{"title":"用横波弹性成像测量肝脏硬度的可靠性:皮肤到包膜距离和皮下组织结构的影响。","authors":"Keisuke Osakabe, Keiko Sugimoto, Hiroji Takai, Yusuke Sano, Keisuke Maeda, Koji Suzuki, Naohiro Ichino","doi":"10.1007/s10396-025-01582-9","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Shear wave velocity (Vs) measurements can be unstable in individuals with obesity or with a skin-capsule distance (SCD) of ≥ 20 mm. However, instability of Vs has also been observed in cases where SCD is not markedly increased; therefore, the present study aimed to investigate the factors contributing to reduced VsN (net amount of effective shear wave velocity), an index that reflects the reliability of individual Vs measurements.</p><p><strong>Methods: </strong>In total, 317 individuals who underwent community-based health screening were included in this study. We investigated the relationship between VsN and physical findings, laboratory data, SCD, and structure of the SCD (single vs. multilayered) to identify factors associated with decreased VsN.</p><p><strong>Results: </strong>Multiple regression analysis revealed that body fat percentage (β = - 0.135, p = 0.011), SCD (β = - 0.347, p < 0.0001), and the structure of the SC region (β = - 0.295, p < 0.0001) were independently associated with decreased VsN. In cases where the SCD exceeded 20 mm, the VsN within the SCD was significantly lower in the multilayer structural type [52 (28-75)] than in the single-layer structural type [95 (77-99)] (p = 0.004). Furthermore, even when the SCD was 20 mm or less, the VsN within the SCD was significantly lower in the multilayer structure type [80 (68-89)] than in the single-layer structure type [99 (92-100)] (p < 0.0001).</p><p><strong>Conclusion: </strong>The results show that, in addition to SCD thickness, its internal structure reduces the VsN and contributes to Vs measurement instability.</p>","PeriodicalId":50130,"journal":{"name":"Journal of Medical Ultrasonics","volume":" ","pages":""},"PeriodicalIF":2.1000,"publicationDate":"2025-10-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Reliability of liver stiffness measurements using shear wave elastography: impact of skin-to-capsule distance and subcutaneous tissue structure.\",\"authors\":\"Keisuke Osakabe, Keiko Sugimoto, Hiroji Takai, Yusuke Sano, Keisuke Maeda, Koji Suzuki, Naohiro Ichino\",\"doi\":\"10.1007/s10396-025-01582-9\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>Shear wave velocity (Vs) measurements can be unstable in individuals with obesity or with a skin-capsule distance (SCD) of ≥ 20 mm. However, instability of Vs has also been observed in cases where SCD is not markedly increased; therefore, the present study aimed to investigate the factors contributing to reduced VsN (net amount of effective shear wave velocity), an index that reflects the reliability of individual Vs measurements.</p><p><strong>Methods: </strong>In total, 317 individuals who underwent community-based health screening were included in this study. We investigated the relationship between VsN and physical findings, laboratory data, SCD, and structure of the SCD (single vs. multilayered) to identify factors associated with decreased VsN.</p><p><strong>Results: </strong>Multiple regression analysis revealed that body fat percentage (β = - 0.135, p = 0.011), SCD (β = - 0.347, p < 0.0001), and the structure of the SC region (β = - 0.295, p < 0.0001) were independently associated with decreased VsN. In cases where the SCD exceeded 20 mm, the VsN within the SCD was significantly lower in the multilayer structural type [52 (28-75)] than in the single-layer structural type [95 (77-99)] (p = 0.004). Furthermore, even when the SCD was 20 mm or less, the VsN within the SCD was significantly lower in the multilayer structure type [80 (68-89)] than in the single-layer structure type [99 (92-100)] (p < 0.0001).</p><p><strong>Conclusion: </strong>The results show that, in addition to SCD thickness, its internal structure reduces the VsN and contributes to Vs measurement instability.</p>\",\"PeriodicalId\":50130,\"journal\":{\"name\":\"Journal of Medical Ultrasonics\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":2.1000,\"publicationDate\":\"2025-10-04\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Medical Ultrasonics\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1007/s10396-025-01582-9\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Ultrasonics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s10396-025-01582-9","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

目的:在肥胖或皮肤包膜距离(SCD)≥20mm的个体中,横波速度(Vs)测量可能不稳定。然而,在SCD没有明显增加的情况下,也观察到Vs的不稳定性;因此,本研究旨在探讨导致VsN(净有效横波速度)降低的因素,VsN是反映单个v测量可靠性的指标。方法:本研究共纳入317名接受社区健康筛查的个体。我们研究了VsN与物理表现、实验室数据、SCD和SCD结构(单层或多层)之间的关系,以确定与VsN降低相关的因素。结果:多元回归分析显示体脂率(β = - 0.135, p = 0.011)、SCD (β = - 0.347, p)与VsN (β = - 0.347, p)之间存在显著差异。结论:除SCD厚度外,其内部结构降低了VsN,导致了VsN测量的不稳定性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Reliability of liver stiffness measurements using shear wave elastography: impact of skin-to-capsule distance and subcutaneous tissue structure.

Purpose: Shear wave velocity (Vs) measurements can be unstable in individuals with obesity or with a skin-capsule distance (SCD) of ≥ 20 mm. However, instability of Vs has also been observed in cases where SCD is not markedly increased; therefore, the present study aimed to investigate the factors contributing to reduced VsN (net amount of effective shear wave velocity), an index that reflects the reliability of individual Vs measurements.

Methods: In total, 317 individuals who underwent community-based health screening were included in this study. We investigated the relationship between VsN and physical findings, laboratory data, SCD, and structure of the SCD (single vs. multilayered) to identify factors associated with decreased VsN.

Results: Multiple regression analysis revealed that body fat percentage (β = - 0.135, p = 0.011), SCD (β = - 0.347, p < 0.0001), and the structure of the SC region (β = - 0.295, p < 0.0001) were independently associated with decreased VsN. In cases where the SCD exceeded 20 mm, the VsN within the SCD was significantly lower in the multilayer structural type [52 (28-75)] than in the single-layer structural type [95 (77-99)] (p = 0.004). Furthermore, even when the SCD was 20 mm or less, the VsN within the SCD was significantly lower in the multilayer structure type [80 (68-89)] than in the single-layer structure type [99 (92-100)] (p < 0.0001).

Conclusion: The results show that, in addition to SCD thickness, its internal structure reduces the VsN and contributes to Vs measurement instability.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
3.30
自引率
11.10%
发文量
102
审稿时长
>12 weeks
期刊介绍: The Journal of Medical Ultrasonics is the official journal of the Japan Society of Ultrasonics in Medicine. The main purpose of the journal is to provide forum for the publication of papers documenting recent advances and new developments in the entire field of ultrasound in medicine and biology, encompassing both the medical and the engineering aspects of the science.The journal welcomes original articles, review articles, images, and letters to the editor.The journal also provides state-of-the-art information such as announcements from the boards and the committees of the society.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信