{"title":"SYSMEX CN-6000与STAGO STA-R对溶血和血脂的影响比较。","authors":"Seren Orhan, Elif İşbilen","doi":"10.5937/jomb0-56377","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Routine coagulation testing is critical in diagnosing hemostatic disorders and monitoring anticoagulant therapy. The SYSMEX CN-6000 and STAGO STA-R analyzers utilise different clot detection methods, which may impact test results. This study evaluates the analytical performance of these two automated coagulation analysers and examines the effects of hemolysis and lipemia on routine coagulation tests.</p><p><strong>Methods: </strong>Blood samples were collected from patients at Gaziantep University Şahinbey Research and Application Hospital and analysed for activated partial thromboplastin time (APTT), prothrombin time (PT), fibrinogen (FBG), and D-dimer using both analysers. Precision, method comparison, and interference studies were conducted following CLSI guidelines. Hemolysis and lipemia were induced in vitro, and their effects on test results were evaluated based on Fraser's criteria.</p><p><strong>Results: </strong>All precision study CV values were within the acceptable limits of biological variation. APTT results exhibited a significant systematic difference between analysers (r= 0.872), whereas PT (INR), FBG, and D-dimer showed strong correlations (r> 0.945). Hemolysis had a minimal impact at lower concentrations (<1 g/L). However, at 4 g/L, PT bias increased to 2.8% for the CN-6000 and 2.0% for the STA-R, with similar increases observed in APTT and FBG. Lipemia significantly affected CN-6000, which failed to produce PT results at triglyceride levels 14 mmol/L and APTT/FBG results at 28 mmol/L. In contrast, STA-R provided results with biases below 7.3% at all lipemia levels.</p><p><strong>Conclusions: </strong>Both analysers demonstrated strong analytical performance, though methodological differences influenced APTT measurements. Hemolysis had a minor impact within Fraser's acceptable bias limits, whereas CN-6000 exhibited performance limitations in severely lipemic samples, necessitating preanalytical lipid-reducing strategies. These findings underscore the need for expanded reference range studies and optimised laboratory workflows to enhance coagulation testing reliability.</p>","PeriodicalId":16175,"journal":{"name":"Journal of Medical Biochemistry","volume":"44 5","pages":"976-984"},"PeriodicalIF":1.5000,"publicationDate":"2025-08-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12433291/pdf/","citationCount":"0","resultStr":"{\"title\":\"Comparison of SYSMEX CN-6000 and STAGO STA-R: Effects of hemolysis and lipemia.\",\"authors\":\"Seren Orhan, Elif İşbilen\",\"doi\":\"10.5937/jomb0-56377\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Routine coagulation testing is critical in diagnosing hemostatic disorders and monitoring anticoagulant therapy. The SYSMEX CN-6000 and STAGO STA-R analyzers utilise different clot detection methods, which may impact test results. This study evaluates the analytical performance of these two automated coagulation analysers and examines the effects of hemolysis and lipemia on routine coagulation tests.</p><p><strong>Methods: </strong>Blood samples were collected from patients at Gaziantep University Şahinbey Research and Application Hospital and analysed for activated partial thromboplastin time (APTT), prothrombin time (PT), fibrinogen (FBG), and D-dimer using both analysers. Precision, method comparison, and interference studies were conducted following CLSI guidelines. Hemolysis and lipemia were induced in vitro, and their effects on test results were evaluated based on Fraser's criteria.</p><p><strong>Results: </strong>All precision study CV values were within the acceptable limits of biological variation. APTT results exhibited a significant systematic difference between analysers (r= 0.872), whereas PT (INR), FBG, and D-dimer showed strong correlations (r> 0.945). Hemolysis had a minimal impact at lower concentrations (<1 g/L). However, at 4 g/L, PT bias increased to 2.8% for the CN-6000 and 2.0% for the STA-R, with similar increases observed in APTT and FBG. Lipemia significantly affected CN-6000, which failed to produce PT results at triglyceride levels 14 mmol/L and APTT/FBG results at 28 mmol/L. In contrast, STA-R provided results with biases below 7.3% at all lipemia levels.</p><p><strong>Conclusions: </strong>Both analysers demonstrated strong analytical performance, though methodological differences influenced APTT measurements. Hemolysis had a minor impact within Fraser's acceptable bias limits, whereas CN-6000 exhibited performance limitations in severely lipemic samples, necessitating preanalytical lipid-reducing strategies. These findings underscore the need for expanded reference range studies and optimised laboratory workflows to enhance coagulation testing reliability.</p>\",\"PeriodicalId\":16175,\"journal\":{\"name\":\"Journal of Medical Biochemistry\",\"volume\":\"44 5\",\"pages\":\"976-984\"},\"PeriodicalIF\":1.5000,\"publicationDate\":\"2025-08-21\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12433291/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Medical Biochemistry\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.5937/jomb0-56377\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"BIOCHEMISTRY & MOLECULAR BIOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Biochemistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.5937/jomb0-56377","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"BIOCHEMISTRY & MOLECULAR BIOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

