{"title":"离心转速及其对诊断准确性影响的比较评价。","authors":"Mohammed A Jeraiby, Jobran M Moshi, Sajad A Dar, Mahmoud M Habibullah, Mjery Yahia, Saif Elden B Abdalla, Raheem M Balkhtab, Renad A Saigh, Nahlah S Burayk","doi":"10.1002/jcla.70120","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Timely laboratory results are essential for effective clinical decision-making, with turnaround time (TAT) serving as a key performance indicator. Pre-analytical steps like centrifugation significantly impact TAT. This study investigates whether a shorter, high-speed centrifugation can reduce TAT while maintaining the analytical accuracy required for routine diagnostic testing.</p><p><strong>Methodology: </strong>This comparative experimental study analyzed blood samples from patients using plasma and serum tubes. Two centrifugation methods (A: (routine) 10 min at 3200 g; B: (faster) 5 min at 4000 g) were compared across chemistry, immunochemistry, coagulation, and interference indices.</p><p><strong>Results: </strong>Most analytes showed no significant difference between the two methods, indicating strong analytical agreement. However, significant differences were observed in sodium (p = 0.003), CKMB (p = 0.044), ALP (p = 0.001), ALT (p = 0.011), FRT4 (p < 0.001), PT (p = 0.008), and INR (p = 0.003). Despite these differences, all values remained within clinically acceptable ranges. Hemolysis rates decreased notably under the high-speed method, dropping from 10.4% to 5.1%. Bias analysis across tube types revealed greater variability for CKMB and ALT under the high-speed method (bias = 2.76 and -2.92; CVs = 68.8% and 84.16%, respectively).</p><p><strong>Conclusion: </strong>High-speed centrifugation significantly reduces processing time without compromising clinical reliability. Despite minor analyte differences, all values ramined within acceptable clinical ranges. Furthermore, the observed reduction in hemolysis rates strengthens the case for adopting the high-speed method as a reliable, efficient, and clinically safe alternative for routine laboratory workflows.</p>","PeriodicalId":15509,"journal":{"name":"Journal of Clinical Laboratory Analysis","volume":" ","pages":"e70120"},"PeriodicalIF":2.9000,"publicationDate":"2025-10-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Comparative Evaluation of Centrifugation Speed and Its Impact on Diagnostic Accuracy.\",\"authors\":\"Mohammed A Jeraiby, Jobran M Moshi, Sajad A Dar, Mahmoud M Habibullah, Mjery Yahia, Saif Elden B Abdalla, Raheem M Balkhtab, Renad A Saigh, Nahlah S Burayk\",\"doi\":\"10.1002/jcla.70120\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Introduction: </strong>Timely laboratory results are essential for effective clinical decision-making, with turnaround time (TAT) serving as a key performance indicator. Pre-analytical steps like centrifugation significantly impact TAT. This study investigates whether a shorter, high-speed centrifugation can reduce TAT while maintaining the analytical accuracy required for routine diagnostic testing.</p><p><strong>Methodology: </strong>This comparative experimental study analyzed blood samples from patients using plasma and serum tubes. Two centrifugation methods (A: (routine) 10 min at 3200 g; B: (faster) 5 min at 4000 g) were compared across chemistry, immunochemistry, coagulation, and interference indices.</p><p><strong>Results: </strong>Most analytes showed no significant difference between the two methods, indicating strong analytical agreement. However, significant differences were observed in sodium (p = 0.003), CKMB (p = 0.044), ALP (p = 0.001), ALT (p = 0.011), FRT4 (p < 0.001), PT (p = 0.008), and INR (p = 0.003). Despite these differences, all values remained within clinically acceptable ranges. Hemolysis rates decreased notably under the high-speed method, dropping from 10.4% to 5.1%. Bias analysis across tube types revealed greater variability for CKMB and ALT under the high-speed method (bias = 2.76 and -2.92; CVs = 68.8% and 84.16%, respectively).</p><p><strong>Conclusion: </strong>High-speed centrifugation significantly reduces processing time without compromising clinical reliability. Despite minor analyte differences, all values ramined within acceptable clinical ranges. Furthermore, the observed reduction in hemolysis rates strengthens the case for adopting the high-speed method as a reliable, efficient, and clinically safe alternative for routine laboratory workflows.</p>\",\"PeriodicalId\":15509,\"journal\":{\"name\":\"Journal of Clinical Laboratory Analysis\",\"volume\":\" \",\"pages\":\"e70120\"},\"PeriodicalIF\":2.9000,\"publicationDate\":\"2025-10-09\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Clinical Laboratory Analysis\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1002/jcla.70120\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Clinical Laboratory Analysis","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/jcla.70120","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

