{"title":"MOPTIM MRT-200屈光仪与2WIN和Retinomax的对比验证。","authors":"Heidi Schumacher, Giancarlo Garaygay, Aana Ayagalria, Bryant Ikaika Lauano, Denise M Marshall, J Richard Townley, Robert W Arnold","doi":"10.2147/OPTH.S545397","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Portable sphero-cylinder refraction is useful for remote medical missions and self-checking. A novel, inexpensive, handheld optical scope with an internal vision chart and adjustable diopter lens, called the Moptim MRT-200, was therefore validated.</p><p><strong>Methods: </strong>Young, capable patients from a pediatric eye practice had dry refraction estimates with the Moptim MRT-200, Adaptica 2WIN photoscreener, and Monocular Retinomax compared with actual refined retinoscopy. Vector-converted sphero-cylinder refractions and the univariate ABCD Ellipsoid were used for comparison as well as monocular visual acuity.</p><p><strong>Results: </strong>206 racially diverse, high-risk refractive patients with median \"Ellipsoid-Gap\" of 3.5 had high correlation spherical equivalent and similar ABCD Ellipsoid comparing Moptim MRT-200, 2WIN and Retinomax (Kruskal Wallis X<sup>2</sup>(5) = 4.2, p = 0.52) however the 61 patients with blur associated with Ellipsoid gap over 5.0 had better refractive accuracy with 2WIN and Retinomax than Moptim MRT-200 (p < 0.001). Visual acuity with Moptim MRT-200 had a weak correlation (ICC.41 AND.47) and a Bland-Altman mean difference of approximately 1 logMAR line less fine than the actual phoropter refraction.</p><p><strong>Conclusion: </strong>Precise sphero-cylinder estimation with the Moptim MRT-200 was too complex for lay screeners and many young patients, especially compared to 2WIN and Retinomax; however, rapid estimation of spherical refraction was feasible and valid.</p>","PeriodicalId":93945,"journal":{"name":"Clinical ophthalmology (Auckland, N.Z.)","volume":"19 ","pages":"3153-3166"},"PeriodicalIF":0.0000,"publicationDate":"2025-09-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12415116/pdf/","citationCount":"0","resultStr":"{\"title\":\"Comparative Validation of MOPTIM MRT-200 Refraction Device with 2WIN and Retinomax.\",\"authors\":\"Heidi Schumacher, Giancarlo Garaygay, Aana Ayagalria, Bryant Ikaika Lauano, Denise M Marshall, J Richard Townley, Robert W Arnold\",\"doi\":\"10.2147/OPTH.S545397\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Portable sphero-cylinder refraction is useful for remote medical missions and self-checking. A novel, inexpensive, handheld optical scope with an internal vision chart and adjustable diopter lens, called the Moptim MRT-200, was therefore validated.</p><p><strong>Methods: </strong>Young, capable patients from a pediatric eye practice had dry refraction estimates with the Moptim MRT-200, Adaptica 2WIN photoscreener, and Monocular Retinomax compared with actual refined retinoscopy. Vector-converted sphero-cylinder refractions and the univariate ABCD Ellipsoid were used for comparison as well as monocular visual acuity.</p><p><strong>Results: </strong>206 racially diverse, high-risk refractive patients with median \\\"Ellipsoid-Gap\\\" of 3.5 had high correlation spherical equivalent and similar ABCD Ellipsoid comparing Moptim MRT-200, 2WIN and Retinomax (Kruskal Wallis X<sup>2</sup>(5) = 4.2, p = 0.52) however the 61 patients with blur associated with Ellipsoid gap over 5.0 had better refractive accuracy with 2WIN and Retinomax than Moptim MRT-200 (p < 0.001). Visual acuity with Moptim MRT-200 had a weak correlation (ICC.41 AND.47) and a Bland-Altman mean difference of approximately 1 logMAR line less fine than the actual phoropter refraction.</p><p><strong>Conclusion: </strong>Precise sphero-cylinder estimation with the Moptim MRT-200 was too complex for lay screeners and many young patients, especially compared to 2WIN and Retinomax; however, rapid estimation of spherical refraction was feasible and valid.</p>\",\"PeriodicalId\":93945,\"journal\":{\"name\":\"Clinical ophthalmology (Auckland, N.Z.)\",\"volume\":\"19 \",\"pages\":\"3153-3166\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2025-09-02\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12415116/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Clinical ophthalmology (Auckland, N.Z.)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.2147/OPTH.S545397\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2025/1/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical ophthalmology (Auckland, N.Z.)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2147/OPTH.S545397","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/1 0:00:00","PubModel":"eCollection","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

