{"title":"B-014在Alinity ci系统上雅培维生素测定的分析性能评价","authors":"Marvin Berman","doi":"10.1093/clinchem/hvaf086.412","DOIUrl":null,"url":null,"abstract":"Background Vitamins can be crucial for diagnosing and managing health conditions related to low or excessive intake and for monitoring the effectiveness of dietary changes or supplement strategies. Abbott provides an extensive menu of vitamin assays which aid in the diagnosis and management of many health conditions such as deficiency or excess. Assays such as Vitamin D, Vitamin B12 and Vitamin C are used to monitor bone health, neurological function and immune system function respectively. Thus, vitamins are valuable for disease risk assessment. Homocysteine is an amino acid that plays a crucial role in cell metabolism and its levels are closely linked to vitamins B12 and folate (vit B9) deficiencies. Recent guideline changes to Vitamin D (Endo Society 2024) and B12 (Nice 2024) support the need for accurate assay performance. The Alinity ci system is part of a unified family of systems that are engineered for flexibility, efficiency, and decreased system down time. The design is based on insights from customers, resulting in a number of benefits including a smaller footprint, improved workflow, and great throughput with up to 200 tests per hour. The objective of this study is to demonstrate the analytical performance of representative assays from the Vitamin Panel of the Alinity i system, which consists of assays that utilize photometric technology for the quantitative determination of analytes in human serum or plasma. Methods Key performance testing including precision, limit of quantitation (LoQ), endogenous interferents, linearity and method comparison were assessed per Clinical and Laboratory Standards Institute (CLSI) protocols. Precision was assessed by testing controls and panels twice per day for 20 days on 2 instruments. Sensitivity was determined with 3 reagent lots on 2 instruments. Endogenous compounds were evaluated at a minimum of one analyte concentration. Linearity was assessed using a minimum of 6 panels that spanned the measuring interval. Method comparison was performed by measuring a minimum of 100 specimens on Alinity and the comparator method. The assay measuring interval was defined by the range for which acceptable performance for bias, imprecision, and linearity was met. Results The observed results for precision were 4.1 to 9%. Alinity method comparison vs ARCHITECT demonstrated slopes of 0.95 to 1.04 with correlation coefficients of 0.96 to 1.00. All assays demonstrated equivalent sensitivities, linearity, interferents and measuring intervals to ARCHITECT. Conclusion Standardization of the Abbott assays ensures the assays meet current guidelines. The vitamin panel of assays on the Abbott Alinity i platform demonstrated acceptable performance for precision, sensitivity, interferents and linearity. Method comparison data showed equivalency with the on-market assays, which verified the performance of these products on the Alinity i system.","PeriodicalId":10690,"journal":{"name":"Clinical chemistry","volume":"19 1","pages":""},"PeriodicalIF":6.3000,"publicationDate":"2025-10-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"B-014 Analytical performance evaluation of the Abbott Vitamin assays on the Alinity ci system\",\"authors\":\"Marvin Berman\",\"doi\":\"10.1093/clinchem/hvaf086.412\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Background Vitamins can be crucial for diagnosing and managing health conditions related to low or excessive intake and for monitoring the effectiveness of dietary changes or supplement strategies. Abbott provides an extensive menu of vitamin assays which aid in the diagnosis and management of many health conditions such as deficiency or excess. Assays such as Vitamin D, Vitamin B12 and Vitamin C are used to monitor bone health, neurological function and immune system function respectively. Thus, vitamins are valuable for disease risk assessment. Homocysteine is an amino acid that