及时的实验室结果对有效的临床决策至关重要,周转时间(TAT)是一个关键的绩效指标。像离心这样的分析前步骤会显著影响TAT。本研究探讨了较短的高速离心是否可以在保持常规诊断测试所需的分析准确性的同时降低TAT。方法:本对比实验研究采用血浆和血清管对患者的血液样本进行分析。两种离心方法:(常规)3200g离心10min;B:(更快)5min, 4000g)在化学、免疫化学、凝血和干扰指标上进行比较。结果:两种方法的大多数分析结果无显著差异,表明分析一致性强。然而,在钠(p = 0.003)、CKMB (p = 0.044)、ALP (p = 0.001)、ALT (p = 0.011)、FRT4 (p)等指标上观察到显著差异。结论:高速离心可显著缩短处理时间,且不影响临床可靠性。尽管分析物有微小差异,但所有值都在可接受的临床范围内。此外,观察到的溶血率降低加强了采用高速方法作为常规实验室工作流程可靠,高效和临床安全的替代方案的案例。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Comparative Evaluation of Centrifugation Speed and Its Impact on Diagnostic Accuracy.

Introduction: Timely laboratory results are essential for effective clinical decision-making, with turnaround time (TAT) serving as a key performance indicator. Pre-analytical steps like centrifugation significantly impact TAT. This study investigates whether a shorter, high-speed centrifugation can reduce TAT while maintaining the analytical accuracy required for routine diagnostic testing.

Methodology: This comparative experimental study analyzed blood samples from patients using plasma and serum tubes. Two centrifugation methods (A: (routine) 10 min at 3200 g; B: (faster) 5 min at 4000 g) were compared across chemistry, immunochemistry, coagulation, and interference indices.

Results: Most analytes showed no significant difference between the two methods, indicating strong analytical agreement. However, significant differences were observed in sodium (p = 0.003), CKMB (p = 0.044), ALP (p = 0.001), ALT (p = 0.011), FRT4 (p < 0.001), PT (p = 0.008), and INR (p = 0.003). Despite these differences, all values remained within clinically acceptable ranges. Hemolysis rates decreased notably under the high-speed method, dropping from 10.4% to 5.1%. Bias analysis across tube types revealed greater variability for CKMB and ALT under the high-speed method (bias = 2.76 and -2.92; CVs = 68.8% and 84.16%, respectively).

Conclusion: High-speed centrifugation significantly reduces processing time without compromising clinical reliability. Despite minor analyte differences, all values ramined within acceptable clinical ranges. Furthermore, the observed reduction in hemolysis rates strengthens the case for adopting the high-speed method as a reliable, efficient, and clinically safe alternative for routine laboratory workflows.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Clinical Laboratory Analysis
Journal of Clinical Laboratory Analysis 医学-医学实验技术
CiteScore
5.60
自引率
7.40%
发文量
584
审稿时长
6-12 weeks
期刊介绍: Journal of Clinical Laboratory Analysis publishes original articles on newly developing modes of technology and laboratory assays, with emphasis on their application in current and future clinical laboratory testing. This includes reports from the following fields: immunochemistry and toxicology, hematology and hematopathology, immunopathology, molecular diagnostics, microbiology, genetic testing, immunohematology, and clinical chemistry.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信