背景:常规凝血检查对诊断止血障碍和监测抗凝治疗至关重要。SYSMEX CN-6000和STAGO STA-R分析仪采用不同的血块检测方法,这可能会影响测试结果。本研究评估了这两种自动凝血分析仪的分析性能,并检查了溶血和血脂对常规凝血试验的影响。方法:收集加济安泰普大学Şahinbey研究与应用医院患者的血液样本,使用两种分析仪分析活化部分凝血活素时间(APTT)、凝血酶原时间(PT)、纤维蛋白原(FBG)和d -二聚体。精确度、方法比较和干扰研究遵循CLSI指南进行。体外诱导溶血和脂血症,并根据弗雷泽标准评估其对试验结果的影响。结果:所有精密度研究CV值均在生物变异可接受范围内。APTT结果在各分析仪之间显示出显著的系统差异(r= 0.872),而PT (INR)、FBG和d -二聚体显示出很强的相关性(r> 0.945)。溶血在较低浓度下影响最小(结论:两种分析仪都表现出很强的分析性能,尽管方法差异影响了APTT测量。溶血在弗雷泽的可接受偏差范围内影响较小,而CN-6000在严重血脂样品中表现出性能限制,需要分析前降脂策略。这些发现强调了扩大参考范围研究和优化实验室工作流程以提高凝血检测可靠性的必要性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

Comparison of SYSMEX CN-6000 and STAGO STA-R: Effects of hemolysis and lipemia.

Comparison of SYSMEX CN-6000 and STAGO STA-R: Effects of hemolysis and lipemia.

Comparison of SYSMEX CN-6000 and STAGO STA-R: Effects of hemolysis and lipemia.

Background: Routine coagulation testing is critical in diagnosing hemostatic disorders and monitoring anticoagulant therapy. The SYSMEX CN-6000 and STAGO STA-R analyzers utilise different clot detection methods, which may impact test results. This study evaluates the analytical performance of these two automated coagulation analysers and examines the effects of hemolysis and lipemia on routine coagulation tests.

Methods: Blood samples were collected from patients at Gaziantep University Şahinbey Research and Application Hospital and analysed for activated partial thromboplastin time (APTT), prothrombin time (PT), fibrinogen (FBG), and D-dimer using both analysers. Precision, method comparison, and interference studies were conducted following CLSI guidelines. Hemolysis and lipemia were induced in vitro, and their effects on test results were evaluated based on Fraser's criteria.

Results: All precision study CV values were within the acceptable limits of biological variation. APTT results exhibited a significant systematic difference between analysers (r= 0.872), whereas PT (INR), FBG, and D-dimer showed strong correlations (r> 0.945). Hemolysis had a minimal impact at lower concentrations (<1 g/L). However, at 4 g/L, PT bias increased to 2.8% for the CN-6000 and 2.0% for the STA-R, with similar increases observed in APTT and FBG. Lipemia significantly affected CN-6000, which failed to produce PT results at triglyceride levels 14 mmol/L and APTT/FBG results at 28 mmol/L. In contrast, STA-R provided results with biases below 7.3% at all lipemia levels.

Conclusions: Both analysers demonstrated strong analytical performance, though methodological differences influenced APTT measurements. Hemolysis had a minor impact within Fraser's acceptable bias limits, whereas CN-6000 exhibited performance limitations in severely lipemic samples, necessitating preanalytical lipid-reducing strategies. These findings underscore the need for expanded reference range studies and optimised laboratory workflows to enhance coagulation testing reliability.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Medical Biochemistry
Journal of Medical Biochemistry BIOCHEMISTRY & MOLECULAR BIOLOGY-
CiteScore
3.00
自引率
12.00%
发文量
60
审稿时长
>12 weeks
期刊介绍: The JOURNAL OF MEDICAL BIOCHEMISTRY (J MED BIOCHEM) is the official journal of the Society of Medical Biochemists of Serbia with international peer-review. Papers are independently reviewed by at least two reviewers selected by the Editors as Blind Peer Reviews. The Journal of Medical Biochemistry is published quarterly. The Journal publishes original scientific and specialized articles on all aspects of clinical and medical biochemistry, molecular medicine, clinical hematology and coagulation, clinical immunology and autoimmunity, clinical microbiology, virology, clinical genomics and molecular biology, genetic epidemiology, drug measurement, evaluation of diagnostic markers, new reagents and laboratory equipment, reference materials and methods, reference values, laboratory organization, automation, quality control, clinical metrology, all related scientific disciplines where chemistry, biochemistry, molecular biology and immunochemistry deal with the study of normal and pathologic processes in human beings.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信