{"title":"髋臼后壁骨折固定的力学评价:患者特异性植入物与常规植入物。","authors":"Miriam G E Oldhoff, Anne M L Meesters, Joep Kraeima, Mehool R Acharya, Kaj Ten Duis, Jean-Paul P M de Vries, Bram B J Merema, Frank F A IJpma","doi":"10.1002/jor.70062","DOIUrl":null,"url":null,"abstract":"<p><p>The aim of the study was to design a standardized mechanical test setup and a corresponding finite element analysis to assess the stability and strength of both patient-specific and conventional implants for posterior wall acetabular fractures. Ten synthetic hemi-pelves with posterior wall fractures were biomechanically tested with two types of implants: a patient-specific implant (PSI) and a seven-hole plate conventional implant. 3D-printed guides ensured reproducibility. The models were tested using an Instron machine. The protocol involved 10,000 cyclic load cycles with static tests at 3200 N before and after to simulate early postoperative weightbearing conditions. Construct stiffness, stiffness over cyclic loading and fracture gapping were measured and compared. A finite element analysis was created with similar conditions to investigate stresses within the synthetic bone and fixation materials. The mechanical tests showed comparable stiffness for PSI (1.75 kN/mm) and the conventional implant (1.71 kN/mm, p = 0.47). Stability over 10,000 cycles was similar, and fracture gapping remained minimal (0.0-0.8 mm) without significant differences. No failure or plastic deformation occurred under 3200 N loading. Finite element analysis confirmed that von Mises stresses remained below the yield stress. This study introduces a reproducible workflow for biomechanical testing of acetabular fractures using synthetic bone models and 3D-printed guides. It serves as a step-by-step guideline and standard reference for pelvic biomechanical testing. Both patient-specific and conventional implants, using a seven-hole plate construct with one or two screws through the plate in the fracture fragment, provide stable fixation for large posterior wall fragments.</p>","PeriodicalId":16650,"journal":{"name":"Journal of Orthopaedic Research®","volume":" ","pages":""},"PeriodicalIF":2.3000,"publicationDate":"2025-09-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Mechanical Evaluation of Posterior Wall Acetabular Facture Fixation: Patient-Specific Implants Versus Conventional Implants.\",\"authors\":\"Miriam G E Oldhoff, Anne M L Meesters, Joep Kraeima, Mehool R Acharya, Kaj Ten Duis, Jean-Paul P M de Vries, Bram B J Merema, Frank F A IJpma\",\"doi\":\"10.1002/jor.70062\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The aim of the study was to design a standardized mechanical test setup and a corresponding finite element analysis to assess the stability and strength of both patient-specific and conventional implants for posterior wall acetabular fractures. Ten synthetic hemi-pelves with posterior wall fractures were biomechanically tested with two types of implants: a patient-specific implant (PSI) and a seven-hole plate conventional implant. 3D-printed guides ensured reproducibility. The models were tested using an Instron machine. The protocol involved 10,000 cyclic load cycles with static tests at 3200 N before and after to simulate early postoperative weightbearing conditions. Construct stiffness, stiffness over cyclic loading and fracture gapping were measured and compared. A finite element analysis was created with similar conditions to investigate stresses within the synthetic bone and fixation materials. The mechanical tests showed comparable stiffness for PSI (1.75 kN/mm) and the conventional implant (1.71 kN/mm, p = 0.47). Stability over 10,000 cycles was similar, and fracture gapping remained minimal (0.0-0.8 mm) without significant differences. No failure or plastic deformation occurred under 3200 N loading. Finite element analysis confirmed that von Mises stresses remained below the yield stress. This study introduces a reproducible workflow for biomechanical testing of acetabular fractures using synthetic bone models and 3D-printed guides. It serves as a step-by-step guideline and standard reference for pelvic biomechanical testing. Both patient-specific and conventional implants, using a seven-hole plate construct with one or two screws through the plate in the fracture fragment, provide stable fixation for large posterior wall fragments.</p>\",\"PeriodicalId\":16650,\"journal\":{\"name\":\"Journal of Orthopaedic Research®\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":2.3000,\"publicationDate\":\"2025-09-23\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Orthopaedic Research®\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1002/jor.70062\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"ORTHOPEDICS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Orthopaedic Research®","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/jor.70062","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

摘要

本研究的目的是设计一个标准化的力学测试装置和相应的有限元分析,以评估髋臼后壁骨折患者特异性植入物和常规植入物的稳定性和强度。对10例后壁骨折的合成半骨盆进行生物力学测试,采用两种类型的植入物:患者专用植入物(PSI)和七孔钢板常规植入物。3d打印导轨确保了再现性。使用Instron机器对模型进行了测试。该方案包括在3200 N的静态测试前后进行10,000次循环载荷循环,以模拟术后早期的负重条件。对结构刚度、循环加载刚度和断裂间隙进行了测量和比较。在类似的条件下创建了有限元分析来研究合成骨和固定材料内的应力。力学测试显示PSI (1.75 kN/mm)和常规种植体(1.71 kN/mm, p = 0.47)的刚度相当。超过10,000次循环的稳定性相似,裂缝间隙保持最小(0.0-0.8 mm),没有显着差异。在3200n载荷下,未发生破坏和塑性变形。有限元分析证实,von Mises应力保持在屈服应力以下。本研究介绍了一种可重复的工作流程,用于使用合成骨模型和3d打印导板进行髋臼骨折的生物力学测试。它作为骨盆生物力学测试的一步一步的指导和标准参考。患者特异性植入物和常规植入物均采用7孔钢板结构,在骨折碎片中通过钢板置入1或2颗螺钉,可为较大的后壁碎片提供稳定的固定。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Mechanical Evaluation of Posterior Wall Acetabular Facture Fixation: Patient-Specific Implants Versus Conventional Implants.

The aim of the study was to design a standardized mechanical test setup and a corresponding finite element analysis to assess the stability and strength of both patient-specific and conventional implants for posterior wall acetabular fractures. Ten synthetic hemi-pelves with posterior wall fractures were biomechanically tested with two types of implants: a patient-specific implant (PSI) and a seven-hole plate conventional implant. 3D-printed guides ensured reproducibility. The models were tested using an Instron machine. The protocol involved 10,000 cyclic load cycles with static tests at 3200 N before and after to simulate early postoperative weightbearing conditions. Construct stiffness, stiffness over cyclic loading and fracture gapping were measured and compared. A finite element analysis was created with similar conditions to investigate stresses within the synthetic bone and fixation materials. The mechanical tests showed comparable stiffness for PSI (1.75 kN/mm) and the conventional implant (1.71 kN/mm, p = 0.47). Stability over 10,000 cycles was similar, and fracture gapping remained minimal (0.0-0.8 mm) without significant differences. No failure or plastic deformation occurred under 3200 N loading. Finite element analysis confirmed that von Mises stresses remained below the yield stress. This study introduces a reproducible workflow for biomechanical testing of acetabular fractures using synthetic bone models and 3D-printed guides. It serves as a step-by-step guideline and standard reference for pelvic biomechanical testing. Both patient-specific and conventional implants, using a seven-hole plate construct with one or two screws through the plate in the fracture fragment, provide stable fixation for large posterior wall fragments.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Orthopaedic Research®
Journal of Orthopaedic Research® 医学-整形外科
CiteScore
6.10
自引率
3.60%
发文量
261
审稿时长
3-6 weeks
期刊介绍: The Journal of Orthopaedic Research is the forum for the rapid publication of high quality reports of new information on the full spectrum of orthopaedic research, including life sciences, engineering, translational, and clinical studies.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信