优化的3D UTE和ZTE MRI用于高分辨率肺部成像的比较研究。

IF 3 3区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Michael Carl, Jiaji Wang, James Lo, Jonathan H Chung, Jiang Du, Yajun Ma
{"title":"优化的3D UTE和ZTE MRI用于高分辨率肺部成像的比较研究。","authors":"Michael Carl, Jiaji Wang, James Lo, Jonathan H Chung, Jiang Du, Yajun Ma","doi":"10.1002/mrm.70100","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To optimize and compare the performance of 3D UTE and zero echo time (ZTE) MRI sequences for high-resolution lung imaging at 3 T.</p><p><strong>Methods: </strong>UTE and ZTE sequences were optimized for lung imaging through phantom studies, and in vivo experiments. Imaging parameters, including bandwidth (BW) and flip angle (FA), were systematically assessed to identify the optimal settings for both sequences. A novel transient UTE sequence was introduced to enhance the SNR by allowing magnetization recovery during the respiratory wait period. Numerical simulations and in vivo experiments were conducted to compare the SNR performance of the proposed transient UTE with conventional steady-state UTE.</p><p><strong>Results: </strong>UTE imaging with a BW of ±125 kHz and an FA of 2° produced the sharpest lung images without noticeable artifacts. In contrast, ZTE imaging was optimal at a lower BW of ±62.5 kHz and an FA of 2°, but showed greater blurriness and more pronounced inhomogeneous excitation artifacts than UTE. The transient UTE sequence increased lung imaging SNR by approximately 11% to 20% compared to steady-state UTE without introducing obvious artifacts.</p><p><strong>Conclusion: </strong>UTE outperformed ZTE in terms of image sharpness and artifact reduction for lung imaging. The proposed transient UTE sequence further enhanced SNR, demonstrating the potential for improved lung MRI quality and diagnostic accuracy.</p>","PeriodicalId":18065,"journal":{"name":"Magnetic Resonance in Medicine","volume":" ","pages":""},"PeriodicalIF":3.0000,"publicationDate":"2025-09-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Optimized 3D UTE and ZTE MRI for high-resolution lung imaging: A comparative study.\",\"authors\":\"Michael Carl, Jiaji Wang, James Lo, Jonathan H Chung, Jiang Du, Yajun Ma\",\"doi\":\"10.1002/mrm.70100\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>To optimize and compare the performance of 3D UTE and zero echo time (ZTE) MRI sequences for high-resolution lung imaging at 3 T.</p><p><strong>Methods: </strong>UTE and ZTE sequences were optimized for lung imaging through phantom studies, and in vivo experiments. Imaging parameters, including bandwidth (BW) and flip angle (FA), were systematically assessed to identify the optimal settings for both sequences. A novel transient UTE sequence was introduced to enhance the SNR by allowing magnetization recovery during the respiratory wait period. Numerical simulations and in vivo experiments were conducted to compare the SNR performance of the proposed transient UTE with conventional steady-state UTE.</p><p><strong>Results: </strong>UTE imaging with a BW of ±125 kHz and an FA of 2° produced the sharpest lung images without noticeable artifacts. In contrast, ZTE imaging was optimal at a lower BW of ±62.5 kHz and an FA of 2°, but showed greater blurriness and more pronounced inhomogeneous excitation artifacts than UTE. The transient UTE sequence increased lung imaging SNR by approximately 11% to 20% compared to steady-state UTE without introducing obvious artifacts.</p><p><strong>Conclusion: </strong>UTE outperformed ZTE in terms of image sharpness and artifact reduction for lung imaging. The proposed transient UTE sequence further enhanced SNR, demonstrating the potential for improved lung MRI quality and diagnostic accuracy.</p>\",\"PeriodicalId\":18065,\"journal\":{\"name\":\"Magnetic Resonance in Medicine\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":3.0000,\"publicationDate\":\"2025-09-29\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Magnetic Resonance in Medicine\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1002/mrm.70100\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Magnetic Resonance in Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/mrm.70100","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

目的:优化和比较3D UTE和零回波时间(ZTE) MRI序列在3t高分辨率肺部成像中的性能。方法:通过假体研究和体内实验,优化UTE和ZTE序列用于肺部成像。系统评估成像参数,包括带宽(BW)和翻转角度(FA),以确定两个序列的最佳设置。引入了一种新的瞬态UTE序列,通过在呼吸等待期间允许磁化恢复来提高信噪比。通过数值模拟和体内实验比较了所提出的瞬态UTE与常规稳态UTE的信噪比性能。结果:超声胸宽为±125 kHz, FA为2°时,肺图像最清晰,无明显伪影。相比之下,中兴通讯成像在较低的BW为±62.5 kHz和FA为2°时是最佳的,但与UTE相比,中兴通讯显示出更大的模糊和更明显的不均匀激发伪影。与稳态UTE相比,瞬态UTE序列将肺部成像信噪比提高了约11%至20%,而不会引入明显的伪影。结论:UTE在肺部成像的图像清晰度和伪影减少方面优于中兴通讯。所提出的瞬时UTE序列进一步提高了信噪比,显示了改善肺部MRI质量和诊断准确性的潜力。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Optimized 3D UTE and ZTE MRI for high-resolution lung imaging: A comparative study.

Purpose: To optimize and compare the performance of 3D UTE and zero echo time (ZTE) MRI sequences for high-resolution lung imaging at 3 T.

Methods: UTE and ZTE sequences were optimized for lung imaging through phantom studies, and in vivo experiments. Imaging parameters, including bandwidth (BW) and flip angle (FA), were systematically assessed to identify the optimal settings for both sequences. A novel transient UTE sequence was introduced to enhance the SNR by allowing magnetization recovery during the respiratory wait period. Numerical simulations and in vivo experiments were conducted to compare the SNR performance of the proposed transient UTE with conventional steady-state UTE.

Results: UTE imaging with a BW of ±125 kHz and an FA of 2° produced the sharpest lung images without noticeable artifacts. In contrast, ZTE imaging was optimal at a lower BW of ±62.5 kHz and an FA of 2°, but showed greater blurriness and more pronounced inhomogeneous excitation artifacts than UTE. The transient UTE sequence increased lung imaging SNR by approximately 11% to 20% compared to steady-state UTE without introducing obvious artifacts.

Conclusion: UTE outperformed ZTE in terms of image sharpness and artifact reduction for lung imaging. The proposed transient UTE sequence further enhanced SNR, demonstrating the potential for improved lung MRI quality and diagnostic accuracy.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
6.70
自引率
24.20%
发文量
376
审稿时长
2-4 weeks
期刊介绍: Magnetic Resonance in Medicine (Magn Reson Med) is an international journal devoted to the publication of original investigations concerned with all aspects of the development and use of nuclear magnetic resonance and electron paramagnetic resonance techniques for medical applications. Reports of original investigations in the areas of mathematics, computing, engineering, physics, biophysics, chemistry, biochemistry, and physiology directly relevant to magnetic resonance will be accepted, as well as methodology-oriented clinical studies.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信