plays a crucial role in cell metabolism and its levels are closely linked to vitamins B12 and folate (vit B9) deficiencies. Recent guideline changes to Vitamin D (Endo Society 2024) and B12 (Nice 2024) support the need for accurate assay performance. The Alinity ci system is part of a unified family of systems that are engineered for flexibility, efficiency, and decreased system down time. The design is based on insights from customers, resulting in a number of benefits including a smaller footprint, improved workflow, and great throughput with up to 200 tests per hour. The objective of this study is to demonstrate the analytical performance of representative assays from the Vitamin Panel of the Alinity i system, which consists of assays that utilize photometric technology for the quantitative determination of analytes in human serum or plasma. Methods Key performance testing including precision, limit of quantitation (LoQ), endogenous interferents, linearity and method comparison were assessed per Clinical and Laboratory Standards Institute (CLSI) protocols. Precision was assessed by testing controls and panels twice per day for 20 days on 2 instruments. Sensitivity was determined with 3 reagent lots on 2 instruments. Endogenous compounds were evaluated at a minimum of one analyte concentration. Linearity was assessed using a minimum of 6 panels that spanned the measuring interval. Method comparison was performed by measuring a minimum of 100 specimens on Alinity and the comparator method. The assay measuring interval was defined by the range for which acceptable performance for bias, imprecision, and linearity was met. Results The observed results for precision were 4.1 to 9%. Alinity method comparison vs ARCHITECT demonstrated slopes of 0.95 to 1.04 with correlation coefficients of 0.96 to 1.00. All assays demonstrated equivalent sensitivities, linearity, interferents and measuring intervals to ARCHITECT. Conclusion Standardization of the Abbott assays ensures the assays meet current guidelines. The vitamin panel of assays on the Abbott Alinity i platform demonstrated acceptable performance for precision, sensitivity, interferents and linearity. Method comparison data showed equivalency with the on-market assays, which verified the performance of these products on the Alinity i system.\",\"PeriodicalId\":10690,\"journal\":{\"name\":\"Clinical chemistry\",\"volume\":\"19 1\",\"pages\":\"\"},\"PeriodicalIF\":6.3000,\"publicationDate\":\"2025-10-02\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Clinical chemistry\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1093/clinchem/hvaf086.412\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical chemistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/clinchem/hvaf086.412","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

维生素对于诊断和管理与低或过量摄入有关的健康状况以及监测饮食变化或补充策略的有效性至关重要。雅培提供了广泛的维生素检测菜单,有助于诊断和管理许多健康状况,如缺乏或过量。测定维生素D、维生素B12和维生素C分别用于监测骨骼健康、神经功能和免疫系统功能。因此,维生素对疾病风险评估很有价值。同型半胱氨酸是一种氨基酸,在细胞代谢中起着至关重要的作用,它的水平与维生素B12和叶酸(维生素B9)缺乏密切相关。最近对维生素D (Endo Society 2024)和B12 (Nice 2024)指南的修改支持对准确分析性能的需求。Alinity ci系统是为灵活性、效率和减少系统停机时间而设计的统一系统系列的一部分。该设计基于客户的见解,带来了许多好处,包括更小的占用空间、改进的工作流程和每小时多达200个测试的高吞吐量。本研究的目的是证明Alinity i系统中维生素面板的代表性分析方法的分析性能,该系统包括利用光度法技术定量测定人血清或血浆中的分析物的分析方法。方法根据临床与实验室标准协会(CLSI)的标准,对精密度、定量限(LoQ)、内源干扰、线性度和方法比较等关键性能指标进行评价。通过在2台仪器上每天两次测试控制和面板来评估精度,持续20天。在2台仪器上用3批试剂测定灵敏度。内源性化合物在至少一个分析物浓度下进行评估。使用跨越测量间隔的至少6个面板来评估线性度。方法比较是通过在Alinity和comparator方法上测量至少100个标本来进行的。测定间隔由可接受的偏差、不精确和线性的范围来定义。结果观察结果精密度为4.1 ~ 9%。Alinity方法与ARCHITECT方法的比较显示斜率为0.95至1.04,相关系数为0.96至1.00。所有的测定都显示出与ARCHITECT相同的灵敏度、线性度、干扰和测量间隔。结论雅培测定法的标准化确保了测定法符合现行指南。雅培Alinity i平台上的维生素检测面板在精度、灵敏度、干扰和线性方面表现出可接受的性能。方法比较数据显示与市场上的分析结果相当,验证了这些产品在Alinity i系统上的性能。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
B-014 Analytical performance evaluation of the Abbott Vitamin assays on the Alinity ci system
