{"title":"种植体宏观设计和植入负荷对人工骨初级稳定性的影响。","authors":"Yiwen Wang, Yoko Yamaguchi, Poyuan Hsueh, Daisuke Higuchi","doi":"10.1186/s40729-025-00644-4","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Implant placement is a critical step for achieving primary stability. During this process, a compressive force, referred to as an \"insertion load,\" is applied through a handpiece or manual driver. However, the influence of the insertion load has not been quantitatively investigated. Since primary stability is essential for predicting successful osseointegration, clarifying the effects of the insertion load may contribute to improving the safety and success of implant treatment.</p><p><strong>Methods: </strong>Straight single-thread implants with pitches of 0.6 mm (S06) and 1.2 mm (S12) and double-thread implants with a 0.6 mm pitch (D06) were used (10 each). Insertion loads of the minimum required load, 5 N, 10 N, and 15 N were applied. Insertion torque values (ITVs), removal torque values (RTVs), and insertion times were measured (p = 0.05). The interface between the implants and the simulated bone was observed with a digital microscope under each load.</p><p><strong>Results: </strong>The minimum insertion load was greatest in the order of S12 < S06 < D06. Variation in the insertion load did not affect the ITV or RTV, but lower loads extended the insertion time. Increased insertion loads led to more voids caused by bone fragment loss, especially in the double-thread designs.</p><p><strong>Conclusions: </strong>While the insertion load did not directly influence the ITV or RTV, it varied by implant design. Higher loads potentially damage the surrounding bone, indicating the importance of careful load management.</p>","PeriodicalId":14076,"journal":{"name":"International Journal of Implant Dentistry","volume":"11 1","pages":"56"},"PeriodicalIF":4.0000,"publicationDate":"2025-08-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12374932/pdf/","citationCount":"0","resultStr":"{\"title\":\"Influence of implant macrodesign and insertion load during implant placement on primary stability in artificial bone.\",\"authors\":\"Yiwen Wang, Yoko Yamaguchi, Poyuan Hsueh, Daisuke Higuchi\",\"doi\":\"10.1186/s40729-025-00644-4\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Implant placement is a critical step for achieving primary stability. During this process, a compressive force, referred to as an \\\"insertion load,\\\" is applied through a handpiece or manual driver. However, the influence of the insertion load has not been quantitatively investigated. Since primary stability is essential for predicting successful osseointegration, clarifying the effects of the insertion load may contribute to improving the safety and success of implant treatment.</p><p><strong>Methods: </strong>Straight single-thread implants with pitches of 0.6 mm (S06) and 1.2 mm (S12) and double-thread implants with a 0.6 mm pitch (D06) were used (10 each). Insertion loads of the minimum required load, 5 N, 10 N, and 15 N were applied. Insertion torque values (ITVs), removal torque values (RTVs), and insertion times were measured (p = 0.05). The interface between the implants and the simulated bone was observed with a digital microscope under each load.</p><p><strong>Results: </strong>The minimum insertion load was greatest in the order of S12 < S06 < D06. Variation in the insertion load did not affect the ITV or RTV, but lower loads extended the insertion time. Increased insertion loads led to more voids caused by bone fragment loss, especially in the double-thread designs.</p><p><strong>Conclusions: </strong>While the insertion load did not directly influence the ITV or RTV, it varied by implant design. Higher loads potentially damage the surrounding bone, indicating the importance of careful load management.</p>\",\"PeriodicalId\":14076,\"journal\":{\"name\":\"International Journal of Implant Dentistry\",\"volume\":\"11 1\",\"pages\":\"56\"},\"PeriodicalIF\":4.0000,\"publicationDate\":\"2025-08-24\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12374932/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"International Journal of Implant Dentistry\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1186/s40729-025-00644-4\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"DENTISTRY, ORAL SURGERY & MEDICINE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Implant Dentistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s40729-025-00644-4","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0
摘要
背景:种植体放置是实现初级稳定的关键步骤。在这个过程中,压缩力,称为“插入载荷”,是通过机头或手动驱动器施加的。然而,插入载荷的影响尚未得到定量研究。由于初级稳定性是预测骨整合成功的关键,因此明确植入载荷的影响可能有助于提高种植体治疗的安全性和成功率。方法:采用节距分别为0.6 mm (S06)和1.2 mm (S12)的单螺纹种植体和节距分别为0.6 mm (D06)的双螺纹种植体各10颗。插入载荷的最小要求载荷,5牛,10牛和15牛的应用。测量插入扭矩值(ITVs)、移除扭矩值(RTVs)和插入次数(p = 0.05)。在不同载荷下,用数码显微镜观察种植体与模拟骨的界面。结论:种植体植入负荷对ITV和RTV无直接影响,但因种植体设计而异。较高的负荷可能会损害周围的骨骼,这表明小心负荷管理的重要性。
Influence of implant macrodesign and insertion load during implant placement on primary stability in artificial bone.
Background: Implant placement is a critical step for achieving primary stability. During this process, a compressive force, referred to as an "insertion load," is applied through a handpiece or manual driver. However, the influence of the insertion load has not been quantitatively investigated. Since primary stability is essential for predicting successful osseointegration, clarifying the effects of the insertion load may contribute to improving the safety and success of implant treatment.
Methods: Straight single-thread implants with pitches of 0.6 mm (S06) and 1.2 mm (S12) and double-thread implants with a 0.6 mm pitch (D06) were used (10 each). Insertion loads of the minimum required load, 5 N, 10 N, and 15 N were applied. Insertion torque values (ITVs), removal torque values (RTVs), and insertion times were measured (p = 0.05). The interface between the implants and the simulated bone was observed with a digital microscope under each load.
Results: The minimum insertion load was greatest in the order of S12 < S06 < D06. Variation in the insertion load did not affect the ITV or RTV, but lower loads extended the insertion time. Increased insertion loads led to more voids caused by bone fragment loss, especially in the double-thread designs.
Conclusions: While the insertion load did not directly influence the ITV or RTV, it varied by implant design. Higher loads potentially damage the surrounding bone, indicating the importance of careful load management.
期刊介绍:
The International Journal of Implant Dentistry is a peer-reviewed open access journal published under the SpringerOpen brand. The journal is dedicated to promoting the exchange and discussion of all research areas relevant to implant dentistry in the form of systematic literature or invited reviews, prospective and retrospective clinical studies, clinical case reports, basic laboratory and animal research, and articles on material research and engineering.