinical evaluation of the Wellvii BP Go blood pressure monitor according to the International Standardization Organization 81060-2:2018.

Objective: This study aimed to evaluate the accuracy of the Wellvii BP Go automated oscillometric finger blood pressure (BP) monitor for self/home BP measurement according to the International Standardization Organization (ISO) 81060-2 : 2018.

Methods: A total of 85 eligible participants was included in the study to fulfill the age, sex, and BP distribution criteria of the ISO standard in the general population. The same arm/finger sequential measurement method was used for comparison with a standard mercury sphygmomanometer, and the assessment were in accordance with the ISO 81060-2 : 2018 protocol.

Results: A total of 255 comparison pairs were obtained for analysis. For validation Criterion 1, the mean ± SD of the differences between the test device and reference systolic BP/diastolic BP readings was 2.12 ± 7.33 and 1.30 ± 6.35 mmHg. For validation Criterion 2, the SD of the averaged systolic BP/diastolic BP differences between the test device and reference BP per participant was 6.58/5.60 mmHg, respectively.

Conclusion: The Wellvii BP Go automated finger BP monitor has passed the requirements of the ISO 81060-2 : 2018, and can be recommended for self/home BP measurement in general population.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Blood Pressure Monitoring
Blood Pressure Monitoring 医学-外周血管病
CiteScore
2.00
自引率
7.70%
发文量
110
审稿时长
>12 weeks
期刊介绍: Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research. This innovative journal contains papers dealing with all aspects of manual, automated, and ambulatory monitoring. Basic and clinical science papers are considered although the emphasis is on clinical medicine. Submitted articles undergo a preliminary review by the editor. Some articles may be returned to authors without further consideration. Those being considered for publication will undergo further assessment and peer-review by the editors and those invited to do so from a reviewer pool.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书