Background Vitamins can be crucial for diagnosing and managing health conditions related to low or excessive intake and for monitoring the effectiveness of dietary changes or supplement strategies. Abbott provides an extensive menu of vitamin assays which aid in the diagnosis and management of many health conditions such as deficiency or excess. Assays such as Vitamin D, Vitamin B12 and Vitamin C are used to monitor bone health, neurological function and immune system function respectively. Thus, vitamins are valuable for disease risk assessment. Homocysteine is an amino acid that plays a crucial role in cell metabolism and its levels are closely linked to vitamins B12 and folate (vit B9) deficiencies. Recent guideline changes to Vitamin D (Endo Society 2024) and B12 (Nice 2024) support the need for accurate assay performance. The Alinity ci system is part of a unified family of systems that are engineered for flexibility, efficiency, and decreased system down time. The design is based on insights from customers, resulting in a number of benefits including a smaller footprint, improved workflow, and great throughput with up to 200 tests per hour. The objective of this study is to demonstrate the analytical performance of representative assays from the Vitamin Panel of the Alinity i system, which consists of assays that utilize photometric technology for the quantitative determination of analytes in human serum or plasma. Methods Key performance testing including precision, limit of quantitation (LoQ), endogenous interferents, linearity and method comparison were assessed per Clinical and Laboratory Standards Institute (CLSI) protocols. Precision was assessed by testing controls and panels twice per day for 20 days on 2 instruments. Sensitivity was determined with 3 reagent lots on 2 instruments. Endogenous compounds were evaluated at a minimum of one analyte concentration. Linearity was assessed using a minimum of 6 panels that spanned the measuring interval. Method comparison was performed by measuring a minimum of 100 specimens on Alinity and the comparator method. The assay measuring interval was defined by the range for which acceptable performance for bias, imprecision, and linearity was met. Results The observed results for precision were 4.1 to 9%. Alinity method comparison vs ARCHITECT demonstrated slopes of 0.95 to 1.04 with correlation coefficients of 0.96 to 1.00. All assays demonstrated equivalent sensitivities, linearity, interferents and measuring intervals to ARCHITECT. Conclusion Standardization of the Abbott assays ensures the assays meet current guidelines. The vitamin panel of assays on the Abbott Alinity i platform demonstrated acceptable performance for precision, sensitivity, interferents and linearity. Method comparison data showed equivalency with the on-market assays, which verified the performance of these products on the Alinity i system.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Clinical chemistry
Clinical chemistry 医学-医学实验技术
CiteScore
11.30
自引率
4.30%
发文量
212
审稿时长
1.7 months
期刊介绍: Clinical Chemistry is a peer-reviewed scientific journal that is the premier publication for the science and practice of clinical laboratory medicine. It was established in 1955 and is associated with the Association for Diagnostics & Laboratory Medicine (ADLM). The journal focuses on laboratory diagnosis and management of patients, and has expanded to include other clinical laboratory disciplines such as genomics, hematology, microbiology, and toxicology. It also publishes articles relevant to clinical specialties including cardiology, endocrinology, gastroenterology, genetics, immunology, infectious diseases, maternal-fetal medicine, neurology, nutrition, oncology, and pediatrics. In addition to original research, editorials, and reviews, Clinical Chemistry features recurring sections such as clinical case studies, perspectives, podcasts, and Q&A articles. It has the highest impact factor among journals of clinical chemistry, laboratory medicine, pathology, analytical chemistry, transfusion medicine, and clinical microbiology. The journal is indexed in databases such as MEDLINE and Web of Science.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信