背景:便携式球柱折射仪用于远程医疗任务和自检。因此,一种新型的、廉价的、带有内部视力表和可调屈光度透镜的手持式光学示波器,被称为Moptim MRT-200,得到了验证。方法:来自儿科眼科诊所的年轻、有能力的患者使用Moptim MRT-200、Adaptica 2WIN光筛和Monocular Retinomax进行干性屈光估计,并与实际的精密视网膜镜检查进行比较。用矢量转换球柱面折射和单变量ABCD椭球进行比较,并对单目视力进行比较。结果:206例不同种族的高危屈光患者中位“椭球间隙”为3.5,与Moptim MRT-200、2WIN和Retinomax相比,具有较高的相关球等效和相似ABCD椭球(Kruskal Wallis X2(5) = 4.2, p = 0.52),而在61例椭球间隙大于5.0的模糊患者中,2WIN和Retinomax的屈光精度优于Moptim MRT-200 (p < 0.001)。与Moptim MRT-200的视力有弱相关性(ICC.41和icc .47), Bland-Altman平均差约为1 logMAR线,比实际的屈光度小。结论:与2WIN和Retinomax相比,使用Moptim MRT-200进行精确的球柱估计对于外行筛查者和许多年轻患者来说过于复杂;然而,球面折射的快速估计是可行和有效的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

Comparative Validation of MOPTIM MRT-200 Refraction Device with 2WIN and Retinomax.

Comparative Validation of MOPTIM MRT-200 Refraction Device with 2WIN and Retinomax.

Comparative Validation of MOPTIM MRT-200 Refraction Device with 2WIN and Retinomax.

Comparative Validation of MOPTIM MRT-200 Refraction Device with 2WIN and Retinomax.

Background: Portable sphero-cylinder refraction is useful for remote medical missions and self-checking. A novel, inexpensive, handheld optical scope with an internal vision chart and adjustable diopter lens, called the Moptim MRT-200, was therefore validated.

Methods: Young, capable patients from a pediatric eye practice had dry refraction estimates with the Moptim MRT-200, Adaptica 2WIN photoscreener, and Monocular Retinomax compared with actual refined retinoscopy. Vector-converted sphero-cylinder refractions and the univariate ABCD Ellipsoid were used for comparison as well as monocular visual acuity.

Results: 206 racially diverse, high-risk refractive patients with median "Ellipsoid-Gap" of 3.5 had high correlation spherical equivalent and similar ABCD Ellipsoid comparing Moptim MRT-200, 2WIN and Retinomax (Kruskal Wallis X2(5) = 4.2, p = 0.52) however the 61 patients with blur associated with Ellipsoid gap over 5.0 had better refractive accuracy with 2WIN and Retinomax than Moptim MRT-200 (p < 0.001). Visual acuity with Moptim MRT-200 had a weak correlation (ICC.41 AND.47) and a Bland-Altman mean difference of approximately 1 logMAR line less fine than the actual phoropter refraction.

Conclusion: Precise sphero-cylinder estimation with the Moptim MRT-200 was too complex for lay screeners and many young patients, especially compared to 2WIN and Retinomax; however, rapid estimation of spherical refraction was feasible and valid.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
4.10
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信